Chapter 7 Flashcards
When were AECs introduced?
1942
Who introduced AECs?
Russell H. Morgan
What is the most critical element in using AEC?
The exact position of the ionization chambers
Automatic exposure devices provide
Diagnostic-quality exposures only for structures positioned directly above the ionization chambers
What is the most common relationship (configuration) of the three chambers?
The center chamber at the center of the image receptor with right and left chambers slightly higher.
Where does the most common three chamber relationship place the center chamber?
Below the duodenum and transverse colon for most abdominal examinations
Where does the most common three chamber relationship place right and left chambers?
Away from the mediastinum and completely within the lives of the lungs during chest x-rays.
How many different combinations of using the three chambers is there?
Seven
What is averaging?
The signals from the cells are sent through i a special operating lap amplifier, which sums the voltages received from each cell and divides by the number of cells that have been activated
Call selection is determined by….
The radiographer based on knowledge of anatomy and positioning
Automatic exposure controls require the use of _______________________ that specify the technical parameters to be used.
Exposure technique charts
If there is fluid in the lungs, does it cause increased or decreased subject density and contrast?
Increased, the AEC would remain on longer, making an aerated lung overexposed for diagnosis
What is the minimum response time?
The length is time necessary for the AEC to respond to the ionization and send a signal to terminate the exposure.
Modern AEC’s have a minimum response time in the region of ____ seconds.
0.001
What does the back up time establish?
The maximum exposure time for the system in order to prevent overexposure.