Flouride Flashcards

1
Q

How often does the Fluoride meter get calibrated?

A

once per shift (every 8 hours)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

How often does the Linearity check (1.0 check) get done?

A

Weekly

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

How often does a QC sample check occure?

A

Monthly

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Maximum hold time for fluoride samples?

A

Immediate analysis recommended, but 48 hours

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

When does the TISAB solution exp?

A

1 year after opening

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Volume of sample and TISAB

A

25 mL of each, for a combined total of 50 mL

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Calibrations values

A

0.5 mg/L and 5.0 mg/L (1.0 mg/L once a week)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Slope acceptance requirements

A

-54.0 to -60.0 mV

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Acceptance criteria for 1.0 standard

A

+/- 10% (0.9-1.10)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly