Fourier Transform Infrared Analyzers Flashcards

1
Q

An ______________ creates interfering mixtures of IR radiation to send through the sample

A

interferometer

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

What category of IR detector is used in an FTIR analyzer and what is the name for the detector signal output over time?

A

An FTIR uses a wideband non-specific detector that outputs an interferogram

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

What can cause the calibration of an NDIR analyzer to drift over time?

A

1) Dirty sample cell
2) Weak IR source
3) Decrease in detector sensitivity

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Explain what would happen to a gas analyzer’s readings if the pressure in the sample cell increased.
Explain what would happen if the temperature in the sample cell increased.

A

Pressure increase: causes density increase therefore increasing the concentration readout

Temperature increase: causes a density decrease therefore decreasing concentration readout

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Identify three common NDIR analyzer maintenance activities.

A

1) Replace faulty IR source bulbs
2) Replace faulty detectors
3) Clean cell windows

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

How are the chemicals in a sample identified by FTIR?

A

A program that assigns each chemical an absorption frequency

How well did you know this?
1
Not at all
2
3
4
5
Perfectly