hardware Flashcards
consistent room tem of 65*75 degree f, humidity of 50-70% and sufficient air flow in the mr scan room is recommended to
make patients less vulnerable to the RF heating effects
the liquid cryogens are cooled in a superconducting magnet at a temperature of
4 degree K
during q quench of the mr system, pts and employees should be evacuated from the scan room to avoid
asphyziation, frostbite and tympanic membranes damage
the intensity limit for a static magnetic field in clinical use is limited by the food and drug administration to
4 T
Radiofrequency shielding can be achieved by lining the scanner room walls with
copper
the location of the coils installed in the MRI scanner from the inner point to the outer most edge are
radiofrequency coils, gradient coils. shim coils, main magnet
the coils construction in an MR system from the innermost aspect to the outer
radiofrequency, gradient. shim. main magnet
doubling the magnetic field strength will
yield an increase of double the SNR
passive shielding can be accomplished by lining the mr room with
steel
active shielding can be defined as
superconducting windings within the mr scanner reducing the fringe magnetic field
_____ could be used to line the walls and or floors of the mr magnet room to reduce the fringe field and is defined as______ shielding
steel; passive
____would be used to contain the RF in an MR magnet room
copper