Chapter 13 Test Flashcards
A radiograph of the lumbar spine is produced using automatic exposure control (AEC). The exposure results in optimal contrast but low receptor exposure. Which of the following options would allow the tech to adjust IR exposure?
a.) increase grid ratio
b.) decrease density setting
c.) increase density setting
d.) increase kvp
c.) increase density setting
What is the primary function of the back up timer?
a.) prevent over exposure to the patient
b.) ensure adequate IR exposure
c.) ensure adequate contrast
d.) prevent over exposure to the IR
a.) prevent over exposure to the patient
When imaging a young child, a short exposure time is needed for an exam using AEC. The mA setting should be:
a.) any
b.) low
c.) mid range
d.) high
d.) high
___________ identifies optimal kvp values and alters the mas for variations in part thickness.
a.) variable kvp, fixed mas technique charts
b.) fixed kvp, variable mas technique charts
b.) fixed kvp, variable mas technique charts
Using fixed mAs and variable kvp technical factors, each centimeter increase in patient thickness requires what adjustment to kvp?
a.) increase 4 kv
b.) decrease 4 kv
c.) decrease 2 kv
d.) increase 2 kv
d.) increase 2 kv
What factors are important when using AEC devices?
1.) detector selection
2.) centering part to detector
3.) back up mas if set
a.) 2 and 3 only
b.) 1 and 2 only
c.) 1 and 3 only
d.) 1, 2 and 3
d.) 1, 2 and 3
Consider a technologist performing a lumbar spine series using an AEC system. After acquiring the lateral image, the technologist angles the beam 5 degrees caudad, correctly centers/positions the patient, and collimates to the area of the L5-S1 joint. Which technical parameter change will ensure proper receptor exposure?
a.) increase mA
b.) increase KVP
c.) no change is needed
d.) decrease SID
c.) no change is needed
If a radiographer is performing a PA chest using AEC at the upright bucky but the table bucky is selected:
a.) the patient will recieve no exposure because theyre no laying on the table bucky
b.) the back up time will likely be reached
c.) exposure will terminate after 5 minutes
d.) minimum exposure time will cut off at 0.5 seconds
b.) the back up time will likely be reached
Automatic exposure control (AEC) devices work by measuring:
a.) radiation that exits the patient
b.) radiation that is absorbed by the patient
c.) attenuation of primary radiation by the patient
d.) radiation exiting the tube
a.) radiation that exits the patient
When performing a lateral cervical spine using an automatic exposure control (AEC) system, the technologist activates the outer ionization chambers. How might this effect the radiographic image?
a.) decreased exposure time and decreased receptor exposure
b.) increased exposure time and decreased receptor exposure
c.) increased exposure time and increased receptor exposure
d.) decreased exposure time and increased receptor exposure
a.) decreased exposure time and decreased receptor exposure
A portable chest xray is to be performed on a patient in the ER and the technique chart suggests the exam should be performed with a grid at 120 kvp, 3.5 mAs, and a 72” SID. However, the room is too small to acheive a 72” SID and the exposure must be made at 55 inches. How should the technique be adjusted to compensate for the change in SID?
a.) 2 mAs
b.) 5 mAs
c.) remove the grid
d.) 21 mAs
a.) 2 mAs
While performing an AP T-spine, the technologist mistakenly centers 3’’ to the right of the vertebral bodies. This is likely to cause:
1.) appropriate exposure to the area of interest
2.) underexposure to the area of interest
3.) appropriate exposure to the part centered over the detector
4.) overexposure to the area of interest
5.) underexposure to the part centered over the detector
a.) 2 and 5 only
b.) 1 and 3 only
c.) 4 and 5 only
d.) 2 and 3 only
d.) 2 and 3 only
Increasing the anode angle will:
1.) decrease the anode heel effect
2.) increase the anode heel effect
3.) decrease IR exposure
4.) increase IR exposure
a.) 1 only
b.) 2 and 4 only
c.) 1 and 4 only
d.) 2 and 3 only
c.) 1 and 4 only
The AEC minimum response time may pose a problem when:
a.) a long exposure time is needed
b.) all options are correct
c.) the patient is unable to hold still for the exposure
d.) the patient can easily hold still for the exposure
c.) the patient is unable to hold still for the exposure
How much mAs is produced when the mA is 400 and the exposure time is 0.5s?
a.) 50 mAs
b.) 200 mAs
c.) 100 mAs
d.) 25 mAs
b.) 200 mAs
Increasing patient thickness while using an AEC device would result in:
a.) decreased brightness
b.) decreased mAs readout
c.) increased exposure time
d.) increased image contrast
c.) increased exposure time
An increase in the image size of an object compared with its true, or actual size is called:
a.) shape distortion
b.) size distortion
c.) sharpness distortion
d.) minification
b.) size distortion
The radiation sensitive portion of an ionization chamber is composed of:
a.) air
b.) scintillation crystal
c.) vibranium
d.) semi conductor crystal
a.) air
What is the effect of using an automatic exposure control (AEC) system to image a patient with a destructive pathology such as osteoporosis?
a.) decreased spatial resolution
b.) decreased exposure time
c.) increased contrast resolution
d.) increased receptor exposure
b.) decreased exposure time
In most AEC units, a typical detector configuration has at least ____________ detectors.
a.) 2
b.) 5
c.) 4
d.) 3
d.) 3
Xrays travel at half the speed of light. T/F
False
What competent within the automatic exposure control (AEC) system is responsible for measuring the quantity of photons exiting the patient?
a.) autotransformer
b.) rectifier bridge
c.) exposure timer
d.) ionization chamber
d.) ionization chamber
When using a radiographic unit with an automatic exposure control (AEC) , which of the following can be employed to reduce patient motion while maintaining proper receptor exposure?
a.) decrease kvp
b.) increase mA
c.) increase back up time
d.) decrease density setting
b.) increase mA
When there is a significant amount of scatter reaching the AEC detector, the exposure time will be:
a.) appropriate
b.) shorter than needed
c.) longer than needed
b.) shorter than needed
As collimation increases, the quanitity of scatter radiation decreases.
a.) true
b.) false
a.) true
Which of the following projections may be improved by using a low mA and long exposure time?
a.) lateral t spine
b.) decubitus chest
c.) supine abdomen
d.) AP axial cranium
a.) lateral t spine
Performing a KUB on a beriatric patient, using AEC results in:
a.) it depends on the type of grid used
b.) underexposure to the IR
c.) overexposure to the IR
d.) proper exposure to the IR
d.) proper exposure to the IR
When using automatic exposure control (AEC), the technologist CANNOT directly assign which of the following parameters?
a.) mAs
b.) kVp
c.) SID
d.) OID
a.) mAs
The 1st generation AEC detector is called:
a.) air filled photocells
b.) geiger counter
c.) ionization chamber
d.) phototimer
d.) phototimer
Which of the following exposure operates by ionizing air that creates an electrical charge?
a.) anatomically programmed technique
b.) none of the above
c.) phototimer
d.) ionization chamber
d.) ionization chamber
Increasing grid ratio:
1.) increases patient dose
2.) decreases IR exposure
3.) increases IR exposure
4.) increases distortion
a.) 2 only
b.) 1 and 2 only
c.) 1 only
d.) 3 and 4 only
b.) 1 and 2 only
Which of the following groups of exposure factors will produce the greatest receptor exposure?
a.) 400 ma, 70 ms
b.) 200 ma, 40 ms
c.) 600 ma, 30 ms
d.) 100 ma, 50 ms
a.) 400 ma, 70 ms