Chemical Examination of Urine Flashcards
*Most conventional method of urine chemical analysis
*Simple and rapid
CHEMICAL ANALYSIS OF URINE:
Reagent strip
*Light reflection from the test pads decreases in proportion to the intensity of color produced by the concentration of the test substance
*The darker the color of the reagent pad, the lesser the light reflection and vice versa
AUTOMATED REAGENT STRIP READERS
What is the principle of automated reagent strip readers?
REFLECTANCE PHOTOMETRY
- Dip the reagent strip briefly into a well-mixed uncentrifuged urine specimen
- Remove excess urine by touching the edge of the strip to the containers as the strip is withdrawn
- Blot the edge of the strip on a disposable absorbent pad
- Wait for the specified time for the reaction to occur
- Compare the color reaction of the strip pad to the manufacturer’s color chart under a good lighting condition
REAGENT STRIP TECHNIQUE
*Store with desiccant in an opaque, tightly closed container
*Store below 30°C (cool dry place); do not freeze
*Do not expose to volatile substances
*Do not use past expiration dates
*Do not use discolored pads
*Remove strips only immediately before use and tightly seal
CARE OF REAGENT STRIPS
CARE OF REAGENT STRIPS:
Store with __________ in an __________, __________ container
desiccant, opaque tightly closed
CARE OF REAGENT STRIPS:
Store below __________ (cool dry place); do not __________
30°C, freeze
CARE OF REAGENT STRIPS:
Do not expose to __________ substances
volatile
CARE OF REAGENT STRIPS:
Do not use ___________ dates
past expiration
CARE OF REAGENT STRIPS:
Do not use __________ pads
discolored
CARE OF REAGENT STRIPS:
Remove strips only __________ and __________
immediately before use, tightly seal
*Always check with a (+) and (-) controls a minimum of once every 24 hours or every shift
*QC must also be performed every opening of a new bottle; when there is a questionable result; when there is a concern in the integrity of the strips
*Dipstick reagent box insert must always be consulted for QC and QA
QUALITY CONTROL OF REAGENT STRIPS
QUALITY CONTROL OF REAGENT STRIPS:
Always check with a (+) and (-) controls a minimum of __________ or __________
once every 24 hours, every shift
QUALITY CONTROL OF REAGENT STRIPS:
QC must also be performed every opening of __________ when there is a questionable result; when there is a concern in the integrity of the strips
new bottle
QUALITY CONTROL OF REAGENT STRIPS:
Dipstick reagent box insert must always be consulted for __________ and __________
QC and QA
PARAMETERS ON A DIPSTICK:
PARAMETER = Glucose
DURATION =
PRINCIPLE =
30 seconds
Double sequential enzymatic reactions