Decontamination Summary Flashcards
What type of water can be used for manual cleaning ?
Tap water.
What type of water can be used in ultrasonic bath ?
Tap water.
What type of water can be used in cleaning stages of AWD ?
Tap water.
What type of water should be used for thermal disinfection during AWD cycle ?
Water with conductivity lower than <30ms recommended.
What type of water should be used for sterilisation ?
Purified water.
What temperature should water be for manual cleaning ?
<35 degrees.
What water temperature should be used for ultrasonic bath cleaning ?
<35 degrees.
What water temperature should be used for AWD cleaning stage ?
Depends on chemical manufacturers instructions - usually <65 degrees.
What water temperature should be used for thermal disinfection stage of AWD ?
90-95 degrees.
What water temperature should be used for sterilisation ?
134-137 degrees.
What is the most common cycle time for ultrasonic bath ?
20 mins - depends on validated cycle time.
What is the most common cycle time for AWD cleaning stage ?
55 mins - depends on validated cycle time.
What is the minimum time for thermal disinfection ?
Hold time of 1 min.
What is the minimum time for sterilisation ?
Hold time of 3 mins.
What type of chemical should be used for manual washing ?
pH neutral or enzymatic detergent.