Exam 2 Flashcards
What are most gauges calibrated for?
Atmospheric pressure
What does non-liquified vs liquefied compress gas give us?
the relationship between pressure and remaining volume and pressure reading on the gauge
What do we apply to liquify a gas?
apply a low temp
Which gases do not liquefy at room temperature regardless of pressure applied?
O2 and Medical Air
What occurs to non-liquified gas contents as volume decreases?
Pressure decreases
What type of relationship is seen between pressure and volume for non-liquified compressed gases?
Linear (P1/v1=p2/v2)= Boyle’s Law
What is a liquefied compressed gas?
A gas that becomes liquid at ambient temperatures and at pressures from 25-1500 PSI
Give an example of liquefied compressed gas that we use
Nitrous oxide
What does the pressure of a liquefied compressed gas depend on?
Vapor pressure which is not an indication of remaining volume
At what level does a liquified compressed gas have to be at to see a pressure change on the gauge?
Over 95% empty
What does the food and drug administration regulate for gas cylinders?
Gas purity
What does the department of labor/OSHA regulate for gas cylinders?
Employee safety
What does the department of transportation regulate for gas cylinders?
Marking, labeling, storing maintenance, transportation and disposition
How often are gas cylinders inspected and tested?
Once every 10 years
What is required of a medical gas cylinder for MRI?
Aluminum = 3AL or 3 ALM
Where would we find the conical depression that fits the retaining screw on the yoke of gas cylinders?
On the cylinder on the back of the machine
4 points about the handle of gas cylinders
- Opens/closes cylinder
- Turns counterclockwise to open for safety
- Also called cylinder wrench
- Must have one for every machine to be readily used
What does the safety relief component of medical gas cylinders allow for?
Spring-loaded mechanism to allow venting of gas, if pressure is increased it allows gas venting and then recloses/seals after pressure is normalized inside cylinder but leaks can occur
What is the ideal gas law?
(P1V1/T1)=(P2V2/T2)
What is the most common gas cylinder size?
E
Color, liters when full and PSI of O2 cylinder
- Green
- 660L
- 1900 PSI
Color, liters when full and PSI of Air cylinder
- Yellow
- 625L
- 1900 PSI
Color, liters when full and PSI of Nitrous Oxide
- Blue
- 1590L
- 745 PSI
What was found regarding cylinders and irregularities?
Some have irregularities or faulty constructions and should be inspected by the facility as well prior to use
Which cylinder must be secured/locked up to prevent access and abuse?
Nitrous oxide
Where are cylinders not to be stored?
In the OR, they must have a designated storage area
5 points to consider when opening a gas cylinder?
- Open valve slowly and slightly prior to installation to clean out the valve port
- Check pressure
- Open away from patient
- Face valve away from people
- Correct any leaks
What piece of a cylinder if missing could potentially cause a leak?
Sealing washer
How many different pin positions are possible for gas the PISS?
7
What is PISS?
Pin Index Safety System (PISS)
Why are pipeline systems used?
Pipeline/central gas supply systems are used to deliver gas to anesthetizing locations and patient care areas because e-cylinder use is not enough
What is the pipeline pressure in the US?
380 KPA or 55 PSI
What is the most frequently reported problem in pipeline systems?
low pressure
What is required of pipeline systems?
Must have a two, 2 day supply banks, one primary and one reserve for a total of 4 days for oth
What is the reserve supply of pipeline systems used for?
Emergencies or failure of primary supply, they are in different areas with different routing
What must occur with liquid oxygen for it to be cost effective?
It must be in constant use to be cost effective or pressure increases as the liquid boils and is then vented in the atmosphere
Why is N20 supplied by a manifold cylinder system?
Because regulator prone to freezing
What do air supply sources use?
Manifold or compressors, important for intake locations are free of contaminants
What do air systems needs to do to qualify for medical use?
Need to dehumidify
What do N20 and Air supply systems have?
A series of valves, pressure regulators, and alarms just like our anesthesia machines to regulate pressure and signify problems
Where do the pipeline systems terminate?
In the OR or ICU
What is an important regulation for pipeline systems?
Name, pressure and flow direction must be clearly marked every 20 feet and in each room
Compare oxygen tank diameter to other gases
It has a different outer diameter than other gases
What do shut off valves allow for?
Certain areas in the piping system to be isolated for maintenance or problems
3 Mandatory locations for shut off valves
- Main supply into building which turns everything off
- One at each riser
- One at each branch except if branch is to anesthetizing area or critical care area
When are pipeline area alarm systems checked? tested?
Checked daily and tested monthly
When do pipeline area alarms sound?
If the pressure increases/decreases at 20% from normal line pressure
3 requirements for pipeline alarm systems
- Must be audible and visible
- Labelled for gas and area
- Personnel must know what to do if sounds
6 Conditions for pipeline alarm systems to sound
- Main supply reaches the daily average
- Main supply not at normal operating pressure
- Reserve supply has been reduced to one day supply
- Reserve supply being used
- Reserve supply decreased pressure
- Humidity of medical air unsuitable
What type of connections do gases have into the wall?
Connection into wall uses quick connectors that have a pair of male and female parts that only connect when the proper alignment is achieved. Each gas has a specific shape and spacing but are ore prone to leaks Vs DISS system
What type of connection do gases use into the machine?
Station outlets uses DISS (Diameter index safety system)
What is a vaporizer?
A device that changes a liquid anesthetic agent into a vapor for inhalation
What must we add to together in a breathing system?
Controlled amount of vapor to fresh gas flow
What are vaporizers calibrated at?
Sea level and is affected by barometric pressure changes
What is our responsibility in regards to vaporizers?
To acquaint ourselves with the vaporizers at your facility
What are the 4 stages of anesthesia?
- Analgesia
- Delirium
- Surgical
- Respiratory cessation
What is the vapor pressure of a liquid?
The equilibrium pressure of the vapor above its liquid
What does the pressure of the vapor result from?
Evaporation of the liquid above the liquid in a closed system at a constant temperature
What does increasing temp do to vapor pressure?
Increasing temp increases vapor pressure, they have a linear relationship
What is the effect on vapor pressure from barometric pressure?
VP is not influenced by barometric pressure
What is vapor pressure dependent on?
liquid and temperature
When is the gas phase above the liquid is said to be saturated?
When it contains all vapor it can hold at a given temperature (Saturated vapor pressure SVP)
What is the saturated vapor concentration (SVC) calculated by
SVP/Atmospheric pressure (must be diluted by a bypass gas flow)
trade name and vapor pressure of isoflurane?
- Forane
2. VP: 238 Torr
trade name and vapor pressure of desflurane?
- Suprane
3. 664 torr
trade name and vapor pressure of sevoflurane?
- Ultane
2. 160 torr
What do we do to common volatile agents to level the playing field?
We express gas concentration in partial pressure and volumes percent (MAC)
Two forms of expression of gas concentration?
- Volumes percent
2. Partial pressure
Describe volumes percent expression of gas concentration
Concentration of a gas in a mixture expressed as a percentage of 100% at 1 atm (PP/TP * 100%)
What is MAC described in terms of
Volume percent
What does volumes percent indirectly relate to
patient uptake and anesthetics depth
Is volume percent influenced by barometric pressure?
yes
Describe partial pressure expression of gas concentration
The pressure exerted by any one gas in a gas mixture on the total gas mixture
What is partial pressure dependent on?
Temperature
What does partial pressure directly relate to?
Patient uptake and anesthetic depth
How does the potency of the partial pressure of gases change when barometric pressure increases or decreases?
No change, there is the same potency no matter what
Definition of heat vaporization
The number of calories necessary to convert 1G or 1mL of liquid into vapor
Where does heat flow from to compensate for the loss of heat in the liquid from entering the gas phase?
surrounding vaporizers
How is equilibrium established with heat vaporization?
Heat lost to vaporization = heat supplied by surrounding vaporizer
How do manufacturers account for the cooling of vaporizers?
they used certain metals
What is the definition of specific heat?
The quantity of heat required to raise the temperature of 1 gram of a substance 1 degree celsius
What does it mean if something has a higher specific heat?
temperature changes more gradually
What is the standard for specific heat?
H20 is the standard, 2 cal per gram per 1 degree
What must specific heat be considered for?
Maintaining a constant temperature to the vaporizer
What is the definition of thermal conductivity?
The speed of which heat flows through a substance
What does higher thermal conductivity mean
Better conductor of heat
What must be considered in choosing a material/metals for vaporizer construction?
Specific heat and thermal conductivity
What do we want in regards to specific heat and vaporizers?
Want a higher specific heat because it is harder for the temperature to change, which means the vaporizer is less likely to change with gas leaving
What are the two kinds of vaporizer design?
bypass vaporizers and electronic vaporizers
How much is FGF through a pipeline system?
35-75L/min
Why are vaporizers placed where they are?
So that liquid can become vapor and then saturate the body
Calculation of vaporizer output?
At a steady state, the total volume of gas leaving the vaporizer is greater than the total volume that entered due to the additional volume attributed to anesthetic vapor at its SVC (saturated vapor concentration)
Compare vapor pressure of an anesthetic gas at room temp to the partial pressure necessary to achieve anesthesia
It is more, that is why the vaporizer dilutes the anesthetic to a useful concentration (it is more potent)
What is the splitting ratio?
Some gas flows into vaporizer chamber, some gas flows into bypass
What does the splitting ratio depend on?
The anesthetic agent/concentration, size of the adjustable orifice, total gas flow, and heat of vaporization
Does the temperature compensating mechanism for vaporizers produce instantaneous results?
No, the anticipated concentration may not be accurate until compensation occurs
Describe bypass vaporizers
No volatile is attached to flow through bypass chamber vs vaporizing chamber
With bypass vaporizers, what does the splitting ratio equal?
vaporizing/bypass
What does increasing the splitting ratio mean with bypass vaporizers?
More fresh gas goes into vaporizing chamber and therefore more goes to the patient
What happens to the splitting ratio if you cool the bypass vaporizer?
it gets smaller, so less goes to to the patient, you need to increase MAC and increase FGF
Describe electronic vaporizers
Computer drive, calculates either volume of carrier gas to produce the desired concentration or the amount of liquid agent needed to be injected into carrier flow
What is the desired concentration based on from electronic vaporizers?
What you have dialed into the dial or machine
Describe injection of vaporizers
Inject known volume liquid anesthetic into known volume of gas
Describe flow-over of vaporizers
Carrier gas passes over surface area of a liquid; increasing surface area increases the efficiency of vaporization
How do we fix the energy lost in vaporization?
Thermocompensation
What does heat of vaporization (HOV) do to the vaporizer?
Decreases the temp and decreases output of gas/vapor and splitting ratio will decrease without thermocompensation, meaning no steady concentration is given to the patient
What is an example of a gas that we use supplied heat to maintain heat?
Desflurane
Where might intermittent back pressure (IBP) come from?
O2 + OR + Pressure from inspiration during ventilator use
What can IBP cause?
A pumping effect or pressurizing effect on vaporizer outputs
What does IBP most commonly have to do with?
Ventilator and flush valve use
What is the end result of the pumping effect from IBP?
Increase of vapor output, causing more to get to the patient
What is the pumping effect due to?
Back pressure during inhalation at low FGF causing increased flow into the vaporizing chamber so more than usual vapor is picked up
When is the pumping effect more commonly seen?
Low FGF, large pressure fluctuations, and low vaporizer settings
What helps minimize the pumping effect?
Presence of the pressurizing valve, unidirectional valve, and pressure relief valve
What is the end result of the pressurizing effect?
decrease in vapor output causing decreased delivery to the patient
What is the pressurizing effect due to?
Back pressure at high FGF causes increased density into vaporizing chambers so less than usual vapor picked up
When is the pressurizing effect more commonly seen?
High gas flows, large pressure fluctuations, and low vaporizer settings
What does FGF directly affect?
Vaporizer output
How many L/min is considered high fresh gas flow?
10L/min
When is there a higher chance of gas being rebreathed by the patient, in high fresh gas flow or low fresh gas flow?
Low fresh gas flow
With what type of fresh gas flow does inspired concentration = vaporizer setting?
high fresh gas flow
With what type of fresh gas flow do we need agent analyzer to get the true value of gas getting to the patient?
low fresh gas flow
Describe high FGF
Gas delivery is higher than patient minute volume, there is no recycling and the patient is getting closer to exact gas concentration and are getting what is on the dial
When is low FGF used?
maintenance phase of anesthesia
How many L/min would be considered low fresh gas flow?
2L/min
What is low fresh gas flow the difference between?
Vaporizer setting and inspired concentration
Which type of fresh gas flow does it take longer for the patient to reach equilibrium?
low fresh gas flow
6 standard regulations for vaporizers
- Average concentration +/- 20% setting
- Gas may not pass through more than 1 vaporizer (safety)
- Output of vaporizer <0.05% in off
- All control knobs turn counterclockwise
- Filling levels displayed
- Can not overfill when in normal operating position
What is the most common type of vaporizer mounting system?
Detachable
When might it be a huge benefit for easy removal of vaporizing mounting systems?
malignant hyperthermia
What does the interlock device for vaporizers prevent?
More than 1 vaporizer being turned on at a time
3 vaporizer hazards?
- Incorrect agent
- Tipping
- Overfilling
2 steps that must be taken if contaminated filling occurs of vaporizers
- Must be completely drained and all liquid discarded
2. FGF run until no vapor detected
Describe tipping of vaporizers
Liquid may get into bypass our outlet causing an increased concentration of agent
What mode should vaporizers be placed in prior to moving to prevent tipping?
Off/travel mode
What do you do if a vaporizer is tipped?
High FGF run with low concentration of vapor until excessive vapor exhausted
3 descriptors of overfilling vaporizers
- Liquid may enter fresh gas line or cause vaporizer failure
- Potential for lethal dose
- Can occur during tipping or filling with the on setting
2 reasons why leaking of vaporizers happens
- Failure to tighten filler cap
2. Fill valve not closed - malfunctioning mount/vaporizer pollutes or can probably smells
What is the most common cause of no vapor output?
vaporizer is empty
3 reasons for no vapor output
- empty
- incorrect mounting
- overfilled = no output because of vaporizer failure
What are anesthesia gas delivery systems composed of?
Anesthesia machine, vaporizers, ventilator, breathing circuit, and scavenging system
Describe the gas delivery from start to end (6)
- Gas source delivers gas to machine
- Gas delivered to flowmeters and vaporizers
- Gas mixture goes to common gas outlet (CGO)
- Gas flows to breathing circuit to the pat
- Gas leaves pt through breathing circuit
- Excess gas either via the APL valve/scavenger or ventilator
Who are the two large manufacturers of anesthesia gas delivery systems in the united states
Draeger Medical and GE
What do older anesthesia machines lack that newer machines do have?
Safety features
What did older anesthesia systems require to work?
Supply of gas under pressure and were pneumatic, they had no electrical functions
What electrical components of an anesthesia machine are active if the master switch is off?
Battery charger and electrical outlets for additional monitors
What pneumatic functions are maintained when the anesthesia machine is off?
Oxygen flush valve and axillary oxygen flowmeter
What occurs when the master switch of an anesthesia machine is on?
the electronics go through a powering up protocol that usually includes an automated machine checkout
When can the anesthesia machine checkout be overridden?
In an emergency
What permits the delivery of gas from the flowmeters and vaporizers?
The pneumatic functions on anesthesia machines
What do the red outlets connect to?
Back up generator
What do all contemporary anesthesia machines incorporate?
Electrical systems and require a connection to electrical power
What occurs if requirements exceed the anesthesia machine outlet?
circuit breaker will activate
When should anesthesia machine checkouts occur?
Take place before the first case of the day and if any changes are made to the system
What does the abbreviated check of the delivery stems between cases include?
High pressure check to check for pressure problems in the low pressure system
What is one of the most important steps often overlooked in the anesthesia precheck?
Machine and airway, the immediate presence of ventilation modalities in the event of a power outage or failure oxygen cylinder supply and ambu bag
What is Ohm’s Law?
Flow = Pressure/Resistance
What is the basic function of any anesthesia machine?
To receive compressed gases from their source and create a gas mixture and flow rate at the CGO to deliver to the patient
What is required of yokes for oxygen and nitrous tanks?
There must be one for each
What does the hanger yoke assembly prevent?
Prevents gas from being transferred from a cylinder with higher pressure to one with lower pressure if both are in a yoke and ON thus prevents the unnecessary depletion of gas
What is the gauge pressure?
measured pressure of a gas above ambient atmospheric pressure
3 descriptors of bourdon tubes (curved hollow tubes) of cylinder pressure indicator/gauges
- Increase pressure straightens curve
- Falling pressure causes curve to redevelop
- Motion is transmitted to gauge
What is a pressure regulator?
Device that converts a high, variable input gas pressure to a constant, lower output presssure
Why are pressure regulators also called reducing valves?
Reduces high, variable pressure in cylinders to lower constant pressure for machine
Which gases are regulators required for?
Each gas supplied by a cylinder
Why do reducing valves help to prevent?
They make it so we do not have to constantly adjust flowmeter to provide constant flow
What gas specific connections are used for entry point of gases from pipelines?
DISS (Diameter index safety system)
Which pipeline inlet connections are required?
Oxygen and nitrous
What must pipeline inlet connections contain?
Unidirectional check valves to prevent gas returning or flowing back into the pipeline
When will the indicator of pipeline pressure be most accurate?
When the cylinder is 0
Describe a high pressure system
Includes parts of upstream of the cylinder pressure regulator aka first stage regulator
What is the first stage regulator?
Converts high pressures to constant lower pressure of 45psi, which is oxygen cylinder pressure regulator
High pressure gas systems have oxygen pressure between what psi’s?
45-2200 psi
Describe intermediate pressure systems
includes parts between cylinder pressure regulator and pipeline gas inlet to gas flow control valves, with oxygen pressures between 16 and 55 psi, can flow and pressurize gas in multiple directions
How does gas enter the machine from the first stage regulator or cylinder pressure regulator?
through intermediate pressure system
What is the intermediate pressure system pressure if the master switch is off?
0
Why is there a deliberate difference in supply pressures between the pipeline (50-55psi) and cylinder oxygen (40-45psi)
Because if the cylinders are open, the machine will preferentially receive oxygen from the pipeline due to this pressure difference
If pipeline pressure drops below the cylinders what occurs?
The pipeline will preferentially receive oxygen from the cylinder
What should be done to the gas cylinder to prevent exhaustion and leakage of gases from cylinders?
It should be turned off
Describe low pressure systems
includes all parts downstream of the gas flow control valves, so extends from flowmeters to CGO with pressures normally slightly greater than atmospheric pressure
Describe pressure in low pressure systems
Pressure is variable and depends on flow from flowmeters and back pressure from breathing circuit
5 parts that low pressure systems include
- Flowmeters
- Hypoxia prevention devices
- Unidirectional valves
- Pressure relief valves
- Common gas outlet (CGO)
What should machine piping leak no exceed?
25mL/min inside machine
Describe the common gas outlet (CGO)
Receives all as from machine and delivers mixture to breathing system to deliver to the patient
What should the common gas outlet not be used for?
supplemental oxygen
What is the unidirectional valve located at the pipeline inlet used for?
to prevent pipeline back flow
Where is the outlet check valve?
Upstream from oxygen flush valve
What does the outlet check valve being upstream from oxygen flush valve prevent/lessen?
Back pressure from oxygen flush or breathing circuit so it prevents reverse gas flow (reduces IBP)
What does the pressure relief valve prevent?
The buildup of pressure upstream of the outlet check valve
Where is the pressure relief valve?
Near the CGO and open to atmosphere to vent gas if preset pressure is exceeded
what does the pressure relief valve limit?
Limits the ability of machine to provide adequate pressure for jet ventilation
What regulates flow of oxygen, medical air and other gases in the anesthesia machine?
flow adjustment controls
How many flow adjustment controls are there?
Only one for each gas
Where must the flow adjustment controls be located?
Adjacent to its flowmeter and turn in only one direction
Describe oxygen flow knobs
Must be fluted and larger than other gases
What do flowmeters indicate?
Rate that gas is passing through piping into CGO then to patient
How is oxygen supply pressure to the flowmeters regulated
A constant, power pressure by a second stage regulator
What type of tube is used for flowmeters?
Thorpe tube
What must flowmeters be marked with?
Appropriate color and chemical symbol of gas
Where does the flowmeter sequence all gas flow from?
bottom to top and left to right
Standards of oxygen flowmeter?
Oxygen flowmeter to be placed on Right side, closest to CGO, that way if a leak occurs with other gases, it is unlikely to result in a hypoxic mixture
Where does the oxygen flush valve receive oxygen from?
Pipeline inlet or cylinder pressure regulator
Where does the oxygen flush valve send high flow O2?
To the CGO and the pressure could increase the supply pressure at the CGO, without the presence of pressure relief valves to appropriately regulate it
5 Requirements of oxygen flush valves?
- Operable with 1 hand
- Single purpose
- Self-closing
- Designed to minimize accidental use
- Have flow between 35-75 L/min
3 hazards of oxygen flush valves
- Potential for sticking of valve
- Barotrauma
- Anesthetic awareness
What does turning the master switch on for the oxygen failure protection device cause?
Causes oxygen to pressurize and holds open a pressure sensor shut-off valve
Describe the Oxygen Pressure Failure aka Oxygen Failure Protection Device (OFPD)
Shuts off or proportionally decreases nitrous to maintain a minimum 19% O2 flow at CGO using oxygen failure safety switch
When does the oxygen failure alarm sound?
When pressure falls below threshold, approx 30 psi, an alarm sounds within 5 seconds
What type of machines are hypoxia prevention devices required on?
Contemporary machines
2 factors of mandatory minimum oxygen flows
- Minimum of 50-250ml/min flow
2. Activated when master switch is on
What does the minimum oxygen ratio use/
Mechanical linkage with nitrous to limit nitrous flow when it is given in tandem with oxygen
When does the mechanical link engage with hypoxia prevention devices?
When oxygen concentration less than 25% to maintain a minimum oxygen concentration delivered and prevent a hypoxic gas mixture
Describe axillary oxygen flowmeters
Delivers oxygen in case of electronic power or system pressure failure, connect ambu bag or modified anesthesia circuit in order to ventilate the patient
What flow of O2 does the axillary oxygen flowmeter allow?
10L/min
When is the axillary oxygen flowmeter active?
when the master switch is off
When should the batter backup of anesthesia machines be at the highest level?
when they are plugged in
What does duration of the anesthesia machine backup depend on?
Power usage, manual ventilation uses much less power than ventilator usage
6 sources of anesthetic gas contamination
- APL valve
- High and intermediate pressure systems
- Low pressure systems
- Ventilator
- Anesthetic errors
- Cryosurgery
What is the outlet for anesthetic gases during spontaneous and assisted ventilation?
APL valve
How much FGF can exit through the APL valve?
5L/min
Describe the APL valve
Spring-loaded and only requires minimal positive pressure to open and allow the exit of waste gas from the circuit