E2 Flashcards
How are compressed gases measured?
Psi = pounds per square inch
Psig = pounds per square inch gauge
What is the relationship of non-liquefied vs liquefied compressed gases.
Relationship btwn pressure and remaining volume and pressure reading on the gauge
What are non-liquefied compressed gases and their properties?
Gases that don’t liquefy @ room temp regardless of pressure applied
O2
Nitrogen
MEDICAL Air
Helium
b/c boiling point is well below ambient temp
Properties:
Will become liquids at very low temp
volume and Pressure
Non-liquefied gas = pressure ↓ as volume ↓
Describe the relationship between pressure and volume of non-liquefied compressed gases.
What law is this?
THERE IS A LINEAR RELATIONSHIP btwn PRESSURE & VOLUME FOR THESE GASES
SO P1/V1=P2/V2
Boyles LAW
What is liquefied compressed gas, examples and it’s properties?
Gas that becomes liquid @ ambient temp and at pressures from 25-1500 psi Liquid at Room temp & Patm N2O CO2 (insufflation)
How do non-liquefied gases differ from liquefied gases?
Liquefied gases do not follow boyles law
They are liquid at ambient room temp and Patm
How is pressure maintained in liquefied compressed gas?
WHEN ONE GAS MOLECULE LEAVES TO BE USED then ANOTHER MOLECULE FROM THE LIQUID TAKES ITS PLACE
WILL MAINTAIN PRESSURE UNTIL BASICALLY EMPTY
HAS TO BE OVER 95% EMPTY BEFORE PRESSURE CHANGE ON GAUGE
Liquid pressures = gas pressure
When will a pressure change be noted on liquefied compressed gas gauge?
HAS TO BE OVER 95% EMPTY BEFORE PRESSURE CHANGE ON GAUGE
How is the volume of liquid compressed gas measured?
Weight
Government regulations of cylinders. FDA, OSHA, DoT
FDA = gas purity
Dept of Labor/OSHA = employee safety
DoT = marking, labeling, storing, maintenance, transportation, and disposition
What are specific DoT gas regulations? (8)
• -Inspected & tested ONCE every 10 years
• -Test date stamped on cylinder
• -Must pass visual inspection & pressure testing
• -Color coded in the US (green)
• but should not be the primary means to identify a gas
• -Diamond shaped label
• identifying fire danger
oxidizer, non-flammable, or flammable
• -Signal word identifying hazard level = CAUTION, WARNING, DANGER
• -Name & address manufacturer & Date of expiration
• -Tag for Full, In USE, Empty to notate gas level
Describe the components of medical gas cylinder? (5)
Body Steel or steel carbon fiber = 3AA Aluminum = 3AL or 3ALM need in MRI Flat or concave base Neck with screw threads
Valve
Bronze or brass screws into neck
Allows refilling and discharge of gas at stem
Port
Point of exit for gas
Take care not to screw retaining screw into port= damage
Conical depression
Fits the retaining screw on the yoke
Handle Opens/closes cylinder Turns counterclockwise** to open Also called cylinder** wrench Must have one for every machine to be readily use
Describe the purpose of the pressure relief device on the cylinder
Vents cylinder to atm if pressure within cylinder becomes too high
• PREVENTS EXPLOSION FROM EXCESSIVE PRESSURE
What are pressure relief device types
Rupture or FRANGIBLE disk
Fusible plug
SAFETY relief
Describe the rupture/frangible disk pressure relief device on the cylinder
- good for venting in high temp or overfilling
- BUT BREAKS AT HIGHER PRESSURES ALLOWING GAS TO ESCAPE
- non reclosing
Describe the fusible plug pressure relief device
- WILL MELT w/ HIGH TEMPS AND ALLOW THE ESCAPE OF GAS
* non reclosing
Describe the safety relief valve in a cylinder
• most common
• SPRING-LOADED MECHANISM TO ALLOW VENTING OF GAS
• IF PRESSURE ↑ IT ALLOWS GAS VENTING
THEN recloses/SEALS after pressure normalized inside cylinder
What is the most common type of pressure relief device on a cylinder?
- SAFETY relief valve
- most common
- SPRING-LOADED MECHANISM TO ALLOW VENTING OF GAS (normalize to Patm)
- IF PRESSURE ↑ IT ALLOWS GAS VENTING
What are the gas cylinder sizes
A = smallest
E = most common on gas machines and for pt transport
Volume and pressure will vary in any given cylinder
Pressure, volume and color of O2, air and N2O full tanks
Full O2 cylinder (GREEN) = 660 L at 1900 psi (some books 625 L @1900 psi)
Full Air cylinder (YELLOW) = 625 L at 1900 psi
Full Nitrous oxide (BLUE) = 1590 L at 745 psi
Cylinder safety considerations (8)
Valves, regulators, gauges do not come into contact with oil, grease, or lubricants
Temperature regulation
< 130F (54C) & > 20F (-7C))
Keep connections Tight
No adapters should be used to change the size of connections for use of w/ hoses, regulators, or gauges
No alteration of markings and labels
No dropping, dragging, sliding of cylinders
Valve kept closed at all times
Cylinders should always be Properly secured to prevent fall
Cylinder storage consideration (8)
Storage in Designated secure areas (NOT the OR)
Adequate ventilation
Signage= no smoking, no combustibles in area of cylinders
Not exposed to corrosive chemicals/fumes
N2O secured/locked up to prevent access and abuse
Stored upright in bins or chained to wall
Wrapping and drapes undesirable
Recent jcaho guidelines
• require that empty or partially empty tagged cylinders
• stored separately from partially full and full cylinders w/ proper gauges
Considerations for the use of cylinders prior to using it(5)
Visible inspection for defects of PISS system, label, regulator
Presence of tamper proof seal around valve (NEW)
must remove prior to attaching to anesthesia machine
Presence of a sealing washer
if absent, you could potentially have a leak
Considerations for opening of cylinder (5)
Open valve slowly and slightly prior to installation to clean out the valve port Check pressure Open away from patient Face valve away from people Correct leaks
Describe the Pin index safety system (PISS)
Two holes on cylinder valve
positioned in an arc = receive pins on the yoke/ pressure regulator
THERE ARE 7 DIFFERENT POSSIBLE PIN POSITIONS
DEPENDING ON THE TYPE OF GAS IN THE CYLINDER
If cylinder valve has no holes = impossible to attach to yoke/regulator with pins
SPECIFIC CONFIGURATIONS BE AWARE OF THIS SAFETY MECHANISM
Pipeline supply source guidelines
Must have 2 days supply “banks”
1 primary
1 reserve
each with 2 days supply so 4 days total for both
Purpose of pipeline system and gases supplied versus cylinder?
Because E- cylinder use is not enough Used to deliver gases to anesthetizing locations & patient care areas • O2 • N2O • MEDICAL Air
What is included in a pipeline supply system?
central supply system
piping to transport gases to the specific locations
branches and terminal units
What is the US pipeline pressure?
What is the most frequent pipeline problem?
pipeline pressures are 380 kPa or approx. 55 psi
Low pressure is the most frequently reported problem in pipeline systems
Purpose and guidelines of reserve gas supply?
Reserve supply
should be used for emergencies or failure of primary supply.
Ideally in different area with different routing
Guidelines for liquid O2 use
Liquid O2 must be in constant use to be cost effective
Otherwise pressure ↑ as the liquid boils and is then vented in the Patm
N2O supply and storage guidelines and regulations
generally supplied by manifold cylinder system
• b/c regulator prone to freezing
Warning signs need to be posted in N2O are warning of asphyxia due to leak
Medical air supply storage requirements
use manifolds or compressors
• important for intake locations are free of contaminants
Air systems need to dehumidify to qualify for medical use
N2O and medical air storage guidelines
Both systems have a series of valves, pressure regulators, & alarms
to regulate pressure and signify problems
What are the components and composition of a pipeline system structure
Main line
• connect gas source to risers
Risers
• vertical pipes connecting mainline w/ branch lines on each level of facility
Branch
• sections supplying a room or group of rooms on one level of the facility
Composition of piping is Copper
What are labeling regulations for pipeline systems
Regulation:
Name, pressure, and flow direction
• must be clearly marked every 20 ft and in each room
What is the difference in O2 pipeline diameter vs other gases and why?
O2 (1/2 in OD) has different outer diameter than other gases (3/8 in OD) safety mechanism
Purpose of pipeline system shut off valve
Allow for certain areas in the piping system to be isolated
for maintenance or problems
ON-OFF, isolation of section or zone
2 types of pipeline system shut off valves and their differences.
Where are they located?
“Manual Shut Offs” = visible & accessible at all times
Mandatory locations:
• Main supply into building= turns everything OFF to building
• One at each riser
• One at each branch
• except if branch is to an anesthetizing area or critical care area-Why?
“Service Shut Offs” = locked box- NOT ACCESSIBLE TO US
What are 2 types of pipeline alarms and their differences?
Master Alarm System
Must be located in 2 different areas
• 1 panel must be in department responsible for maintaining system
Monitors entire pipeline system
Area Alarm System- check daily, test monthly
In critical care areas- ICUs and ORs
Alarms if the pressure ↑/↓ at 20% from normal line pressure***
May trigger master alarm
What are pipeline alarm requirements
Must be audible and visible
Must be labeled for gas and area
Will alarm with a 20% increase or decrease
What are terminal units of a pipeline system
Point where piped gas is accessed
by user through hose connections (flow-meter)
i.e. wall connector-hose-station outlet
Describe pipeline terminal unit connections
Connection into wall uses quick connectors
Pair of male and female parts
• only connect w/ proper alignment.
Each gas has a specific shape and spacing more prone to leaks vs diss system
What is the station outlet of the pipeline terminal unit?
Connection into machine
uses DISS (diameter index safety system) system
Vaporizer definition and additional requirement
Vaporizer = device that changes a liquid anesthetic into a vapor for inhalation
Changes liquid to gas
Must add a controlled amount of vapor to FGF in a breathing system
How is the vaporizer calibrated?
Calibrated at sea level
Affected by barometric pressure changes
What are the stages of anesthesia?***
STAGES OF ANESTHESIA
1) ANALGESIA
2) DELIRIUM
3) SURGICAL
4) RESP CESSATION
How is the vaporizer affected by pressure and what should the anesthetist do prior to using it?
BMP changes volatile given
READ THE MANUAL FOR THE VAPORIZER AND MACHINE
• PRESSURE MUST BE TAKEN INTO ACCOUNT
When considering vapor administration and pressure changes what are the 3 most important considerations for the anesthetists?
GIVE THE PT WHAT THEY NEED
LOOK AT WHOLE PICTURE
MAKE THE CLINICAL DECISION
What is the vapor pressure of a liquid
the equilibrium pressure OF the vapor ABOVE ITS LIQUID
What is pressure of vapor
RESULTS FROM EVAPORATION OF THE LIQUID
ABOVE THE LIQUID IN A CLOSED system at a CONSTANT temp**
How does temperature relate to VP
increase temp = increase VP
How is VP affected by barometric pressure
It is not
VP is dependent on the liquid and temperature
What is VP dependent on?
The liquid and the temperature
What is saturated vapor pressure
WHEN THE GAS CONTAINS ALL THE VAPOR IT CAN HOLD AT A GIVEN TEMP
What is saturated vapor concentration and important consideration during administration
SVP/atm pressure
THE SVC MUST BE DILUTED BY A BYPASS GAS FLOW
What is important to note about each gases VP?
Cannot compare gases by VP
What is the best way to compare gas agents
Gas concentration in partial pressure or vol % (MAC)
Describe volume percent expression of gas concentration
Concentration of a gas in a mixture
• expressed as percentage of 100% @ 1 ATM
• PP/TP X 100% = volumes %
• MAC is described in terms of volume percent
What affects volumes percent and how is it related to anesthetic uptake?
Indirectly relates to patient uptake & anesthetic depth
• IS INFLUENCED BY BAROMETRIC PRESSURE
Describe the partial pressure expression of gas concentration and what is it dependent on?
Pressure exerted by any one gas in a gas mixture on the total gas mixture
• Total pressure of the mixture is a sum of all the PP
- Dependent on temperature
- Not the pressures exerted by total pressure of the gas mixture
What affects partial pressure of a gas and how is PP related to anesthetic uptake?
Affected by: temperature
• Not affected by barometric pressure
• SAME POTENCY NO MATTER WHAT THE BAROMETRIC PRESSURE IS
•Directly relates patient uptake and anesthetic depth
What is heat of vaporization
The number of calories necessary to convert 1G (or 1ml) of liquid into vapor**
How does gas flow affect the heat of vaporization?
As carrier gas flows through the vaporizers
Vapor molecules leave and more liquid is vaporized
As more molecules enter the gas phase, the liquid begins to cool.
How does heat flow as gas flows pick up vapor? When is equilibrium established?
The liquid begins to cool
Heat will flow from the surrounding VAPORIZER
To compensate for the loss of heat in the liquid
Heat lost to vaporization = heat supplied by surrounding VAPORIZER
What happens to the liquid in the gas state in the vaporizer when the temp drops? Why is this significant?
IT DECREASES:
• IF MY VAPORIZER IS SET AT 1, BUT ITS COOLING
• THERE IS LESS GAS LEAVING THE VAPORIZER
EVEN THOUGH NOTHING HAS CHANGED
What happens to the temp as gas leaves the vapor
It cools the whole vaporizer
How do manufacturers account for heat of vaporization effects?
MANUFACTURERS MUST ACCOUNT FOR THIS PHENOMENA
INCORPORATE A SYSTEM TO REPLACE AND EQUALIZE LOST HEAT
FOR ACCURATE GAS ADMINISTRATION
• THEY CHOSE CERTAIN METALS TO ACCOMMODATE HOV
• SO WE GET A CONSISTENT ANESTHETIC
Specific heat definition
the quantity of heat required to raise the temp of 1gram of a substance 1degree C
What is the standard measure of specific heat
H2O is the standard
• 1cal/gram/1 degree C
Specific heat considerations for vaporizers
SH must be considered for maintaining a constant temperature to the vaporizer
Higher SH = temp changes more gradually
Thermal conductivity definition
the speed at which heat flows through a substance
Thermal conductivity considerations with vaporizers
Higher TC = better conductor of heat
Specific heat and thermal conductivity considerations with vaporizers
must be considered in choosing a material/METALS for vaporizer construction
↑ SH=THE HARDER IT IS FOR THE TEMP TO CHANGE-
Higher SH/TC is desirable b/c changes in the the vaporizers are less likely when the gas leaves the vaporizer
Vapor concentration calibration and location
Calibrated by agent concentration
Single knob calibrated in volumes percent
Located between flow meter and common gas outlet (PT) - so it isn’t flushed with O2
Where is the vapor concentration calibration and Why is the location of vapor calibration site important.
Located between flow meter and common gas outlet (CGO)
So gas isn’t flushed w/ O2
How does the vaporizer maintain steady state?
- THE TOTAL VOLUME OF GAS LEAVING THE VAPORIZER IS»_space; THE TOTAL VOLUME THAT ENTERED
- D/T ADDITIONAL VOLUME ATTRIBUTED TO ANESTHETIC VAPOR AT ITS SVC
2 Types of vaporizers**
bypass
electronic
How does vapor pressure and partial pressure of a gas at room temp relate to anesthetic depth?
Why is this important?
Vapor pressure of an anesthetic gas at room temp is GREATER than the partial pressure necessary to achieve anesthesia
so the vaporizer dilutes to a useful concentration
↓ potency of gas w/ mix of vapor and gas flow
What is a variable bypass vaporizer?
Vaporizer splits flow
some gas flows into vaporizing chamber
some gas flows into bypass
THIS IS KNOWN AS THE SPLITTING RATIO
Gas that passes through the vaporizing chamber will have volatile agent attached
Both flows are then rejoined before exiting vaporizer
Process of the variable bypass vaporizer
Gas flow is split
Portion of gas picks up volatile in vaporizing chamber
Portion of gas bypasses vaporizer chamber
Both flows are rejoined before exiting vaporizer
What is splitting ratio dependent upon? (4)
the ANESTHETIC AGENT/CONCENTRATION
size of the adjustable orifice
total gas flow
heat of vaporization
How do temperature compensating mechanisms affect variable bypass vaporizers?
THEY DO NOT PRODUCE INSTANTANEOUS RESULTS
T/F ANTICIPATED CONCENTRATION MAY NOT BE ACCURATE UNTIL COMPENSATION OCCURS
What is a bypass vaporizer
NO VOLATILE ATTACHED TO FLOW THROUGH BYPASS CHAMBER VS VAPORIZING CHAMBER
How is splitting ratio determined for bypass vaporizer
Splitting ratio = vaporizing/bypass
Higher ratio means
MORE GOES INTO VAPORIZING CHAMBER
t/f MORE GOES TO THE PATIENT
What happens to the splitting ratio if the vaporizer is cooled?
How is the problem fixed?
• IT GETS SMALLERSO LESS GOES OUT TO PATIENT
Increase MAC or FGF to increase depth
How does an electronic vaporizer work?
Computer driven Calculates either: volume of CARRIER gas • to produce the desired concentration OR amount of liquid agent needed to be injected into carrier flow To EQUAL desired concentration
2 methods of vaporization delivery
Injection
Flow-over
What are the differences between injection and flow-over vapor delivery
Injection
Inject known volume liquid anesthetic into known volume of gas
Flow-over
Carrier gas passes over surface area of a liquid
↑ surface area= ↑ efficiency of vaporization
Most common
How is temperature accounted for when delivering volatile gas
Thermocompensation
must maintain constant ANESTHETIC output
Can be mechanical or computer driven
Splitting ratio changes as temp changes
How is the splitting ratio affected by temperature
Splitting ratio changes as temp changes
• ↓ TEMP AND ↓ OUTPUT OF GAS/VAPOR
• SPLITTING RATIO WILL ↓ W/O THERMOCOMPENSATION
What will happen to splitting ratio w/o thermocompensation
Splitting ratio will decrease
Purpose of regulating intermittent back pressure on vaporizers
To limit the change in vaporizer concentration from IBP
To keep a steady state of volatile delivery
Causes of intermittent back pressure
O2+ flush
Positive pressure from inspiration during vent use
What can IBP cause during vaporizer use?
Most common reason?
Can Cause pumping effect or pressurizing effect on vaporizer outputs
Most common: d/t vent and flush valve use
Problem of having intermittent back pressure occur
Can cause small changes in the amount of gas delivered
IBP CAN AFFECT WHAT THE PATIENT IS GETTING
How does intermittent back pressure affect delivery of gases
By leading to the pumping or pressurizing effect
What is the pumping effect due to and how does it affect volatile delivery?
D/t back pressure during inhalation @ LOW FGF
• Causes INC flow into the vaporizing chamber
• MORE than usual vapor picked up
Effect
= ↑ of vapor output
• COMMON WITH OLDER VAPORIZERS
What can lead to pumping effect and how can it be minimized?
More common
• @ LOW FGF
• large pressure fluctuations
• low vaporizer settings
Minimize effect
• Presence of the pressurizing valve
• unidirectional valve
• pressure relief valve
How do pressurizing and pumping effect alter gas delivery
pressurizing = not enough vapor output
pumping= too much vapor output
What is the pressurizing effect due to and how does it affect volatile delivery?
D/t back pressure at HIGH FGF
• causes INC density into the vaporizing chamber
• LESS than usual vapor picked up
Effect
=↓ of vapor output
What can lead to pressurizing effect and how can it be minimized?
More common • at high gas flows • O2 flush valve use • large pressure fluctuations • low vaporizer settings
Minimize effects
NEWER, CONTEMPORARY VAPORIZERS
USE VALVES AND OTHER MECHANISMS TO MINIMIZE THESE ISSUES
How does FGF affect vaporizer?
It directly affects vaporizer output
Definition and effects of high FGF
Flow > pt minute ventilation
Little gas rebreathed
Inspired concentration = vaporizer setting
HIGH FGF»_space; THAN PATIENT’S MIN VOLUME
• Brand new breath w/ brand new volatile
• NO RECYCLING
• Pt is getting closer to exact gas concentration
• Pt gets what is on the dial
Definition and effects of low FGF
Flow < pt minute ventilation Significant rebreathing Difference btwn vaporizer setting & inspired concentration • Difference is in bellows or bag Takes longer to achieve equilibration
How can equilibration be reached faster with low FGF
increase volatile concentration
Increase flow
Difference in high and low FGF monitoring, effects and use?
Low FGF Flow << min ventilation There's a difference between setting and inspired concentration Need agent analyzer to get true value significant rebreathing Used during maintenance
High FGF Flow >> min ventilation inspired concentration = vaporizer setting NO rebreathing useful during induction
Where is the difference between min volume and flow w/ low FGF use?
In the vent bellows or reservoir bag
What is the problem with rebreathing during low FGF use?
Recycling present = patient not necessarily getting dialed concentration
Takes longer to reach equilibrium
What are standard regulations for vaporizers? (6)
Average concentration +/- 20% setting
Gas may not pass through more than 1 vaporizer
- SAFETY=to prevent mixing of gases
Output of vaporizer <0.05% in OFF
All control knobs counterclockwise
Filling levels displayed
Cannot overfill when in normal operating position
Types of vaporizer mounting systems.
Which is most common
Permanent
Detachable = most common
Describe the permanent vaporizer mounting system including advantages and disadvantages
Tool required Always filled in upright position Advantage • fewer leaks/damage Disadvantage • limited mounting locations • not easily exchanged
Describe the detachable vaporizer mounting system including advantages and disadvantages
Most common Weight of vaporizer & “O” ring create seal • O ring damage = can lead to leak Locking lever on back Control must be OFF before mounting Easily removed/replaced • think MH= machine has to be flushed Disadvantages: • Site for leak from • damaged O ring • unlocked lever • Manufacturer compatibility
Purpose of interlock device on vaporizer
– prevents more than 1 vaporizer being turned ON at a time
Allows only ONE gas to be administered at a one time
Prevents mixture of gases
4 vaporizer hazards
Incorrect agent
Tipping
Overfilling
No Vapor output-
How is the vaporizer hazard of incorrect agent prevented and how is it fixed if it occurs.
Filling systems agent specific=not likely
• Must have correct/compatible key to fill system
If contaminated filling occurs:
• Must be completely drained
• all liquid discarded
• FGF run until no vapor detected
What happens to potency when the incorrect agent is put in the vaporizer?
- Won’t achieve 1 MAC
- Won’t deliver what the pt needs
- dec potency
How does the vaporizer hazard of tipping occur, prevented, the effects and how is it fixed if it occurs.
Liquid may get into bypass or outlet
INC concentration of agent
• Bypass carrier flow will pick & deliver more agent
Prevented:
Should be placed in OFF/ travel mode when moved
Newer vaporizers prevent
If tipped:
• High FGF run with low concentration of vapor
• until excessive vapor exhausted
• Until agent analyzer reads concentration that is set on dial
How does the vaporizer hazard of overfilling occur, the hazard and what can this lead to
Liquid may enter FG line or cause vaporizer failure
Potential for lethal dose
Can occur during tipping or filling on “ON”
Leads to: • failure to tighten filler cap • fill valve not closed • malfunctioning mount/vaporizer • pollutes OR can probably smell
Causes of no vapor output
Most common cause = empty
Incorrect mounting
Overfilled = no output because of vaporizer failure
What is the anesthesia gas delivery system composed of? (5)
anesthesia machine vaporizers ventilator breathing circuit, scavenging system
Who sets the standards for anesthesia workstations? When were standards published
American Society for Testing and Materials (ATSM)
starting in 2000
reviewed in 2005
What does the term “workstation” include and the basic operations.
include the ventilator & associated monitoring devices used
operations
components have become more technologically advanced (i.e. virtual flowmeters)
What should workstations be able to do? (5)
Provide: • accurate & safe gas delivery • a means for ventilating patients • electrical outlets • a housing for monitoring devices like vaporizers • storage/shelving for other equipment
How does gas travel through the entire workstation?
Gas source delivers gas to machine
gas delivered to flowmeters & vaporizers
gas mixture goes to common gas outlet (CGO)
then flows to breathing circuit to the pt
gas then leaves the pt through breathing circuit
excess gas exits either via APL valve (spontaneous resp) scavenger or ventilator
(Mech resp)
Disadvantages to older workstations
may lack the safety features of newer machines
may be considered obsolete
there are ASA guidelines to consult to determine if this is the case
What are workstation system components (19)
- Master switch
- Hanger yoke assembly
- Cylinder pressure indicator/gauges
- Cylinder pressure regulators
- Pipeline inlet connections
- Pipeline pressure indicators/gauges
- Gas pathways
- Machine piping
- Common gas outlet***
- Unidirectional valves
- Pressure relief valve
- Flow adjustment controls
- Flowmeters
- O2 flush valve
- O2 failure protection device
- Hypoxia prevention devices
- O2/N2O linkage
- Auxiliary O2 flowmeter
- Power backup/battery
Components of the workstation (16)
Master switch (ON/OFF) Hospital backup/electrical outlets Hanger yoke assembly Cylinder pressure gauges/regulators Pipeline pressure gauges Gas pathways Pressure relief valves Flow controls/flowmeters Vaporizers Unidirectional valves (outflow check valve) Common gas outlet (CGO) Oxygen flush valve Oxygen failure protection device (OFPD) Minimum oxygen flow and ratio Axilllary oxygen flowmeter Power failure/battery backup