Cedo - (Section 6) Magic Numbers Flashcards

0
Q

What is the radiation storage area limit?

A

0.025 mSv/hr

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
1
Q

What is the radiation barrier limit?

A

0.1 mSv/hr

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

What’s the radiation limit on the surface of a device?

A

2 mSv/hr

- Before use and upon installation of new source.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

A working calibrated survey meter must be capable of measuring in what range?

A

Between 20μSv to 100 mSv/hr

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Survey meters must be calibrated once every?

A

12 months.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

If a DRD reading exceeds 2 mSv during a shift, what should be done?

A

Stop work and investigate?

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

When must a leak test be performed?

A

Once a year or upon new source installation.

If a reading of 200 Bq or higher is obtained withdraw from use.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Long-handled tongs must be a minimum of what length?

A

1.5m

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Collimators must have a minimum shielding factor of?

A

100 SF

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Audible dosimeters must be calibrated how often?

A

12 months

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Audible dosimeters must alarm when intensity reaches a momentary dose of what? And a total absorbed dose of?

A
  • 5 mSv/hr momentary

- 2 mSv total absorbed dose

How well did you know this?
1
Not at all
2
3
4
5
Perfectly