Doppler Pt 1 Flashcards

1
Q

occurs when blood clots in vessels

A

thrombosis

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

narrowing of a blood vessel

A

stenosis

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

something closed of blocked off

A

occlusion

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

does occluded artery or thrombosed vein show color?

A

no

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

___ is a normal cause of turbulence

A

tortuosity

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

arterial stenosis can lead to

A

post stenotic turbulence

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

filling in of spectral window on doppler tracing caused by turbulence

A

spectral broadening

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

pressure gradient = 4 (velocity)^2

A

Bernoulli equation

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

peak systolic = 3 m/s

peak pressure gradient = ?

A

36 mmHg

pressure gradient = 4 (velocity)^2 = 4 (3)^2 = 36

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

doppler evaluation provides speed and direction of flow

A

velocity

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

location where doppler shift is being obtained

A

sample volume

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

pressure is lowest at

A

stenotic site

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

increasing threshold ___ the amount of color displayed in slow moving vessels

A

increases

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

PSV - EDV / PSV

A

resistive index

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

if EDV = 0, RI =

A

1

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

reducing sample size improves

A

temporal resolution

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

changing color map in real time

A

preprocessing

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

waveform with 2 components of antegrade flow during cardiac cycle

A

triphasic

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

increased frame rate =

A

more color displayed

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

flow distal to stenosis

A

reduced doppler shift (velocity), low resistance turbulent waveform, and greater bandwidth

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

stenosis demonstrates

A

high velocity, higher resistance, and lower pressure

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

____ can reduce beam width artifact, clutter, and crosstalk

A

increasing wall filter

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
23
Q

beam width artifact, clutter, and crosstalk all involve

A

incorrect display of low level doppler shifts

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
24
Q

when referring to blood flow volume throughout the body, ____ and ____ are the best answer

A

cardiac function

vessel resistance

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
25
if speed of sound in soft tissue increases, doppler shift
decreases
26
FFT requires a minimum of __ samples/cycle to process and produce a _____
2 | doppler tracing
27
FFT cannot correctly process a frequency shift greater than
1/2 PRF | Nyquist limit, aliasing will occur
28
most systems obtain ___ doppler samples per scan line
100-400
29
what is the appearance of an aliasing waveform?
wrapping around the baseline
30
to correct spectral doppler wrap around
increase PRF to allow proper display of high velocities, BUT use a lower frequency to provide more depth penetration to fully resolve the issue
31
PRF =
velocity scale
32
brightness of the reflection on the doppler tracing represents
the # of RBCs moving at each velocity displayed
33
lower reflectivity makes blood
more difficult to see at greater depths than ST | reason you can see deep vessel with no color flow
34
process to produce color doppler shift info
autocorrelation
35
used to produce spectral doppler settings
FFT
36
determines direction of flow, either forward or reverse with respect to transducer but does not produce spectral tracing
quadrate phase detection
37
unable to detect flow less then 6 cm/s, unable to detect post stenotic turbulence and spectral broadening, and usually underestimates high velocities and over estimates low velocities
analog doppler
38
increasing sweep speed will aide in ability to properly
trace the contour of the doppler waveform
39
larger packet in color doppler =
better flow sensitivity
40
color is assigned to flow based on # of moving RBCs, it is NOT related to velocity of the movement
power doppler
41
compares peak to mean velocity throughout 1 cardiac cycle
pulsatility index
42
noise can be reduced by
decreasing doppler gain, increasing wall filter, or trying another wall outlet
43
used to improve SNR on color image and defined as averaging of previous image frames to produce newest frames
persistence
44
highest velocity throughout a stenosis is
immediately distal to focal point of stenosis
45
to reduce clutter, ___ color priority
decrease
46
reduced sample size = ____ spectral window
larger
47
for the best velocity calculation, what must be known?
angle between flow and sound beam
48
doppler shift and frequency relationship
direct
49
frequency is reduced by 1/2, ___ decrease in detected flow, machine will display ___ velocity
50% | same
50
resistance of a vessel is determined by
the part it is supplying
51
____ on doppler exam will cause false increase in peak velocity
overgaining
52
what is necessary to quantify doppler shift?
cos of angle frequency speed of sound in ST
53
to eliminate crosstalk
lower gain
54
color is not angle dependent when using
power doppler
55
anemia affect on blood velocity
``` increased velocity (thinner blood = reduced viscosity = increased velocity) ```
56
2D imaging = ___ pulse per scan line | Color doppler = ___ pulses per scan line
1 | 3-32
57
power doppler has ___ frame rates than color
slower
58
to measure acceleration time
place a caliper at the onset of the systolic upstroke and at the early systolic peak
59
measurement used to assess flow in organs, extremities, and through cardiac valves
acceleration time
60
radius decreases by 1/2, flow volume
decreased by 2^4 (16x)
61
repetitive echoes above the baseline are consistent with
electrical interference
62
color doppler displays ___ velocity
mean
63
PW displays ___ velocity
peak
64
perceived color used to demonstrate direction; reds and blues; determined by direction (+ or -) of the shift
color hue
65
brightness of the color displayed; determined by magnitude of the reflection
luminance