Chromatography 2 Flashcards
Chromatograms
These are graphs which show when different components reach the detector.
Ideally each component should be seen as a separate peak.
Peak Area
The size of the peak produced for each component relates to the amount present in the original sample.
Computers attached to chromatographic equipment use the peak area to indicate the amount present.
The peak is seen where the response of the detector increases from the baseline, which represents the zero response of the instrument when no component is being detected.
Co-elution
This is where two components have not been completely separated and some of each component reached the detector at the same time.
It is desirable to have at least some baseline between the peaks.
Selectivity Factor and Resolution
Selectivity factor and resolution are two ways of expressing how well two components of a mixture are separated.
For qualitative analysis the components should be separated so that it is easy to see the different components and make identification using retention times.
For quantitative analysis, the components should be separated as the size of each peak related to how much was present in the sample analysed.
Selectivity Factor
The separation is expressed using only the retention times of each component.
Problems occur as some components will produce sharp peaks and other broader peaks.
Resolution
This utilises not only the retention time but also the point where each peak starts and ends.
This accounts for the difference in peak width.
Chromatography Detectors
Two different types of analysis can be conducted with chromatography, firstly the identity of components can be established based on the retention time or retardation factor.
The quantity of an identified component in a mixture can be established, by comparison to standards of differing concentration or by standard addition.
Different Detectors
Flame ionisation detector
Thermal conductivity detector
Nitrogen-phosphorous detector
Electron-capture detector
Atomic emission detector
Fourier-Transform Infra-Red detector
Mass spectrometer
UV-Vis detectors
Photodiode array dectors
Refractive index detector
Fluorescence detector
Electrochemical detectors
Flame Ionisation Detector (FID)
Responds to all organic compounds, anything that combusts.
With this the components leaving the column are burnt in a small flame fed by a mixture of hydrogen and air.
The temperature of the flame is such that the combustion products are immediately broken into ions. This generates more ions in the flame.
Ions are measured by conductance between two electrodes.
Thermal Conductivity (TCD)
This detector uses a heated filament immersed in the components leaving the column. Only the mobile phase is present, so the heat from the filament is constant.
When other components are present, the filament heats up. As the temperature changes, the electrical resistance does too.
Nitrogen-Phosphorous Detector (NPD)
An alkali bead is heated in the flame of an FID, which emits electrons which are collected at an electrode above the bead.
When compounds containing nitrogen or phosphorous are present, these coat the surface of the bead and less electrons reach the electrode.
Electron-Capture Detector
The detector uses a heated radioactive bead, which emits electrons and ionizes some of the carrier gas as it travels past, which produces a current between the two electrodes.
When compounds containing electronegative elements pass the bead, they absorb electrons and reduce the current between the electrodes.
Atomic Emission Detector
Compounds enter a heated chamber containing a microwave element. The microwave destroys the compounds and the atoms are excited by the heat of the chamber.
Light emitted by the excited atoms is separated into individual lines, similar to a prism with normal white light.
A detector then records each wavelength simultaneously. The chromatogram produced shows the elemental composition simultaneously, each line representing a chemical element.
Qualitative Use
Using detectors, retention time and retardation factor can be used to qualitatively identify components within a mixture.
Quantitative Use
The linear ranges for different detectors relative to how they respond to different concentrations of compounds or elements.
The larger the linear range, the less sample preparation required.
The sensitivity and limit of detection indicate how low a concentration the detector will actually respond to.