HARDWARE QA/QC Flashcards
The location of the coils installed in the MRI scanner, from the inner point closest to the patient to the outer most edge, are:
- RF radiofrequency coils
- gradient coils
- shim coils
- main magnet
The gradient rise time is defined as
the time it takes for the gradient to reach its full amplitude
measured in microseconds — mT/m (milliTesla/meter) or G/cm(Gauss/centimeter).
The** gradient slew rate** is
the speed rate of ascent or descent of a gradient from zero to its maximum amplitude, either positive or negative
Which is the amplitude divided by the rise time in msec. Measured in mT/m/msec or T/m/sec.
The duty cycle is defined as
the time the gradients are on during a TR period
“gradient working time”
Parameter changes affecting the pulse duty cycle include:
- Increased # slices
- Employment of fat suppression pulses (SPAIR requires increased TR
compared to SPIR) - Utilization of presaturation slabs/bands
- Increased ETL
The primary cause of the noise in the MR scan room is
the rapid, successive switching of the gradient coils
RF heating is more of a concern in ________ sequences due to the multiple echo train lengths, leading to potential increases in tissue heating.
fast spin echo
Doubling field strength results in ____ increase in SAR potential.
fourfold (4x)
The primary biological effect of radiofrequency fields is
tissue heating
Consistent room temperature of ____________ and sufficient air flow in the MR scan room is recommended to make patients less vulnerable to the RF heating effects.
65-75°F, humidity of 50-70%
For energy to be efficiently transferred between a transmitter and receiver, the two must be
at the same frequency.
Utilization of ___________ yields a more localized, smaller FOV (field of view) capability, with increased SNR, thus providing opportunity for improved image quality on most systems.
surface coils (linear coils)
A 6 inch surface coil is able to receive signal ____ inches deep into the patient
3
(SNR penetration depth = ½ coil diameter)
Circularly polarized transmission or receiver coil with two orthogonal
transmission and/or receiver channels are
CP coil, also known as a quadrature coil
This yields better signal-to-noise than a linear coil
When going from a linear coil to a quadrature coil, SNR is increased by
40%
A ________ is comprised of multiple coil elements combined with multiple receiver channels.
phased array coil
The patient image display with sequence parameter visualization is located in the
scan control room on the operator console
The CPU used to plot and adjust slices in an MR sequence is located in the
scan control room at the technologist operator console
The array processor is responsible for reconstructing the collected MR
images/data using the Fourier transform, and is located in
the MR equipment room.
The RF power amplifier is located within
an equipment room