ch11,12,&15 Flashcards
every diagnostic imaging system MUST have
protective tube housing & a correctly functioning control panel
when operating at the highest voltage leakage cannot exceed
100mR/hr at a distance of 1 meter (exposure rate)
control panel (console) must indicate
condition of exposure and have light or audible x ray indicator
absorbs only minimal amount of radiation - commonly carbon fiber
radiolucent
must have a means of measurement from anode focal spot to IR
SID
devices that limits the x ray beam before it enters a patient (pre exposure)
beat limiting devices- collimator
tight collimation will
decrease patient dose
to minimize the skin exposure, the patients skin surface should be
at least 15 cm below collimator for fixed equipment
portable units, the source to skin distance is
at least 30 cm
positive beam limitation - PBL
automatic collimation
purpose of positive beam limitation
is to restrict the size and shape. so it does not exceed the size and shape of the IR
effects of filtration
increase quality of the beam & decreased quantity of x rays (by absorbing some low energy beams - fewer overall beams are left)
2 types of filtration
inherent & added
half value layer - HVL
decrease the intensity of the primary beam by 1/2 (50%) of its original value
wedge filter
bone & joint - t spine, feet (thicker part towards bones)
trough wedge
(bilateral wedge) chest PA (thin center)
consistent output in radiation intensity for identical generator setting from one exposure to the next
reproducibility
AEC back up timer
is set at 150%
when image is shown in display
digital image