Healthcare Technology and Function Flashcards
You have a reading on 2 different manufacturer’s pulse oximeters. The nurse says they are reading wrong because there is a 2% difference in the blood oxygen levels between them. What is wrong?
a. Both of them need to be re-calibrated
b. Trust the one from a known manufacturer
c. This is normal operation
d. There is too much light from overhead affecting the readings.
c. This is normal operation
A sterilizer has a jacket pressure of 30 psi and a chamber pressure of 28 psi. What is wrong with it?
a. This is normal operation when a cycle is running.
b. The chamber solenoid is defective
c. The jacket solenoid is defective.
d. The overpressure solenoid is defective
a. This is normal operation when a cycle is running.
The EEG is used to detect:
a. Murmurs
b. Arrhythmias
c. Brain Tumors
d. Dermatitis
c. Brain Tumors
Steam Sterilization Temperatures are usually above:
a. 62 Degrees Celsius
b. 102 Degrees Celsius
c. 132 Degrees Celsius
d. 162 Degrees Celsius
c. 132 Degrees Celsius
Ultrasound is best described as:
a. Radiation
b. Flow of Sonic Particles
c. Electromagnetic Waves
d. Acoustic Compressions
d. Acoustic Compressions
During Synchronous Cardioversion
a. The defibrillator is set to fire approx 20ms after the peak of the “R” wave
b. The heart is in fibrillation
c. The heart is sometimes in fibrillation
d. The defibrillator is set to fire approx 20ms before the peak of the “R” wave
a. The defibrillator is set to fire approx 20ms after the peak of the “R” wave
What is the Highest Temp that an infant warmer should be before the alarm is tripped?
a. 37 degrees Celsius
b. 39 degrees Celsius
c. 43 degrees Celsius
d. 50 degrees Celsius
b. 39 degrees Celsius
The treatment of high bilirubin in babies is treated by;
a. Blood Transfusion
b. Drug Therapy
c. Bile Duct surgery
d. Light Therapy
d. Light Therapy
Ultrasound Energy for imaging is produced by;
a. Piezoelectric Crystal
b. Charging Amp
c. Feedback oscillator
d. Doppler effect
a. Piezoelectric Crystal
What is the most life-threatening problem associated with an IV Pump?
a. Air Embolism
b. Free uncontrolled flow to the patient
c. Occlusion to patient
d. Vein closing when infusion is complete
a. Air Embolism
A paper speed of 25mm/sec is set on an ECG recorder. The patient’s heart rate is 60 beats per minute. In a normal ECG waveform, howoften will the R wave show up on the paper?
a. Every 25 cm
b. Every 3 big boxes on the paper
c. Every 5 big boxes on the paper
d. Ever 20 mm on paper fed
c. Every 5 big boxes on the paper
Using your test equipment you feed a 1milli volt signal into ECG leads.You have the monitor set to read on lead 2. You are using a 3 lead set. The monitor ECG size is set to 1 milli volt. You run a strip on the recorder. What should you see for amplitude on the recording?
a. One small square on the paper positive deflection
b. One big square on the paper positive deflection
c. Two small squares on the paper positive deflection
d. Two big squares on the paper positive deflection
d. Two big squares on the paper positive deflection
What selection on an electro surgical unit most resembles a sine wave?
a. Pure cut
b. Coag
c. Blend 1
d. Bipolar coag
a. Pure cut
The purpose of a grid on an x-ray bucky is to:
a. Stop scatter radiation from going to the film or detector
b. Focus the x-ray beam
c. Magnify the x-ray beam
d. Add lines to the film for measurement purposes
a. Stop scatter radiation from going to the film or detector
The normal test load for a defibrillator is:
a. 50 ohms
b. 75 ohms
c. 100 ohms
d. 50 ohms non-inductive
d. 50 ohms non-inductive
An ECG monitor acquires the respiration wave and rate from:
a. ECG lead 1
b. ECG lead 2
c. ECG lead 3
d. Calculates it with a formula based off of heart rate
b. ECG lead 2
A tourniquet stops the flow of blood to a patient’s body part that isbeing operated on by:
a. Pumping up to a pressure lower than the patient’s blood pressure and restricting flow.
b. Pumping up to a pressure higher than the patient’s blood pressure and restricting flow.
c. Pressurizing and de-pressurizing based on the needs of the patient and the surgeon.
d. Alarming when it is over or under pressure.
b. Pumping up to a pressure higher than the patient’s blood pressure and restricting flow.
A centrifuge spins at 3432 RPM at one scheduled preventive maintenance check. This centrifuge is checked on a quarterly schedule. At the next scheduled preventive maintenance it reads 2600 RPM. What would the most likely case for this problem be?
a. Worn armature.
b. Defective motor speed controller circuit.
c. Different rotor head was installed.
d. The wrong weight of sample tubes being used in the centrifuge.
c. Different rotor head was installed.
A nurse calls you to repair a non-invasive blood pressure unit. The nurse tells you that the blood pressures are too high when using this unit. What is the most likely cause of this problem?
a. Blood pressure cuff is too big.
b. Blood pressure cuff is too small.
c. The blood pressure algorithm is set up wrong and needs to be recalibrated.
d. The pressure sensor is defective in the unit and needs to be replaced.
b. Blood pressure cuff is too small.
Name the A) standard speed and B) two optional speeds on an ECG machine.
a. 25 mm/s, 10 and 50 mm/s
b. 25 mm/s, 50 and 100 mm/s
c. 50 mm/s, 10 and 25 mm/s
d. 10 mm/s, 25 and 50 mm/s
b. 25 mm/s, 50 and 100 mm/s