Ch.2 Mammo Flashcards
What is the frequency of ACR Acceditation?
Every 3years
The material used for the exit port of the mammography tube is necessary because
Regular glass would harden the emerging beam
The intensity of the X-Ray beam from the cathode side of the tube is generally higher because
Q.3 pg.48
The heel effect causes variations in the intensity of the X-ray beam
The design of the lip of the compression paddle (both height and angle along the chest wall) affects all the following except
Q. 4 pg.48
It ensures higher compression of the anterior breast tissue
The primary goal of compression is to
Q.5 pg 48
Allow uniform penetration of structures within the breast
What features of the high-transmission cellular (HTC) grid allow these grids to maintain equal or less radiation dose to the patient compared with linear grids ?
Q6 pg. 48
They have copper as the strips and air as the interspace material
The grid ratio can vary in modern mammography units. A common grid ratio used is
5:1
SN: can be as low as 3:1
The chest wall edge of the compression paddle should be aligned just beyond the chest wall edge of the detector to
Question 8 on Pg 48
Avoid projecting the chest wall edge of the paddle on the mammogram
Which of the following affects focal spot size ?
Angle of the anode
SN: the focal spot is the area where electrons strike on the target
SN: effective focal spot is the area projected onto the patient or detector
SN: spatial resolution is related to focal spot size, BUT CHANGES IN SID, OID, and Size of Collimation Field do NOT affect the focal spot size !
Which of the following is the commonly used focal spot size for routine work in mammography?
0.3mm
Which of the following mammography quality control test is performed as needed?
Repeat analysis
The general criteria to pass the new American College of Radiology-Digital Mammography (ACR DM) phantom require a minimum of
2.0 masses
See question 12 on PG 49
See answer question on PG 56-57
What is the compression thickness indicator test?
Q.13 pg 49
A check to ensure that the indicated compression thickness is within normal compression
What is the name given to the test to determine whether the unit has consistent detector signal-to-noise level for all breast thickness?
Visual checklist
A technologist test that is performed semiannually is
Q. 15 on pg 49
Compression force test
Repeats are
Images that involve unnecessary exposure to the patient
To get a meaningful analysis, the repeat analysis needs a patient volume of at least ?
250 patients
Mammography facilities can receive certification from
1. ACR
2 FDA
3. States as Certifiers (SACs)
2 or 3 only
Aluminum filtration is likely to be matched with _____ targets
Tungsten
Which of the following visual checklist tests must be performed before clinical use ?
Paddles/ face shield must not be cracked
What tool is used to evaluate for artifacts ?
Phantom
A single binary digit of data, such as “0” or “1” is called a
Bit
In general, higher magnification will require the use of a
Q.23 pg50
Smaller focal spot size
A grid is not necessary during magnification because
Q.24 pg50
The large OID produces the same effect as a grid