Chapter 10-Bioimaging Flashcards
Advantages of inverted microscope
- compatible with live cell imaging
Features of phase contrast microscopy
- does not require cell staining
- special objective required
- based on the principle that diff organelles have diff refractive indices, resulting in diff phase shift and hence intensity diff in phase rings
Define fluorescence
- process where substance absorbs photons and subsequently emits photons as emission light
- molecular: transition of electron from one molecular orbital (ground state) to another with higher energy
Define stokes shift
Emission fluorescent light has a longer wavelength (lower energy per photon) than the absorbed light (due to energy loss between the time a photon is absorbed and emitted)
Define photobleaching
A fluorophore permanently loses its fluorescence due to photon-induced chemical damage
Why is detergent used in immunofluorescence labelling?
Low concentration dissolves part of the cell membrane to increase permeability of the cell to allow antibodies to enter
Function of excitation filter in wide-field fluorescence microscopy
- select bandwidth of light to excite fluorophore
- as narrow as possible
Function of dichroic mirror in wide field fluorescence microscopy
- Reflect excitation light
2. Transmit emission light and suppress scattered excitation light
Function of emission filter in wide field fluorescent microscopy
- pass fluorescence emission light and further block scattered excitation light
Limit of light microscope resolution
d=200nm
What does the numerical aperture (NA) determine?
- resolution of microscope system
- NA=nsina
What is spatial resolution (d)?
- smallest distance of the 2 resolvable point light sources
- the smaller the d value, the higher the resolution
Practical resolution of TEM
0.1nm
TEM
- image area brightness is proportional to electrons collected by the detector
- transmitted electrons detected as light areas
- darker areas occur where electrons have been scattered or absorbed by sample
Define FSC
- refracted by cell in the forward direction and in the same direction the light was already travelling
- used to determine cell size
- the smaller the cell, the less scattered light