Exam 1- Clinical Monitoring (6/12/23) Flashcards
What are the two sampling sites depicted by the two arrows?
- Elbow
- Y-piece
What are the two types of gas sampling systems?
- Side-stream/ diverting analyzer
- Mainstream/ non-diverting analyzer
Which gas sampling system will have more lag time (transit time)?
- Side-stream/ diverting analyzer
What is rise time in terms of the gas sampling system?
- The time taken by the analyzer to react to the change in gas concentration
The mainstream analyzer will have a faster rise time.
Side-stream responses is dependent on what factors?
- Sampling tubing inner diameter
- Length of tubing
- Gas sampling rate (50 - 250 mL/min)
What are gas sampling challenges with mainstream analyzers?
- Water vapor (can block IR waveforms)
- Secretions
- Blood
- More interfaces for disconnections
What are gas sampling challenges with side-stream analyzers?
- Kinking of sampling tubing (can’t break over time)
- Water vapor (can block IR waveforms)
- Failure of sampling pump
- Leaks in the line
- Slow response time
The total pressure exerted by a mixture of gases is equal to the sum of the partial pressures exerted by each gas in the mixture. What law is this?
- Dalton’s Law
At sea level, what is the total pressure of all anesthetic gases in the system?
- 760 mmHg
Calculate the partial pressure of O2 at room air
- 159.6 mmHg
760 mmHg x 21% = 159.6 mmHg
Calculate the partial pressure of inspired O2 at room air.
- 149.7 mmHg
PIO2 = FIO2 (PB -PH2O)
21% (760 - 47) = 149.7 mmHg
_________ is an instrument that allows the identification and quantification, on a breath-by-breath basis, of up to eight of the gases commonly encountered during administering an inhalational anesthetic.
- Mass Spectrometry
This tool uses a high-powered argon laser to produce photons that collide with gas molecules in a gas sample. The scattered photons are measured in a spectrum that identifies each gas and concentration.
- Raman Spectrometry (Raman Scattering)
No longer in use
What is typically used in anesthesia machines to determine the concentration of gas?
- IR Analysis
What is the most common gas analyzer?
- Non-dispersive IR analyzer
IR analysis measures energy absorbed from a narrow band of wavelengths of IR radiation as it passes through a gas sample
What gases are measured using a non-dispersive IR analyzer?
- CO2
- Nitrous Oxide
- Water
- Volatile Anesthetic Gases
O2 does not absorb IR radiation
How does Infrared Analysis (IR Analyzer) work?
- Gas will enter the sample chamber
- Each gas has a unique IR transmission spectrum absorption band
- Strong absorption of IR light occurs at specific wavelengths
- IR light is transmitted through the gas sample and filtered
- The amount of IR light that reaches the detector is inversely related to the concentration of the gas being measured
- Less light = high concentration of gas
Do side-stream analyzers take into account of water vapors?
- No
- Side-stream analyzers report ambient temperature and pressure dry values (ATPD).
What are the two types of oxygen analyzers?
- Fuel or Galvanic Cell O2 Analyzer
- Paramagnetic O2 Analyzer
What are the drawbacks of a Fuel/ Galvanic Cell O2 Analyzer?
- Short life span (months) depending on the length of O2 exposure
- Slow response time (best to measure O2 in the inspiratory limb)
What oxygen analyzer is used in most side-stream sampling multi-gas analyzers?
What is the benefit of this analyzer?
- Paramagnetic O2 Analyzer
- Benefit: Rapid response, breath-by-breath monitoring
Purpose of gas sampling inside the inspiratory limb.
- Ensures oxygen delivery
- Analyzes hypoxic mixtures
Purpose of gas sampling inside the expiratory limb.
- Ensure complete pre-oxygenation/ “denitrogenation”
- ET O2 above 90% adequate
What can trigger a low O2 alarm?
- Pipeline crossover
- Incorrectly filled tanks
- Failure of a proportioning system