Exam 2 - Heater-Coolers Flashcards

1
Q

Heater-Cooler Functions

A
  • Heat/cool Cardioplegia (usually cool)
  • Regulate fluid temps going into and out of patient
  • Patient blanket temp regulation
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

SARNS

A
  • Manufacturer of Cardiovascular equipment

- Purchased my Terumo

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Jostra

A
  • Purchased by Maquet

- Common in PEDs

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Lobe

A
  • Made heart-lung machines and disposable supplies

- Purchased by Sorin….and then by LivaNova

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

LivaNova

A
  • Most popular Heater-Cooler
  • 3T
  • 60% market share
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

TCM

A

Temp control module

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Common Heater Coolers

A
  • Terumo TCM II
  • Cincy sub zero
    - only one output…cannot separately control cardioplegia
  • Terumo dual H/C
    - can only heat/cool one system at a time….cooled manually via ice
  • LivaNova 3T
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Why clean Heater-coolers?

A
  • NTM - Non-tuberculosis micro bacteria

……can lead to TB in patients

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Cleaning 3T

A
  • Fill with filtered water / Clorox solution (180ml Clorox)
  • cycle between cardioplegia (10C) & patient 1 (20C) for 5min
  • Create short circuits for cardioplegia, patient 1, patient 2
  • open all valves and cycle 10 min
  • drain device completely
  • refill w/ filtered water and cycle all circuits 3 minutes
  • drain device completely
  • refill w/ filtered water and cycle all circuits 3 minutes
  • drain device completely
  • refill device with filtered water and 150 ml hydrogen peroxide
  • create short circuit from cardioplegia to patient 1
  • cycle 5 min and power off
How well did you know this?
1
Not at all
2
3
4
5
Perfectly