Electrodes Flashcards

1
Q

Describe the electrode evolution.

A

Spiral Electrode (1990-1997)
Spiral Electrode with Positioning System (1998-1999/2000)
HiFocus Electrode with Positioning (1999-2000)
HiFocus Electrode (present)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

How do previous electrode designs differ?

A

Wiring of electrode contacts
Location and shape of the contacts
Curvature of the elctrode carrier.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

How are electrodes stimulated?

A

Stimulated in pairs with one active and one ground electrode.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

How can electrodes be coupled?

A

Monopolar (what AB uses)-stimulating/active electrode is intracochlear and the ground electrode is extracochlear
**Allows the ability to recruit more nerve fibers to obtain adequate loudness levels with less current.
Bipolar-both stimulating/active electrode and ground electrode are intracochlear. This can allow more accurate stimulation and recruits fewer nerve fibers, which requres more current to obtain adequate loudness levels. *Power hungry

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

What was the most recently released HiFocus Electrode?

A

Mid Scala

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

What is unique about the HiFocus MidScala Electrode?

A

Designed for MidScala placement
Designed to be inserted with minimal trauma to the delicate structures of the cochlea
Smaller Pre-curved electrode
Designed for hearing preservation…but we can’t say that per FDA

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

In what year did AB initiate a voluntary recall after observing a small number of failed devices due to moisture?

A

September 2004

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

What did AB do after voluntary recall 2004 to improve manufacturing process?

A

Bake test and helium test

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Describe the Vendor B feedthru issue.

A

In 2006, after manufacturing processes had been improved, a high number of ex planted devices showed moisture ingress. After investigation vendor b was found to be the source of a bad feedthru. The glass insulator used was too porous and over time found to corrode and leak body fluid in to the casing.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

What year did AB stop using Vendor b?

A
  1. All vendor b devices were recalled and pulled off the shelves.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

What is the 7 year CSR?

A

98.31

How well did you know this?
1
Not at all
2
3
4
5
Perfectly