Exam 2 - Heater-Coolers Flashcards
1
Q
Heater-Cooler Functions
A
- Heat/cool Cardioplegia (usually cool)
- Regulate fluid temps going into and out of patient
- Patient blanket temp regulation
2
Q
SARNS
A
- Manufacturer of Cardiovascular equipment
- Purchased my Terumo
3
Q
Jostra
A
- Purchased by Maquet
- Common in PEDs
4
Q
Lobe
A
- Made heart-lung machines and disposable supplies
- Purchased by Sorin….and then by LivaNova
5
Q
LivaNova
A
- Most popular Heater-Cooler
- 3T
- 60% market share
6
Q
TCM
A
Temp control module
7
Q
Common Heater Coolers
A
- Terumo TCM II
- Cincy sub zero
- only one output…cannot separately control cardioplegia - Terumo dual H/C
- can only heat/cool one system at a time….cooled manually via ice - LivaNova 3T
8
Q
Why clean Heater-coolers?
A
- NTM - Non-tuberculosis micro bacteria
……can lead to TB in patients
9
Q
Cleaning 3T
A
- Fill with filtered water / Clorox solution (180ml Clorox)
- cycle between cardioplegia (10C) & patient 1 (20C) for 5min
- Create short circuits for cardioplegia, patient 1, patient 2
- open all valves and cycle 10 min
- drain device completely
- refill w/ filtered water and cycle all circuits 3 minutes
- drain device completely
- refill w/ filtered water and cycle all circuits 3 minutes
- drain device completely
- refill device with filtered water and 150 ml hydrogen peroxide
- create short circuit from cardioplegia to patient 1
- cycle 5 min and power off