Equipment + Measurement Flashcards
Gas
Exists in the gaseous state at room temp. Its liquefaction at room temperature is impossible, since the room temperature is above its critical temperature.
Vapour
Gaseous state of a substance below its critical temperature. At room temperature and atmospheric pressure, the substance is a liquid
Critical temperature
Temperature above which a substance cannot be liquefied no matter how much pressure is applied
Pulse Oximetry
- Probe with 2 LEDs (red 660nm, infrared 940nm) and a photodetector
- A microprocessor analyses the signal (calculates R:IR absorbance ratio = R value)
- 3 phase cycle >200Hz
=> Red on, IR off
=> Red off, IR on
=> Both off (baseline - so interference from ambient light can be detected and subtracted)
Isobestic point = 590nm + 805nm (wavelength at which oxy- and deoxyHb absorb the same proportion of the two wavelengths of light)
Pros
- Safe, non invasive
- Accurate within clinical range (70-100%)
- Fast response time, cont reading
- Changes in audio tone
- Plethysmography trace gives index of peripheral perfusion
- Accuracy not effected by HbF or HbS
Cons
- Susceptible to errors -> inaccurate reading
- Potential for burns if fault
Sources of error
- Low signal to noise ratio (pulsatile component only 1-2% of total signal, hence reading prone to inaccuracies is any further reduction is signal)
- Arrhythmias
- Different Hb (metHb - absorbs red and IR light 1:1 => falsely low readings. COHb => falsely high)
- Electromagnetic interference (e.g diathermy)
- Coloured dyes and nail polish (falsely low)
- Sats <75% (reading extrapolated)
- Lag in reading change (up to 30s)
- Venous pulsations (may be interpreted as arterial -> falsely low)
Humidifiers
Passive - HMEs
Active - THRIVE
HME
- Pleats of paper provide large SA which condensation/ vaporisation can occur
- Hygrosopic (water retaining)
- Hydrophobic (water repelling), or combination of both
MOA
- Expire warm, fully saturated air
- Gas passes through HME, loses heat to cooler ambient temps => condensation + secondary hygroscopic + hydrophobic properties
- Condensed water evaporated and returned to patient on next inspiration
- Relative humidity of 60-70% achieved
Pros
- Cheap
- Lightweight
- Portable
- May contain anti-bacterial filter
Disadvantages
- Incr resistance (require 5-10cmH2O PS to compensate)
- Incr apparatus deadspace
- Obstruction with mucous + water accumulation => incr resistance
- Not as effective cf active humidifer
- Efficiency decr with large VT
- Only lasts 24hrs
Humidity and intubation
Absolute humidity - mass of water vapour present in a given volume of air (mg/L or g/m3)
Relative humidity - the ratio of the mass of water vapour present in a given volume of air to the mass of water vapour required to saturate that volume at that temp (%)
Intubation bypasses humidification function of nose
- Dehydration of mucosa
- Altered ciliary function
- Inspissation of secretions -> obstruction, atelectasis, V/Q mismatch
- Loss of body heat, esp neonates
DINAMAP
Components
- Cuff with tube
- Case with microprocessor, pressure transducer and solenoid valve
Arterial Pressure waveform
- Primary/ fundamental frequency = HR (60-120bpm = 1-2Hz)
- Dicrotic notch represents change in pressure due to closure of AV
- Rate of rise of upstroke reflect myocardial contractility. Slow rise requires inotropic support. Max upward slope related to speed of ventricular ejection
- Position of dicrotic notch - PVR. In vasodilated - lower on curve. Vasoconstrictor - high on curve
- Downstroke slope - resistance to outflow. Slow fall seen in vasoconstriction
- SV can be estimated by measuring area from start of upstroke to dicrotic notch, then x HR = CO
- Systolic time = myocardial O2 supply
- Diastolic time = myocardial O2 demand
- MAP = average pressure throughout cardiac cycle
CO measuring - non-invasive
Calibrated
- PiCCO - uses CVL and thermistor tipped art line, CO estimated by pressure waveform analysis. Calibrated using trans pulmonary thermodilution method
- LiDCO - standard art line, calibrated by lithium dye injected into peripheral (or central vein), fall in conc measured by lithium electrode sampling art line. Need recal Q8hrly.
=> avoids error of pulm heat dissipation
=> cannot be used in pt on lithium
=> freq calibration introduces error
=> muscle relaxants may cross react with lithium
=> underestimated CO when afterload is low
=> cardiac shunts introduce error in both LiDCO and PiCCO
Non-calibrated
- FloTrac/ Vigileo - pressure sensor attached to art line, waveform analysis
- LiDCO rapid - same algorithm as LiDCO but uses nomogram for calibration
Sources of error
- Under or over damped art line
- Used of IABP
- Arrhythmias
- Aortic regurgitation
- SV variation
CO measurement - invasive
Pulmonary artery catheter
- Inserted into IJ or subclavian vein, proximal lumen connected to transducer with continuous pressure monitoring and inserted into RA/ RV, then balloon inflated and floats until it is in the PA/ branch
- RA 3-8mmHg, RV 25/0, PA 25/10, PACWP 4-12
Indicator dilution methods
- Uses Fick principle - based on law of conservation of mass. Uptake or excretion of a substance is equal to the difference between the amount of substance entering and the amount leaving the organ
Thermodilution technique
- Requires PAC with thermistor
- Heat lost from blood = heat gained by injectate
- 10-15ml of normal saline at room temp with rapid injection
- Change in PA temp measured and temp-time graph plotted
- CO calculated using mod Stewart-Hamilton equation (large AUC = small CO)
- Gold standard
- Errors - speed of injection too slow (do average of 3), low CO (injectate may increase preload briefly and thus CO, overestimates CO), inaccurate temp recording of injectate, false assumption (thermal equilibrium established by the time mixture reaches thermistor, injected warmed by blood only, flow unidirectional, cold injectate does not depress CO)
Dye dilution technique
- Known amount of indicator dye injected into PAC and conc continuously samples at peripheral art line
- Indocyanin green (short T1/2 and low toxicity), Li also used
- Change in conc over time graph
- Calculated from AUC
- Error introduced from recirculation of the indicator dye which causes a second peak on the graph
USS oesophageal doppler
- Minimally invasive
- Small USS transducer on tip of probed inserted via mouth or nose to oesophagus to T5-6 (descending aorta adjacent and parallel to oesophagus)
- USS beam directed at 45 degrees to direction of aortic blood flow
- Doppler equation -> velocity of flow in aorta calculated, + CO, SV, volaemic status, SVR and myocardial function
- Pros - minimally invasive, nil CVL or art line. No calibration. Continual CO measurement
- Cons - movement of probe may lead to poor trace, frequent probs repositioning required. Estimates may be inaccurate
Critical temp of O2 and N2O
O2 = -115C (at room temp = gas)
N2O = 36.5C (at room temp = gas + liquid)
TOF
- 4x 0.1ms stimuli 0.5s apart (2Hz)
- TOF ratio = ratio of the 4th twitch height to the 1st twitch height
=> TOF >0.9 = adequate level of reversal for extubation - Depolarising blockade - reduced the amplitude of all 4 contractions with NO fade
- Non-depolarising blockade - causes fade, reduction of twitch height with each subsequent stimuli
- TOF fade is difficult to evaluate manually or visually when the ratio is >0.4
- TOF count correlates with receptor occupancy
- 2 twitches present (~80% receptors occupied) = anatgonisable block
Tetanic stimulation
- 50Hz stimulation for 5s -> sustained muscle contraction
- Non-depolarising blockade
=> Tetanic stimulation fades with time
=>Blockade of pre-synaptic AchR by NDMR -> no positive feedback -> decr production of Ach vesicles -> decr mobilisation of Ach at time of peak activity - Effect may last up to 6mins -> underestimation of neuromuscular block
- Painful in awake
Post-tetanic potentiation and count
Post-tetanic potentiation - incr response to a given stimulus following tetanic stimulation. Due to incr synthesis and mobilisation of ACh. Additional ACh is able to compete with NDMRs for receptors.
PTC
- Single twitch at 1Hz are started 3s after tetanic stimulation for 10s
- Number of twitches is inversely related to the level of block
- Useful when profound neuromuscular blockade is required
- 1st twitch of TOF returns with PTC of 9
- PTC should not be repeated within 6mins, as result inaccurate
Double burst stimulation
- 2x short burst 50Hz tetanic stimulation 750ms apart
- each short burst consists of 3x0.2ms stimuli separated by 20ms
- Fade is easier to detect manually
=> Bigger response
=> Two twitches are easier to assess than four - Same requirements as TOF
=> Ratio >0.9 required for adequate reversal - can detect fade 60% difference cf 40 in TOF