Fluids 2 Flashcards
What should you always do when moving patients on IV fluids? Why?
Always turn off fluid infusion
Risks fluid overload
What should you do with short term moving of a patient (from induction to OR or recovery)
Close slide clamps in 2 sites if possible
Closest to patient and closest to IV bag
What do you do when moving a patient on IV fluids for long periods of time
Turn off flow
Close all the clamps (closest to patient and bag)
Disconnect extension set from IV line (put on sterile caps on both ends)
Secure extension set to patient
Hang IV line from pole with bag
What do you need to do when reconnecting catheters
Flush the catheter to ensure it has not clotted
If it is patent, reconnect drip line to extension set, Open clamps, reset drip rate
Save sterile caps for re use
What is a Y injection port used for
To inject medications and other fluids into the line to be infused into patient
Can be needle ports or needless ports
Always wipe port before use
What is an IV push/bolus
Giving a medication all at once over 30-60 seconds
Must clamp or pinch line above the port
The must flush line to get drug into patient (flush with double the volume of the line between the port and the patient) flush over same amount of time as drug was given)
What is slow IV
one shot of medication given over 10-20 minutes
These drugs have more toxicity so they need to be given even slower
What are infusion pumps AKA volumetric pumps
Forces fluid into the vein under pressure -does not use gravity
Consistent and accurate
Enter infusion rate into pump (ml/h)
Problems:
Risk of infiltration (won’t stop if fluid is running)
May not alarm with problems
Occlusion
What is the minimum infusion rate needed to keep veins open
5ml/h
What is a volume-controlled burette (buretrol)
Gravity based, placed between the IV bag and patient
Used for greater accuracy in measuring smaller volumes and delivering medications
Commonly used with chemotherapy drugs
What are flow regulators
Part of an infusion set or can be added as an extension set
Regulates flow of IV from the infusion set to the cannula
Controls the size of opening that the line runs through
Gravity sets only
No drips to count
Single use only
Explain a piggy back system
Purpose: increases the amount of reservoir fluids and to deliver an intermittent drug or second type of fluid
Each bag has its own drip chamber
Second bag is placed higher than the primary bag
Secondary/higher bag will empty first and then the primary bag will flow
Requires a specific piggyback infusion set with a Y port or split
What are in-line filters
Precision filters as small as 0.2 micrometers
Placed in line
Can remove air or particulate matter
Blood transfusion sets have a 160-270 micrometer filter to remove small clots and clumps in blood due to storage
What is a syringe pump
Delivers small volumes of fluid in a syringe at a constant rate over a period of time
Volume determined by size of syringe
Used for:
Medications that can be diluted in a IV bag
Small patients
Slow IV
What is CRI
Constant/continuous rate infusion
Dosing regimen used to deliver a constant amount of drug per unit of time
Usually ml/kg/h
Used in IV infusion (drip rates, syringe pumps, transdermal patches)
Flush is normally done with
3-5 ml of 0.9% saline or LRS