Exam #2 Flashcards

1
Q

Mass Analyzers can use static/dynamic ________ or _______ fields to separate gas phase ions based on their ___________.

A

Electric ; magnetic ; m/z values

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Can mass analyzers be used alone or with other mass analyzers?

A

Both

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Ions are accelerated out of the source to the mass analyzer by ______________.

A

electrostatic focusing

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Why is it good that ions from the ion source usually have a negligible KE value?

A

Good because some mass analyzers measure the velocity and use it in their calculations. You don’t want to add to that.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

KE is

A

kinetic energy of fully accelerated ion.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

V is

A

Potential difference you’re accelerating ions with. You know what this value is and you can control it.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

With mass analyzers, ions move _______ to the _______.

A

from the ion source ; mass analyzer

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

In mass analyzers, how are the ions moved through the vacuum chamber?

A

Via electrostatic focusing

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

What are the two definitions of resolution?

A

1) The separation of two ions of nearly identical m/z (chromatography)
2) Sharpness of a mass spectral peak (of just one single peak)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

When talking about one peak’s sharpness (resolution), what is it talking about?

A

Full Width Half Maximum Method

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

What is the Full Width Half Maximum Method?

A

Resolution = m/Δm ; the m/z at the top of the peak divided by the width of the peak at half peak

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

What are wider peaks called?

A

Gaussian

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

What are sharper peaks called?

A

Lorentzian peaks

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

As a good rule of thumb, what m/z value is considered high resolution?

A

100,000 or more

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

High resolution mass spectrometers

A
  • Double Focusing Sectors
  • Orbitraps
  • FT-ICR (100,000,000 resolution)
  • Some TOFs (40,000 resolution right now but they are improving)
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Why do we need high resolution?

A

1) Detection of isobaric ions
2) Exact Mass Measurements
3) Determination of an ion’s charge from m/z using C13 (or C12) isotope pattern
4) Increase selectivity and/or sensitivity (incr. resolution decreases the chemical noise; lower number of ions detected but again, the noise is lowered). Increases S/N

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

Small width of peak means

A

High Resolution

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

What is the highest form of resolution?

A

FT-ICR

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

What is meant by “Double Focusing”

A

electric sector followed by a magnetic sector (EB or BE)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

What is the mass accuracy with high resolution?

A

+/- 0.0003 m/z

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

What calibration compound is used to calibrate the x-axis?

A

FC-43 (Perfluorotributylamine, aka PFTBA); Used for low volatile analysis (EI or CI)
-Calibrates the x-axis

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

What calibration compound is used to calibrate the y-axis?

A

Decafluorotriphenylphosphine (DFTPP)

-For calibrating ion intensity (y-axis)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
23
Q

Why are there so many hydrogens on the calibrants?

A
  • Presence of Fluorines significantly decreases boiling point (30-40ºC). More volatility b/c there is less hydrogen bonding
  • Because of its larger mass, it creates a larger range of analysis for your sample to peak
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
24
Q

Time of Flight (TOF)

A
  • Simplest mass analyzer (in terms of hardware and understanding it)
  • Measures time (t) required for ions to travel down a flight tube (L)
  • Not as sturdy (electronics get outdated over time)
  • Pulse Mass analyzer
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
25
Q

How do ions move out of the source / how are ions directed to the mass analyzer?

A

Ions are accelerated out of the source to the mass analyzer by electrostatic focusing.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
26
Q

Write out the kinetic energy equation for an ion fully accelerated from the ion source.

A

KE = (mv^2)/2 = zV

*Note: “z” is the charge on ions and “V” is the voltage in electrostatic focusing

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
27
Q

What is the mass spect. definition of resolution?

A

Resolution is the sharpness of the spectral peaks.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
28
Q

What is the chromatography definition of resolution?

A

The separation of two ions of nearly identical m/z

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
29
Q

Where are Lorentzian peaks typically found (what computational conversion)?

A

Fourier-Transform

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
30
Q

With mass analyzers, in addition to having high resolution, what parameter also needs to be well-defined?

A

m/z range

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
31
Q

Kicking out all chemical noise increases ______ and ______.

A

Resolution and specificity

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
32
Q

What chemical is used to calibrate the x-axis for mass analysis? What is it calibrating?

A

Perfluorotributylamine (FC-43) ; m/z range

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
33
Q

What chemical is used to calibrate the y-axis for mass analysis? What is it calibrating?

A

Decafluorotriphenylphosphine (DFTPP) ; ion intensity

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
34
Q

What is the mass range for the x-axis in mass analysis?

A

50 to 502 m/z

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
35
Q

Why does TOF require a high accelerating voltage?

A

-This is to minimize the KE spread that may occur from the ionization event. When ioniziation event occurs, KE is added to the ions. So you want to add such a large accelerating voltage (V) so that the KE added by the ionization event is negligible.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
36
Q

Why does fragmentation within the flight tube, along L, not effect the time of flight?

A

After the accelerating voltage is applied, it does not matter of there is fragmentation or not because the ions will still travel at the constant velocity. Velocity of the ions won’t change.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
37
Q

Name two ways to increase resolution by decreasing time spread.

A

1) Delayed Extraction - where you allow the ionization pulse to complete (200-350 seconds) before applying the pulse to extract the ions from the ion source.
2) Reflectron (or Reflector) - Ions are exposed to an electromagnetic field in the middle of the flight path. This fixes the time spread, not the velocity spread.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
38
Q

What is the purpose of reflectrons?

A
  • Reflectrons are used to fix the time spread.
  • Ions that are moving at a greater velocity have to travel a longer flight path, which means it takes longer to reach the detector. This is so that it and another ion with the same m/z (but may be moving at a slower velocity) will reach the detector at the same time.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
39
Q

What is TOF often coupled with (ionization technique(s))?

A

Any continuous source of ions. Also works really well with pulsed ionization methods.

  • MALDI, LD, & 252CfPD (Good because of their pulsed ionization method)
  • TOF/SIMS (more common with newer TOF machines)
  • ESI and APCI (ions are made continuously but can be “trapped” and pulsed into the extraction region as needed.
  • EI, CI, FAB, GC, LC-MS
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
40
Q

Reflectrons increase resolution by (2 methods)

A

1) Decrease the time spread for ions of the same m/z

2) Increase the flight path, (L)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
41
Q

What detector is used for TOF?

A

Multichannel Plate (MCP)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
42
Q

Describe the detector.

A

-A circular device that is about 2-3 mm wide in diameter. -Composed of multiple glass capillary tubes that each are 0.5-1.0 mm long and have a 5 micrometer diameter. The capillaries are coated with an electron emissive material such as PbO.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
43
Q

Why is the detector a good choice for TOF?

A
  • It is fast
  • Can go up or down as the ions hit the detector
  • Good multiplication of signal.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
44
Q

What is signal averaging?

A

Also known as ensemble averaging

  • Not related to Fourier Transform
  • Recorded mass spectrum is usually an average of spectra from 100 or more individual ionization events (i.e. laser shots)
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
45
Q

Signal/Noise ratio is equal to

A

The square root of the number of scans (n)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
46
Q

Overall features of TOF

A
  • Entire mass spectrum is obtained in 500 microseconds
  • Pulsed ; can be combined with MALDI, LD and PD
  • Can be used with continuous ion sources as well
  • Moderate to high resolution (better than quadrupoles, less better than double-focusing, much less better than FT-ICR)
  • Has the widest dynamic and linear range of any technique
  • The fastest mass analyzer
  • Can be used on large molecules (both volatile and nonvolatile)
  • Has very high efficiency of ion transmission, meaning it can detect extremely small amounts of ions
  • Very low limits of detection
  • Uses signal averaging
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
47
Q

What feature is needed for a TOF digitizer?

A

Needs a fast analog to digital converter

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
48
Q

With a magnetic sector, what two forces must be equal to each other in order for the turn/curve to take place?

A

Magnetic force and the centrifugal force

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
49
Q

What is the equation for magnetic force?

A

Bzv

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
50
Q

What is the equation for centrifugal force?

A

mv^2/r

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
51
Q

How do magnetic sectors discriminate m/z?

A

Using momentum to charge ratio

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
52
Q

Derive the BS equation for m/z.

A

mv/z=Br

m/z=B2z2/2V

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
53
Q

What is the special focusing feature of BS?

A

Directional Focusing-

ions with the same m/z and velocity but take different paths through B will come into focus at the exit slit.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
54
Q

Name two parameters which affect ions reaching the exit slit.

A

1) Scanning V while B is held constant (needs continuous ion source)
2) Scanning B while V is held constant
- Most common
- Scanning analyzer
- Yields a non-linear mass scan

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
55
Q

How does a magnetic sector work?

A
  • At given values of B & V, only ions with a particular m/z have a circular path of radius, r, that allows them to pass through the exit slit of the analyzer.
  • Ions of other m/z will hit the walls of the analyzer
56
Q

Magnetic Sectors require a __________ source of ions.

A

Constant

57
Q

With directional focusing, the time required to obtain a complete spectrum depends on _________ and _______.

A

the scan rate ; m/z range

58
Q

The slower the scan rate the ___________ measurement of B and more accurate measurement of ___________.

A

more accurate ; m/z measurement

59
Q

The accelerating voltage (V) range of the magnetic sector is __________ with ______ being the more common.

A

1-10 kV ; 3 kV

60
Q

What are the common sector angles for magnetic analyzers?

A

60, 90, and 180 degrees

61
Q

With magnetic analyzers, lowering the accelerating voltage will _______ the mass range.

A

increase

62
Q

How is the accelerating voltage of the magnetic analyzer related to the spread in the ion beam?

A
  • The greater the accelerating voltage causes the initial velocity of the ions (from the ionization event) to minimalized relative to their velocity when accelerated by V.
  • This gives more monoenergetic ions
63
Q

How are slit width, resolution, and ion abundance related / how do they effect each other?

A

Narrowing the entrance and/or exit slit width INCR. resolution but also decreases ion abundance

64
Q

What ion sources can be coupled with BS?

A
  • Literally anything that has a continuous source of ions

- EI, SIMS and CI

65
Q

What mass analyzers are considered continuous/scanning?

A

E, B and Quadrupoles

66
Q

What mass analyzers are considered pulsed?

A

TOF, TOF relfectron, FTICR, QIT

67
Q

Rank the mass analyzers from least resolution to highest resolution

A

Q, QIT, TOF, TOF R, EB, FTICR

68
Q

What ES forces must be perfectly balanced for ions to make the turn/curve?

A

Electrostatic and centrifugal forces

69
Q

What is a special focusing feature of ES?

A
  • It filters ions based on energy

- For ions with the same mass, ESA also provides velocity focusing

70
Q

Why does ESA have sharper spectra peaks?

A

Because it has velocity focusing and energy focusing

71
Q

How does ESA scan a spectra?

A
  • Scans a mass spectrum by scanning the accelerating voltage (V) and scanning the electric field strength (by scanning V prime)
  • This brings the ions of varying KE through the exit slit
  • Only ions with a particular KE will pass through the circular path of radius, r, and emerge from the exit slit
  • It filters ions based on energy. Ions w/ different KEs emerging on different paths.
72
Q

What order of sectors is considered normal? What order is considered reverse?

A

EB is normal, BE is reversed

73
Q

Pros of EB

A
  • Gives enhanced energy and direction focusing to ion beam
  • Increases resolution (up to 10^5 at 1000 m/z)
  • Better detection limits b/c of more intense peak
  • Better sensitivity
  • Can accept ions with a higher range of velocities
  • Sharpens the range of velocities
74
Q

How do the peaks of EB differ from peaks made using B alone?

A

They are sharper, more narrow peaks. Specifically, it sharpens the velocity range.

75
Q

E sector does what

A

separates ions based on velocity (or KE)

76
Q

B sector does what

A

separates ions based on momentum

77
Q

What is normal geometry?

A

EB ; Used for high resolution. Magnetic sector acts as the actual mass analyzer. The E sector sharpens the peaks by giving energy focusing (velocity focusing). The E sector has a slit to allow ions to enter the magnetic sector.

  • Gives a more accurate mass measurement
  • The magnetic sector focuses the ions to a point. This is where the detector needs to be placed.
78
Q

What is Mattauch-Herzog geometry?

A
  • All values of m/z are brought into focus along a plane
  • B is not scanned
  • Instead, a multichannel detector is used.
  • Simultaneously detects all ions. All m/z are detected at once
  • Uses a curved electromagnetic field
79
Q

What is the significance of SIMS?

A
  • Commonly used with scanning mass analyzers to improve the limit of detection or increase selectivity.
  • Only a few m/z of interest are allowed through the analyzer at a time, all others are kicked out.
  • Commonly used for Quant. Analysis
80
Q

Describe the shape of the quadrupole (Q) rods and how their shape affects their use.

A
  • Circular or half-moon shaped rods that serve as mass to charge filters.
  • Only the curved part of the rods is where ions hit.
81
Q

Describe the potentials applied to each/all rods.

A

Combinations of dc (direct current) and radio frequency (rf) alternating current (ac) voltages are applied to the rods.
-Adjacent rods have opposite polarities

82
Q

What is the typical length and diameter of a Quadrupole rod?

A

20 cm and 1 cm

83
Q

What is the radio frequency equation?

A

RF = Vcos wt

84
Q

What is unique about the quadrupole?

A
  • Has the lowest accelerating voltage of all mass analyzers (0-30 V)
  • Only continuous ion sources
  • Scans the electric field
  • Scan one m/z at a time
85
Q

Why are Q called mass filters instead of mass analyzers?

A
  • Because only one m/z is being scanned at a time

- Only ions of a specific m/z value are able to travel through the rods and to the detector.

86
Q

How does the position of the scan line on the Mathieu diagram affect detection limits, sensitivity, ion transmission, and resolution?

A

-Everything in the stability region is stable enough to hit the detector. Anything outside of that is too unstable and will hit the rods.
To have stable trajectories U < V/6

87
Q

Increasing the slope of the scan line on the Mathieu diagram

A

Incr. resolution, becomes more selective for that one m/z, and lowers sensitivity and detection.

88
Q

Decreasing the slope of the scan line on the Mathieu diagram

A

Incr. Transmission efficiency (more ions can come in), better sensitivity and detection, but lowers the resolution

89
Q

What is meant by “dynamic” mass analyzers?

A

Mass analyzers that are dependent on time-varying electric fields or on time-dependent ion-movement.
Ex: TOF, Q, ICR, Orbitrap
-Anything with frequency

90
Q

What is meant by “static” mass analyzers?

A

Mass analyzers that are dependent on the effects of electrical or magnetic fields and NOT on TIME.
Ex: Sectors

91
Q

Why would someone have to increase the voltage applied to a quadrupole over time

A

Over time, the Q rods will become dirty due to ions colliding the rods. More voltage will be needed to get through this grime so that the ions traveling will feel it.

92
Q

Compare and contrast a regular linear quadrupole and a QIT

A
  • QIT is a 3D quadrupole device that has 2 dome-shaped end cap electrodes and a central ring electrode.
  • It most commonly uses “Mass Selective Instability Mode” (aka rf-only mode).
  • AC Voltage in radio frequency range is applied to the ring electrode.
  • In MS mode, the end cap electrodes are held at ground potential.
  • Voltages are applied to the end cap electrodes to kick out ions from the trap.
  • Linear quadrupole just has rods
93
Q

How are the ions trapped in a QIT

A

-At a fixed value of V, all ions above a certain m/z are trapped within the quadrupole field. Each m/z has a specific oscillating frequency (secular frequency)
At a fixed frequency, the storage voltage determines the range of m/z that are trapped by the field.

94
Q

Describe the motion of the ions once trapped

A

They oscillate at a frequency related to their m/z and this motion is imperfect. As the RF voltage is raised ions of increasing m/z expand their range of motion and escape from the trap and are detected via an EM

95
Q

How does trap size correlate to resolution

A

Smaller trap size = better resolution

96
Q

What is the purpose of collisional cooling

A

Used to collide with the oscillating ions to keep them in the center of the trap. This is so they can lose some of their KE and won’t leave the trap until you’re ready

97
Q

Where is the filament placed in QIT for EI CI versus ESI APCI and FAB

A

a. The filament is beside the trap and ions are pulsed in at the electron gate which involves accelerating voltages for EI CI
b. Ions are produced externally and injected as pulses into the trap by electrostatic focusing for ESI APCI and FAB

98
Q

What is the upper m/z range for QIT?

A

650 m/z

99
Q

In QIT, how long can ions be stored for?

A

milliseconds to seconds

100
Q

What is the sample pressure in QIT?

A

10^-6 to 10^-7 Torr

101
Q

Why is the dynamic range of QIT so small (10^3)?

A

Because only a small number of ions can be trapped. If you increase the trap size, the radius between the rods gets bigger = you lose resolution and performance; Too many ions cause charge effects and causes ions to fly out of the trap

102
Q

Can QIT hold positive and negative ions at the same time

A

Yes

103
Q

Linear Ion Trap

A
  • 2D form of ion trapping
  • Rod-like trap
  • Same charge is applied to both ends of the rods to trap the ions in the middle.
  • Not good for trapping positive and negative ions at the same time.
104
Q

Advantages of LIT over QIT

A

1) Bigger than QIT. Can hold more ions, better dynamic range b/c it can hold more. More sensitive. More ions = better detection limits
2) Improved trapping efficiency
3) Improved ion-ejection
4) Faster scan speeds
5) improved detector sensitivity

105
Q

How is a mass spectra obtained using HCT

A

Detecting the number of resonantly ejected ions as a function of applied RF voltage amplitude

106
Q

In tandem MS^n what are two things resonance excitation can do

A

1) Cause unwanted ions to be ejected out of the trap to isolate the precursor ion
2) Provide precursor ions with KE for collision induced dissociation

107
Q

What is the upper range of HCT?

A

6000 m/z

108
Q

What three things of HCT that contribute to it having better detection limits, speed resolving power, and mass accuracy than QIT?

A

1) geometry of the device
2) rf voltage application
3) ejection methods are optimized

109
Q

What mass analyzers can use EM as a detector

A

Quads and sectors

110
Q

What magnitude (unit) describes the response time of an EM

A

nanoseconds

111
Q

What is the purpose of the dynode

A

Converts the ion beam to an electron beam

112
Q

How many stages does a discrete stage dynode have

A

14-16

113
Q

What is the difference between a continuous and discrete dynode

A
  • With discrete dynode, there are multiple stages where a different voltage is applied.
  • With continuous dynode, it is a horn-shaped that has the same voltage applied to it but the ions reach the detector via a series of “bumps” down the funnel of the horn.
114
Q

Are there any advantages of a continuous dynode over a discrete dynode

A

The continuous dynode is smaller and requires less power to operate

115
Q

What is the material used to coat photomultiplier detector versus a dynode

A

-Organic phosphor

With a dynode, there is only copper beryllium

116
Q

With the Daly photomultiplier, what is the ion path?

A
  • The ions are accelerated with a 20-40 kV into a conversion dynode, which emits electrons.
  • The electrons hit an organic phosphor which then produces photons
  • The photons pass through a quartz or sapphire window and into a photomultiplier tube.
  • Inside the PMT, the photon beam is converted back to electrons. the electron beam is multiplied, and electrical current is detected
117
Q

What is the most common MS detector?

A

Daly Photomultiplier Detector

118
Q

This it the worst ion current detector

A

Faraday Cup ; good only really for isotopic ratios

  • Simple detector in which the ion beam impinges against a metal collector and the electrical current produced by the ion beam is measured.
  • No amplification
119
Q

Discharge sources and plasma sources may still use ______ as their detector

A

Photographic Plates

120
Q

How does FTICR separate ions

A

It separates ions based on the cyclotron frequencies at which they travel in the presence of a strong magnetic field. Frequency of circular motion = m/z

121
Q

The cube/cylinder of FT-ICR is made of how many plates?

A

6

122
Q
  1. Ions exposed to a strong magnetic field are constrained in circular orbits _______ to the magnetic field
A

perpendicular

123
Q

What is the equation for angular frequency?

A

v/r

124
Q

Ions have trapping times of

A

milliseconds to hours

125
Q

10’s of ___________ are produced per second with FT-ICR.

A

spectra

126
Q

Name and explain three fundamental modes of motion

A

1) Cyclotron motion – rotation of an ion about a fixed applied magnetic field. Used in m/z analysis
2) Trapping motion – axial back and forth motion of ions between two plates which have a DC potential. Used to trap ions in cell
3) Magnetron motion – slow circular drift of an ion along a path of constant electrostatic potential. Occurs as a result of the cross radial electric field and axial magnetic field

127
Q

How is an ICR spectrum obtained

A

1) Scan magnetic field, keep frequency fixed
* m/z sequentially come into resonance, their image currents are detected (conventional ICR)
2) Scan excitation frequency and keep magnetic field constant
* m/z sequentially come into resonance, their image currents are detected (rapid san ICR) as frequency can be scanned more rapidly than magnetic field
3) Excite all m/z’s simultaneously with a very fast frequency sweep, keep B constant
* The image currents of all m/z’s are detected simultaneously in time and FT converts time data to frequency data (FTICR)

128
Q

Considerations of TOF

A
  • The fastest mass analyzer (Good choice to couple with chromatography)
  • Unlimited mass range of the mass analyzer itself
  • Pulsed (most common with pulsed ionization)
  • Resolution is at the low end of the high resolution range (but improving)
  • Moderate to high resolution
  • Analyzer is relatively inexpensive
  • Good mass analyzer for ultrafast chromatography
  • Commonly couple with GC-MS and LC-MS
  • Most TOFs are MALDI-TOFs
  • Single Ionization event gives a complete mass spectrum, so very small samples can be used.
  • Very low LOD
  • Requires good electronics
  • Widely commercially available
  • Used primarily with MALDI in the analysis of large molecules such as biomolecules, organic polymers, and inorganics) or with chromatography (GC-MS or LC-MS)
129
Q

Microchannel Plates (MCP)

A
  • Detector
  • Based on fiber optics
  • Measuring the electrical current
  • Very fast response times
  • Coupled with TOF
  • Entire mas spectrum may be obtained in 500 microseconds
  • Very sensitive
  • Requires a high accelerating voltage.
  • HIGH SOURCE PRESSURES (LIKE LC-MS) MAY LEAD TO ARCING. ADDITIONAL PUMPING MAY BE REQUIRED
  • Less rugged and more expensive than quadrupoles, but more versatile because of higher resolution
  • WORKS WELL W/ SIMS
  • HIGHEST RESOLUTION OF ALL
130
Q

Magnetic Sectors’ ion source is

A

EI or CI

131
Q

__________ will not work with MALDI because it needs a continuous source of ions.

A

Magnetic Sectors

132
Q

________ are slower than other machines but the slower the scan rate gives a __________ measurement of B and m/z measurement

A

Sectors ; more accurate

133
Q

Quadrupoles

A
  • Simplest MS in operation
  • Any continuous ion source ( DART, DESI, ESI)
  • CAN’T USE W/ MALDI
  • SCANNING MASS ANALYZER
  • Requires DC and AC voltages
  • Has the lowest accelerating voltage of all mass analyzers
  • Low resolution
  • m/z range up to 4000
  • Precise rod alinement is key
  • GOOD FOR EI, not MALDI
  • WORKS WELL WITH SIM
  • MORE RUGGED AND LESS EXPENSIVE THAN SECTORS, BUT LESS VERSATILE
  • RODS ARE DIFFICULT TO ALIGN AND COSTLY TO REALIGN
  • HIGH MAINTENANCE
  • CONSTANT CLEANING
  • READILY COMPATIBLE WITH “HIGH PRESSURE” TECHNIQUES (LC-MS) B/C IT DOESN’T CAUSE ELECTRICAL ARCING
  • M/Z ANALYSIS IS NOT DEPENDENT ON VELOCITY, SO A LARGE RANGE OF KEs ARE ACCEPTED
134
Q

Quadrupole Ion Trap (QIT)

A
  • Best-selling Mass spectrometer of all time
  • Pulsed
  • Storage voltage determines the range of m/z that are trapped by the field
  • Pressures are 10^-6 to 10^-7 Torr
  • Very simple to operate
  • Relatively inexpensive, versatile, rugged, small, and mobile
  • Not normally high resolution
  • Can achieve m/z ranges up to 5,000
  • Ions trapped by high pressure helium
  • Ions are detected by an electron multiplier (EM) detector
  • Common with GC-MS
  • Can trap positive and negative ions at the same time
135
Q

HCT ion trap

A
  • 10x greater ion capacity
  • LOD is 10x better
  • Most sensitive ion traps available
  • Better resolution
  • Upper m/z range of 6,000
  • Fast, can scan up to 26,000 m/z per second
  • Medium resolution
136
Q

MALDI is a ______________ ion source.

A

pulsed

137
Q

QIT is not usually coupled with MALDI because ________________.

A

QIT’s low upper m/z range isn’t capable with MALDI’s unlimited m/z range.