Cold sterilisation Flashcards
What are the THREE factors to gas sterilisation?
- Ethylene oxide
- Factors influencing efficiency
- Control of process
What are the three factors to sterilisation by ionising radiation?
- Ionising radiation
- Factors influencing efficiency
- Control of process
What are the TWO main categories of gaseous sterilising agents? How are they distinguished? Explain in further detail about these agents.
Distinguished by their antimicrobial action
- Alkylating agents are highly reactive compounds and interact with many components, such as amino, sulfhydryl and hydroxyl groups in proteins and purine bases in nucleic acids
> Ethylene oxide is an alkylating agent that is associated with cytotoxic, carcinogenic and mutagenic effects
- Oxidising agents are highly reactive, toxic compounds. Such compounds currently used as sterilising agents include hydrogen peroxide and peracetic acid
For ethylene oxide (alkylating agent);
A) Describe its structure
B) When is it explosive
A)
- Simplest cyclic ether
- A colourless gas at room temperature
- Liquefaction point, about 12°C (becomes a liquid at 12°C or below)
- Water and organic solvents dissolve it readily
B)
- [>3%] in air are highly inflammable
- Explosive when mixed with air over a wide range of concentrations
What are the properties of ethylene oxide? What types of toxicity does it result in?
Non- selective in action so equally toxic to the operator as to the bioburden
Contact toxicity
- oedema, erythema, skin eruptions, allergic reaction
Vapour toxicity
- eye & nose irritation, nausea, dizziness (all below smell detection levels)
What is the action of ethylene oxide?
Alkylates –SH; amines; -COOH; -OH groups occurring on enzymes & nucleic acids
> chemical reaction that inactivates/kill micro-organism (see attached image)
- cyclic ether = very reactive
- Very wide spectrum –> vegetative bacteria; large viruses
- Less reliable when compared to heat sterilisation
For ethylene oxide applications;
A) What types of thermolabile equipment is it used in
B) What type of medicaments is it used in
A)
- plastics
- lensed equipment
- electrical
- Rubber
B)
- thermolabile solids (outside only), rarely use nowadays
- Alkylating agent –> toxic to the operator –> remove EO before opening sterilisation chamber
For gas sterilisation (BP) –> by ethylene oxide;
A) What is gas sterilisation of surfaces used for?
B) What is essential in order for the material to be sterilised?
A)
- It may be used for the sterilisation of primary packaging materials, equipment and some pharmaceuticals
B)
- It is essential that penetration by gas and moisture into the material to be sterilised is ensured
- Followed by a process whereby the gas is eliminated under conditions that have been previously established as sufficient to ensure that any residues of gas or related transformation by-products are below concentrations that could give rise to toxic effects during product use
For factors influencing efficiency: concentration of EO;
A) What does doubling gas concentrations do? (with limiting concentrations)
B) What concentrations are insufficient to achieve sterility within practical process time?
C) What are the usual concentrations in practice?
D) What other factors are interrelated?
A)
- Doubles the rate of inactivation
B)
- Concentrations of less than 300 mg/L are insufficient to achieve sterility
C)
- Concentrations are usually within a range of 500 to 800 mg/L
D)
- Influence of concentration and temperature are interrelated
For factors influencing efficiency;temperature of EO;
A) Under constant conditions and within the range of limiting gas concentrations, what kinetics does ethylene oxide follow?
B) When is rate of inactivation approximately doubled?
C) Why are the upper limits of temperature of less importance?
A)
- Follows first order chemcial kinetics
B)
- Every 10°C rise in temperature
C)
- The whole point of using ethylene oxide is to achieve cold sterilisation
For factors influencing efficiency; relative humidity of EO;
A) When is ethylene oxide ineffective?
B) How would a surface film of water promote contact?
C) What is the optimum humidity?
A)
- Ethylene oxide is quite simply ineffective against dehydrated microorganisms in a dry environment
B)
- A surface film of water would promote contact by dissolving ethylene oxide to form a concentrated solution
C)
- Optimum humidity level is 33 per cent
What are the SIX steps in the process of gas sterilisation by ethylene oxide?
- Remove air from chamber (EO dangerous in presence of air)
- Introduce moisture into chamber
- Introduce EO into chamber
- Once there is enough EO in the chamber (500-800 mg/L) –> sterilisation starts (55OC at 100mins –> double-check from diagram)
- Remove EO
- Air purge –> aeration of the load –> get rid of EO
For aeration of the load (get rid of EO);
A) What must all the air used for aeration pass through?
B) What is the aeration time affected by?
A)
- All air used for aeration must pass through a bacterial-retentive filter of 99.97 percent efficiency
B)
The aeration time is affected by
- Composition, thickness, design configuration and weight of the device and wrapping material
- Ethylene oxide concentration and time of the sterilisation process
- Rate and pattern of airflow
- Intended use of the device
For process monitoring of EO;
A) What is specialist equipment required to monitor during validation and routine operation
B) What is the effectiveness of the process dependent on?
C) What type of indicators is necessary for all gas sterilisation procedures, both for the validation of the cycles and for routine operations? What is used specifically for EO?
D) Where are these indicators placed?
A)
- Monitor temperature, humidity and gas concentration
B)
- number of parameters, including gas concentration, temperature, humidity, exposure time, load configuration and characteristics of the product and its packaging materials
- The effect on the process effectiveness of any change in one or more of these parameters shall be investigated
C)
- The use of biological indicators is necessary for all gas sterilisation procedures
- The use of spores of Bacillus subtilis is recommended for ethylene oxide
D)
- The use of spores of Bacillus subtilis is recommended for ethylene oxide
> chemical indicators can also be used
A suitable sample of each batch is tested for sterility, before the batch is released –> this is needed to be done for membrane sterilisation and sterilising material by EO but not for?
Not necessary for dry heat+moist heat sterilisation
> also occupational health and safety is necessary –> need to look after operators