CC - ONLINE EXAM Flashcards
- 10 6th
A. Kilo
B. Mega
C. Milli
D. Micro
B. Mega
- The prefix which means 10 -9 is:
A. Micro
B. Milli
C. Nano
D. Pico
C. Nano
- Concentration expressed as the amount of solute per 100 parts of solution:
A. Molarity
B. Normality
C. Percent solution
D. Ratio
C. Percent solution
- Indication of relative concentration:
A. Dilution
B. Molarity
C. Normality
D. Ratio
A. Dilution
- What is the molarity of a solution that contains 18.7 grams of KCl in 500 mL (MW 74.5)?
A. 0.1
B. 0.5
C. 1.0
D. 5.0
B. 0.5
Rationale:
M = 18.7 g / (74.5)(0.5L) = 0.5 M
- How much 95% v/v alcohol is required to prepare 5L of 70% v/v alcohol?
A. 2.4 L
B. 3.5 L
C. 3.7 L
D. 4.4 L
C. 3.7 L
Rationale:
C1V1 = C2V2
V1 = (70)(5L) / 95 = 3.7 L
- A colorimetric method calls for the use of 0.1 mL serum, 5 mL of the reagent and 4.9 mL of water. What is the dilution of the serum in the final solution?
A. 1 to 5
B. 1 to 10
C. 1 to 50
D. 1 to 100
D. 1 to 100
Rationale:
Amount of serum: 0.1 mL
Total volume: 10 mL
Dilution: 0.1:10 or 1:100
- Convert 72 Fahrenheit to its Celsius equivalent:
A. 12.2C
B. 22.2C
C. 40.2C
D. 44.4C
B. 22.2C
Rationale:
C = 5/9 (F - 32)
= 5/9 (72 - 32)
= 22.2C
- Convert 100 Celsius to its Kelvin equivalent:
A. 73.15K
B. 173.15K
C. 273.15K
D. 373.15K
D. 373.15K
Rationale:
K = C + 273.15
= 100 + 273.15
= 373.15K
- Most basic pipette:
A. Automatic pipette
B. Glass pipette
B. Glass pipette
- Does not have graduations to the tip:
A. Mohr pipet
B. Serologic pipet
C. Micropipet
D. None of these
A. Mohr pipet
Rationale:
A Mohr pipet does not have graduations to the tip.
It is a self-draining pipet, but the tip should not be allowed to touch the vessel while the pipet is draining.
- Pipets are used with biologic fluids having a viscosity greater than that of water:
A. Mohr pipets
B. Ostwald-Folin pipets
C. Pasteur pipets
D. Volumeteric pipets
B. Ostwald-Folin pipets
Rationale:
Ostwald-Folin pipets are used with biologic fluids having a viscosity greater than that of water. They are blowout pipets, indicated by two etched continuous rings at the top.
- Pipette with BULB CLOSER TO THE DELIVERY TIP and are used for accurate measurement of VISCOUS FLUIDS, such as blood or serum:
A. Ostwald-Folin pipette
B. Volumetric pipette
A. Ostwald-Folin pipette
- Pipette with cylindrical glass bulb near the CENTER of the pipette that helps to distinguish them from other types of transfer pipettes.
A. Ostwald-Folin pipette
B. Volumetric pipette
B. Volumetric pipette
- Extremely inert, excellent temperature tolerance and chemical resistance; used for stir bars, stopcocks and tubing:
A. Polyethylene
B. Polycarbonate
C. Polystyrene
D. Teflon
D. Teflon
Rationale:
POLYETHYLENE
● Widely used in plastic ware, too, including test tubes, bottles, graduated tubes, stoppers, disposable transfer pipets, volumetric pipets, and test tube racks.
● May bind or absorb proteins, dyes, stains, and picric acid
POLYCARBONATE
● Used in tubes for centrifugation, graduated cylinders, and flasks
● Usable temperature range is broad: –100° C to +160° C
● Very strong plastic but is not suitable for use with strong acids, bases, and oxidizing agents
● May be autoclaved but with limitations
POLYSTYRENE
● Rigid, clear type of plastic that should not be autoclaved
● Used in an assortment of tubes, including capped graduated tubes and test tubes
● Not resistant to most hydrocarbons, ketones, and alcohols
TEFLON
● Widely used for manufacturing stirring bars, tubing, cryogenic vials, and bottle cap liners
● Almost chemically inert and is suitable for use at temperatures ranging from –270° C to +255° C
● Resistant to a wide range of chemical classes, including acids, bases, alcohol, and hydrocarbons
- Horizontal-head centrifuge:
A. Cytocentrifuge
B. Fixed-angle head centrifuge
C. Swinging bucket centrifuge
D. Ultracentifuge
C. Swinging bucket centrifuge
Rationale:
HORIZONTAL-HEAD OR SWINGING BUCKET CENTRIFUGE
● HORIZONTAL WHEN MOVING OR SPINNING
● VERITICAL WHEN NOT MOVING
Cups holding the tubes of material to be centrifuged occupy a vertical position when the centrifuge is at rest but assume a horizontal position when the centrifuge revolves
- It is used when rapid centrifugation of solutions containing small particles is needed; an example is the microhematocrit centrifuge:
A. Horizontal-head centrifuge
B. Fixed-angle head centrifuge
C. Ultracentrifuge
D. Cytocentrifuge
B. Fixed-angle head centrifuge
- High-speed centrifuges used to separate layers of different specific gravities, commonly used to separate lipoproteins:
A. Horizontal-head centrifuge
B. Fixed-angle head centrifuge
C. Ultracentrifuge
D. Cytocentrifuge
C. Ultracentrifuge
Rationale:
● High-speed centrifuges used to separate layers of different specific gravities
● Commonly used to separate lipoproteins
● Usually refrigerated to counter heat produced through friction
- Uses a very high-torque and low-inertia motor to spread MONOLAYER OF CELLS rapidly across a special slide for critical morphologic studies:
A. Horizontal-head centrifuge
B. Fixed-angle head centrifuge
C. Ultracentrifuge
D. Cytocentrifuge
D. Cytocentrifuge
Rationale: (6mm)
Uses a very high-torque and low-inertia motor to spread monolayers of cells rapidly across a special slide for critical morphologic studies
Used for blood, urine, body fluid, or any other liquid specimen that can be spread on a slide
- The speed of the centrifuge should be checked every 3 months with:
A. Tachometer
B. Wiper
C. Potentiometer
D. Ergometer
A. Tachometer
- Calibration of centrifuges is customarily performed every ______.
A. Daily
B. Weekly
C. Every 3 months (quarterly)
D. Yearly
C. Every 3 months (quarterly)
Rationale:
Photoelectric tachometer or strobe tachometer
CAP recommends that the number of revolutions per minute for a centrifuge used in chemistry laboratories be checked every 3 months and the appropriate relative centrifugal force for each setting is recorded.
- Centrifuges are routinely disinfected on a ___ basis.
A. Daily
B. Weekly
C. Monthly
D. Quarterly
B. Weekly
- HIGHLY PURIFIED SUBSTANCES of a known composition:
A. Control
B. Standard
B. Standard
Rationale:
A standard may differ from a control in its overall composition and in the way it is handled in the test.
Standards are the best way to measure ACCURACY. Standards are used to establish reference points in the construction of graphs (e.g., manual hemoglobin curve) or to calculate a test result.
- It represents a specimen that is SIMILAR IN COMPOSITION TO THE PATIENT’S WHOLE BLOOD or plasma:
A. Control
B. Standard
A. Control
Rationale:
The value of a control specimen is known. Control specimens are tested in exactly the same way as the patient specimen and are tested daily or in conjunction with the unknown (patient) specimen.
Controls are the best measurements of PRECISION and may represent normal or abnormal test values.
- Water produced using either an anion or a cation EXCHANGE RESIN, followed by replacement of the removed ions with hydroxyl or hydrogen ions.
A. Deionized water
B. Distilled water
C. RO water
A. Deionized water
- The PUREST TYPE OF REAGENT WATER is:
A. Type I
B. Type II
C. Type III
D. All are equal
A. Type I
- Chemicals that are used to manufacture drugs:
A. Technical or commercial grade
B. Analytical grade
C. Ultrapure grade
D. USP and NF chemical grade
D. USP and NF chemical grade
- Basic unit for mass:
A. Gram
B. Kilogram
C. Mole
D. Pound
B. Kilogram
Rationale:
BASE QUANTITY
- Length (meter)
- Mass (kilogram)
- Time (second)
- Electric current (ampere)
- Thermodynamic temperature (Kelvin)
- Amount of substance (mole)
- Luminous intensity (Candela)
- Which of the following is NOT A COLLIGATIVE PROPERTY of solutions?
A. pH
B. Freezing point
C. Osmotic pressure
D. Vapor pressure
A. pH
Rationale:
The properties of osmotic pressure, vapor pressure, freezing point, and boiling point are called COLLIGATIVE PROPERTIES.
When a solute is dissolved in a solvent, these colligative properties change in a predictable manner for each osmole of substance present:
● FREEZING POINT IS LOWERED by −1.86°C
● VAPOR PRESSURE IS LOWERED by 0.3 mm Hg or torr
● OSMOTIC PRESSURE IS INCREASED by a factor of 1.7 × 104 mm Hg or torr
● BOILING POINT IS RAISED by 0.52°C
- Most clinical microbiology laboratories are categorized at what biosafety level?
A. 1
B. 2
C. 3
D. 4
B. 2
- Degree of hazard #2:
A. Slight
B. Moderate
C. Serious
D. Extreme
B. Moderate
- Electrical equipment fire:
A. Class A
B. Class B
C. Class C
D. Class D
C. Class C
Rationale:
Fires have been divided into four classes based on the nature of the combustible material and requirements for extinguishment:
● Class A: ordinary combustible solid materials, such as paper, wood, plastic, and fabric
● Class B: flammable liquids/gases and combustible petroleum products
● Class C: energized electrical equipment
● Class D: combustible/reactive metals, such as magnesium, sodium, and potassium
- Type of extinguisher for CLASS A FIRES:
- Pressurized water
- Dry chemical
- Carbon dioxide
- Halon
A. 1 and 2
B. 1 and 3
C. 1, 2 and 3
D. Only 1
A. 1 and 2
Rationale:
TYPE OF EXTINGUISHER (Bishop page 47)
● Class A: Pressurized water and dry chemical
● Class B: Dry chemical and carbon dioxide
● Class C: Dry chemical, carbon dioxide and halon
● Class D: Metal X
- All of the following are CRYOGENIC MATERIALS HAZARDS, EXCEPT:
A. Asphyxiation
B. Fire or explosion
C. Shock
D. Tissue damage similar to thermal burns
C. Shock
Rationale:
Liquid nitrogen is probably one of the most widely used cryogenic fluids (liquefied gases) in the laboratory.
There are, however, several hazards associated with the use of any cryogenic material: fire or explosion, asphyxiation, pressure buildup, embrittlement of materials, and tissue damage similar to that of thermal burns.
- Repetitive strain disorders such as tenosynovitis, bursitis, and ganglion cysts:
A. Cryogenic materials hazards
B. Electrical hazards
C. Ergonomic hazards
D. Mechanical hazards
C. Ergonomic hazards
Rationale:
The primary contributing factors associated with repetitive strain disorders are position/posture, applied force, and frequency of repetition.
Remember to consider the design of hand tools (e.g., ergonomic pipets), adherence to ergonomically correct technique, and equipment positioning when engaging in any repetitive task. Chronic symptoms of pain, numbness, or tingling in extremities may indicate the onset of repetitive strain disorders. Other hazards include acute musculoskeletal injury. Remember to lift heavy objects properly, keeping the load close to the body and using the muscles of the legs rather than the back. Gradually increase force when pushing or pulling, and avoid pounding actions with the extremities.
- The first step to take when attempting to repair electronic equipment is to:
A. Check all electronic connections
B. Turn instrument off and unplug it
C. Reset all the printed circuit boards
D. Review instrument manual
B. Turn instrument off and unplug it
Rationale:
Before REPAIR OR ADJUSTMENT of electrical equipment
The following should be done
- unplug the equipment
- make sure the hands are dry
- remove jewelry.
- When a person is receiving an electrical shock, all of the following should be done EXCEPT:
A. Pull the person away from the electrical source
B. Turn off the circuit breaker
C. Move the electrical source using a glass object
D. Move the electrical source using a wood object
A. Pull the person away from the electrical source
- Most common source of light for work in the visible and near-infrared regions:
A. Deuterium discharge lamp and mercury arc lamp
B. Incandescent tungsten or tungsten-iodide lamp
B. Incandescent tungsten or tungsten-iodide lamp
- The lamps most commonly used for ultraviolet (UV) work are:
A. Deuterium discharge lamp and mercury arc lamp
B. Incandescent tungsten or tungsten-iodide lamp
A. Deuterium discharge lamp and mercury arc lamp
- Which is the most sensitive detector for spectrophotometry?
A. Photomultiplier
B. Phototube
C. Electron multiplier
D. Photodiode array
A. Photomultiplier
Rationale:
Because of this amplification, the PM tube is 200 times more sensitive than the phototube. PM tubes are used in instruments designed to be extremely sensitive to very low light levels and light flashes of very short duration.
- Reflectance spectrometry uses which of the following?
A. Luminometer
B. Tungsten–halogen lamp
C. Photomultiplier tube
D. UV lamp
E. Thermometer to monitor temperature in reaction
B. Tungsten–halogen lamp
Rationale:
● Slide technology depends on reflectance spectrophotometry.
● For colorimetric determinations, the light source is a tungsten–halogen lamp.
- Which of the following light sources is used in atomic absorption spectrophotometry?
A. Hollow-cathode lamp
B. Xenon arc lamp
C. Tungsten light
D. Deuterium lamp
E. Laser
A. Hollow-cathode lamp
Rationale:
The usual light source, known as a hollow-cathode lamp, consists of an evacuated gas-tight chamber containing an anode, a cylindrical cathode, and an inert gas, such as helium or argon.
- Used to measure concentration of LARGE PARTICLES:
- Nephelometry
- Turbidimetry
- Absorption spectroscopy
A. 1 only
B. 2 only
C. 1 and 2
D. 1, 2 and 3
C. 1 and 2
Rationale:
Nephelometry and turbidimetry are used to measure the concentrations of large particles (such as antigen–antibody complexes, prealbumin, and other serum proteins) that because of their size cannot be measured by absorption spectroscopy
- Temperature is _______ proportional to fluorescence.
A. Directly proportional
B. Inversely proportional
C. No effect
B. Inversely proportional
- Low temperature:
A. Increase in fluorescence
B. Decrease in fluorescence
A. Increase in fluorescence
- Which of the following techniques measures light scattered and has a light source placed at 90 degrees from the incident light?
A. Chemiluminescence
B. Atomic absorption spectrophotometry
C. Nephelometry
D. Turbidimetry
C. Nephelometry
- Which of the following instruments is used in the clinical laboratories to detect beta and gamma emissions?
A. Fluorometer
B. Nephelometer
C. Scintillation counter
D. Spectrophotometer
C. Scintillation counter
- Liquids (reagents, diluents, and samples) are pumped through a system of [continuous] tubing:
A. Continuous flow analysis
B. Centrifugal analysis
C. Discrete analysis
D. None of these
A. Continuous flow analysis
- Which of the following types of analyzers offers RANDOM-ACCESS CAPABILITIES?
A. Discrete analyzers
B. Continuous-flow analyzers
C. Centrifugal analyzers
D. None of these
A. Discrete analyzers
Rationale:
All three can use batch analysis (i.e., large number of specimens in one run), but only discrete analyzers offer random-access, or stat, capabilities.
- Checking instrument calibration, temperature accuracy, and electronic parameters are part of:
A. Preventive maintenance
B. Quality control
C. Function verification
D. Precision verification
C. Function verification
Rationale:
Function verification includes monitoring temperature, setting electronic parameters, calibrating instruments and analyzing control data.
- PLEASE CHECK THREE (3) BOXES: Measures of center.
o Coefficient of variation
o Mean
o Median
o Mode
o Range
o Standard deviation
o Mean
o Median
o Mode
- PLEASE CHECK THREE (3) BOXES: Measures of spread.
o Coefficient of variation
o Mean
o Median
o Mode
o Range
o Standard deviation
o Coefficient of variation
o Range
o Standard deviation
Rationale:
The spread represents the relationship of all the data points to the mean. There are three commonly used descriptions of spread: (1) range (2) standard deviation (SD), and (3) coefficient of variation (CV).
- Systematic errors include: PLEASE CHECK 3 BOXES.
o Calibrator reconstitution
o Electro-optical mechanism
o Environmental conditions
o Fluctuations in line voltage
o Instability of instrument
o Reagent dispensing
o Reagent lot variability
o Sample evaporation
o Temperature of analyzer
o Variation in handling techniques: pipetting, mixing, timing
o Variation in operators
o Wear and tear of instrument
o Fluctuations in line voltage
o Reagent lot variability
o Wear and tear of instrument
Rationale:
A SYSTEMATIC ERROR, on the other hand, will be seen as a trend in the data. Control values gradually rise (or fall) from the previously established limits. This type of error includes improper calibration, deterioration of reagents, sample instability, instrument drift, or changes in standard materials. All the Westgard rules that indicate trends identify systematic errors. 2(2S), 4(1S) and 10(x) rule.
SYSTEMATIC ERRORS MAY BE DUE TO:
● Aging reagents
● Aging calibrators
● Instrument components
● Optical changes
● Fluctuations in line voltage
● Wear and tear of instrument
● Reagent lot variability
● Calibration differences
● Technologist interactions
- Random errors include: PLEASE CHECK 3 BOXES.
o Aging reagents
o Aging calibrators
o Calibration differences
o Instrument components
o Fluctuations in line voltage
o Optical changes
o Reagent lot variability
o Reagent dispensing
o Technologist interactions
o Variation in handling techniques: pipetting, mixing, timing
o Variation in operator
o Wear and tear of instrument
o Reagent dispensing
o Variation in handling techniques: pipetting, mixing, timing
o Variation in operator
Rationale:
RANDOM ERROR is one with no trend or means of predicting it. Random errors include such situations as mislabeling a sample, pipetting errors, improper mixing of sample and reagent, voltage fluctuations not compensated for by instrument circuitry, and temperature fluctuations. Violations of the 1(2S), 1(3S) and R(4S) Westgard rules are usually associated with random error. To assess the situation, the sample is assayed using the same reagents. If a random error occurred, the same mistake may not be made again, and the result will be within appropriate control limits.
RANDOM ERRORS MAY BE DUE TO:
● Reagent dispensing
● Sample evaporation
● Temperature of analyzer
● Electro-optical mechanism
● Calibrator reconstitution
● Environmental conditions
● Instability of instrument
● Variation in handling techniques: pipetting, mixing, timing
● Variation in operators
- Most frequently occurring value in a dataset:
A. Mean
B. Median
C. Mode
D. Range
C. Mode
Rationale:
The mode is the most frequently occurring value in a dataset. Although it is seldom used to describe data, it is referred to when in reference to the shape of data, a bimodal distribution, for example.
- Type of systemic error in the sample direction and magnitude; the magnitude of change is constant and not dependent on the amount of analyte.
A. Constant systematic error
B. Proportional systematic error
A. Constant systematic error
- Type of systemic error where the magnitude changes as a percent of the analyte present; error dependent on analyte concentration.
A. Constant systematic error
B. Proportional systematic error
B. Proportional systematic error
- Difference between the observed mean and the reference mean:
A. Bias
B. Confidence interval
C. Parametric method
D. Nonparametric method
A. Bias
Rationale:
Bias: Difference between the observed mean and the reference mean.
Negative bias indicates that the test values tend to be lower than the reference value, whereas positive bias indicates test values are generally higher.
Bias is a type of constant systematic error.
- Ability of a test to detect a given disease or condition.
A. Analytic sensitivity
B. Analytic specificity
C. Diagnostic sensitivity
D. Diagnostic specificity
C. Diagnostic sensitivity
- Ability of a test to correctly identify the absence of a given disease or condition.
A. Analytic sensitivity
B. Analytic specificity
C. Diagnostic sensitivity
D. Diagnostic specificity
D. Diagnostic specificity
- Ability of a method to detect small quantities of an analyte.
A. Analytic sensitivity
B. Analytic specificity
C. Diagnostic sensitivity
D. Diagnostic specificity
A. Analytic sensitivity
- Ability of a method to detect only the analyte it is designed to determine.
A. Analytic sensitivity
B. Analytic specificity
C. Diagnostic sensitivity
D. Diagnostic specificity
B. Analytic specificity
- Positive predictive value:
A. Ability of a test to detect a given disease or condition.
B. Ability of a test to correctly identify the absence of a given disease or condition.
C. Chance of an individual having a given disease or condition if the test is abnormal.
D. Chance an individual does not have a given disease or condition if the test is within the reference interval.
C. Chance of an individual having a given disease or condition if the test is abnormal.
- Negative predictive value:
A. Ability of a test to detect a given disease or condition.
B. Ability of a test to correctly identify the absence of a given disease or condition.
C. Chance of an individual having a given disease or condition if the test is abnormal.
D. Chance an individual does not have a given disease or condition if the test is within the reference interval.
D. Chance an individual does not have a given disease or condition if the test is within the reference interval.
- What percentage of values will fall between ±2 s in a Gaussian (normal) distribution?
A. 34.13%
B. 68.26%
C. 95.45%
D. 99.74%
C. 95.45%
Rationale:
● 68.26% will lie within ±1s
● 95.45% will lie within ±2s
● 99.74% will lie within ±3s
- Two (2) consecutive control values exceed the same 2 standard deviation limit:
A. 1:2S
B. 2:2S
C. R:4S
D. 4:1S
B. 2:2S
Rationale:
Westgard multirule is a control procedure that utilizes control rules to assess numerical quality control data; the control rules establish the limits for data rejection in a system with two controls. Other rules apply when three controls are used.
1:2s = 1 control value exceeds the mean ±2 standard deviations; warning rule that triggers inspection of control values using the other rejection rules that follow; only rule that is not used to reject a run; results are reportable
1:3s = 1 control value exceeds the mean ±3 standard deviations; detects random error
2:2s = 2 consecutive control values exceed the same 2 standard deviation limit (same mean +2 s or same mean -2 s); detects systematic error
R:4s = 1 control value in a group exceeds the mean +2 s and a second control value exceeds the mean -2 s, creating a 4 standard deviation spread; detects random error
4:ls = 4 consecutive control values are recorded on one side of the mean and exceed either the same mean +1 s or the same mean -1 s; detects systematic error
10:x =10 consecutive control values are recorded on one side of the mean (either above or below the mean); detects systematic error
- The term R:4S means that:
A. Four consecutive controls are greater than ±1 standard deviation from the mean
B. Two controls in the same run are greater than 4s units apart
C. Two consecutive controls in the same run are each greater than ±4s from the mean
D. There is a shift above the mean for four consecutive controls
B. Two controls in the same run are greater than 4s units apart
Rationale:
The R:4s rule is applied to two control levels within the same run. The rule is violated when the algebraic difference between them (level 1 – level 2) exceeds 4s.
The R:4s rule detects random error (error due to poor precision).
- Error always in one direction:
A. Random error
B. Systematic error
B. Systematic error
Rationale:
Random error: error varies from sample to sample
- Which of the following terms refers to deviation from the true value caused by indeterminate errors inherent in every laboratory measurement?
A. Random error
B. Standard error of the mean
C. Parametric analysis
D. Nonparametric analysis
A. Random error
Rationale:
Random errors are deviations from the true value caused by unavoidable errors inherent in laboratory measurements.
The standard error of the mean is a statistical concept reflecting sampling variation. It is the standard deviation of the entire population.
Parametric statistics refer to a Gaussian (normal) distribution of data.
Nonparametric statistics are more general and require no assumptions.
- A trend in QC results is most likely caused by:
A. Deterioration of the reagent
B. Miscalibration of the instrument
C. Improper dilution of standards
D. Electronic noise
A. Deterioration of the reagent
- Which of the following plots is best for detecting all types of QC errors?
A. Levy–Jennings
B. Tonks–Youden
C. Cusum
D. Linear regression
A. Levy–Jennings
Rationale:
The Levy–Jennings plot is a graph of all QC results with concentration plotted on the y axis and run number on the x axis.
- Which of the following plots is best for comparison of precision and accuracy among laboratories?
A. Levy–Jennings
B. Tonks–Youden
C. Cusum
D. Linear regression
B. Tonks–Youden
Rationale:
The Tonks–Youden plot is used for interlaboratory comparison of monthly means. The method mean for level 1 is at the center of the y axis and mean for level 2 at the center of the x axis. Lines are drawn from the means of both levels across the graph, dividing it into four equal quadrants. If a laboratory’s monthly means both plot in the lower left or upper right, then systematic error (SE) exists in its method.
- Which plot will give the earliest indication of a shift or trend?
A. Levy–Jennings
B. Tonks–Youden
C. Cusum
D. Linear regression
C. Cusum
Rationale:
Cusum points are the algebraic sum of the difference between each QC result and the mean. The y axis is the sum of differences and the x axis is the run number. The center of the y axis is 0. Because QC results follow a random distribution, the points should distribute about the zero line. Results are out of control when the slope exceeds 45° or a decision limit (e.g., ±2.7s) is exceeded.
- Which of the following terms refers to the closeness with which the measured value agrees with the true value
A. Random error
B. Precision
C. Accuracy
D. Variance
C. Accuracy
Rationale:
The accuracy of an analytical result is the closeness with which the measured value agrees with the true value.
Precision is reproducibility.
Accuracy and precision are independent, but it is the goal of the clinical laboratory to design methods that are both precise and accurate.
- Relatively easy to measure and maintain:
A. Accuracy
B. Precision
C. Sensitivity
D. Specificity
B. Precision
Rationale:
Accuracy is easy to define but difficult to establish and maintain.
Precision is relatively easy to measure and maintain.
- Which of the following describes the ability of an analytical method to maintain both accuracy and precision over an extended period of time?
A. Reliability
B. Validity
C. Probability
D. Sensitivity
A. Reliability
Rationale:
The reliability of an analytical procedure is its ability to maintain accuracy and precision over an extended period of time during which supplies, equipment, and personnel in the laboratory may change. It is often used interchangeably with the term “consistency.” It is the goal of every clinical laboratory to produce reliable results.
- Which of the following statistical tests is used to compare the means of two methods?
A. Student’s t test
B. F distribution
C. Correlation coefficient (r)
D. Linear regression analysis
A. Student’s t test
Rationale:
T TEST: COMPARES MEANS, ACCURACY
F TEST: COMPARES S.D., PRECISION
Student’s t test is the ratio of mean difference to the standard error of the mean difference (bias/random error) and tests for a significant difference in means.
The F test is the ratio of variances and determines if one method is significantly less precise.
- In a person with normal glucose metabolism, the blood glucose level usually increases rapidly after carbohydrates are ingested but returns to a normal level after:
A. 30 minutes
B. 45 minutes
C. 60 minutes (1 hour)
D. 120 minutes (2 hours)
D. 120 minutes (2 hours)
- Glucose measurements can be ____ mg/dL erroneously higher by reducing methods than by more accurate enzymatic methods.
A. 1 to 5 mg/dL
B. 5 to 15 mg/dL
C. 20 to 25 mg/dL
D. 30 to 35 mg/dL
B. 5 to 15 mg/dL
Rationale:
Reduction methods also measure other reducing sugars.
- Select the enzyme that is most specific for beta D-glucose:
A. Glucose oxidase
B. Glucose-6-phosphate dehydrogenase
C. Hexokinase
D. Phosphohexose isomerase
A. Glucose oxidase
Rationale:
Glucose oxidase is the most specific enzyme reacting with only beta D-glucose. Glucose oxidase converts beta D-glucose to gluconic acid.
Mutarotase may be added to the reaction to facilitate the conversion of alpha-D-glucose to beta-D-glucose.
- Select the coupling enzyme used in the hexokinase method for glucose:
A. Glucose dehydrogenase
B. Glucose-6-phosphatase
C. Glucose-6-phosphate dehydrogenase
D. Peroxidase
C. Glucose-6-phosphate dehydrogenase (G6PD)
Rationale:
The hexokinase method is considered more accurate than the glucose oxidase methods because the coupling reaction using glucose-6-phosphate dehydrogenase is highly specific; therefore, it has less interference than the coupled glucose oxidase procedure.
- Which of the following is a potential source of error in the HEXOKINASE METHOD?
A. Galactosemia
B. Hemolysis
C. Sample collected in fluoride
D. Ascorbic acid
B. Hemolysis
Rationale:
Hemolyzed samples require a serum blank correction (subtraction of the reaction rate with hexokinase omitted from the reagent).
- Gross hemolysis and extremely elevated bilirubin may cause ______ in HEXOKINASE RESULTS.
A. False increase
B. False decrease
C. No effect
D. Variable
B. False decrease
Rationale:
Generally accepted as the reference method, hexokinase method is not affected by ascorbic acid or uric acid. Gross hemolysis and extremely elevated bilirubin may cause a false decrease in results.
- VERY LOW OR UNDETECTABLE C-peptide:
A. Type 1 diabetes mellitus
B. Type 2 diabetes mellitus
A. Type 1 diabetes mellitus
- DETECTABLE C-peptide:
A. Type 1 diabetes mellitus
B. Type 2 diabetes mellitus
B. Type 2 diabetes mellitus
- A 62-year-old patient presents to the physician with report of increased thirst and increased urination, particularly at night. The physician requests a series of tests over the next few days. The following data are received:
RANDOM GLUCOSE: 186 mg/dL
FASTING GLUCOSE: 114 mg/dL
2-HOUR OGTT: 153 mg/dL
HbA1c: 5.9%
Which of the following conclusions may be made regarding these data?
A. Data represents normal glucose status
B. Data represents an impaired glucose status
C. Data represents the presence of insulinoma
D. Data represents diagnosis of diabetes
B. Data represents an impaired glucose status
Rationale:
Resting plasma glucose
● NORMAL <140
● IMPAIRED 140 – 199
● DIAGNOSTIC ≥ 200 mg/dL
Fasting plasma glucose
● NORMAL <100
● IMPAIRED 100 – 125
● DIAGNOSTIC ≥ 126 mg/dL
2-hour OGTT
● NORMAL <140
● IMPAIRED 140 – 199
● DIAGNOSTIC ≥ 200 mg/dL
HbA1c
● NORMAL <5.7 %
● IMPAIRED 5.7 – 6.4%
● DIAGNOSTIC ≥ 6.5%
- What is the recommended cutoff value for adequate control of blood glucose in diabetics as measured by glycated hemoglobin?
A. 5%
B. 6.5%
C. 9.5%
D. 11%
B. 6.5%
Rationale:
If the result is 6.5% or more, the treatment plan should be adjusted to achieve a lower level, and the test performed every 3 months until control is improved.
- A factor, other than average plasma glucose values, that can affect the HbA1c level is:
A. Serum ketone bodies
B. Red blood cell life span
C. Ascorbic acid intake
D. Increased triglyceride levels
B. Red blood cell life span
Rationale:
Two factors determine the glycosylated hemoglobin levels: the average glucose concentration and the red blood cell life span.
If the red blood cell life span is decreased because of another disease state such as hemoglobinopathies, the hemoglobin will have less time to become glycosylated and the glycosylated hemoglobin level will be lower.
- Most widely used to assess SHORT-TERM (3 to 6 weeks) glycemic control:
A. Glycosylated hemoglobin (HbA1c)
B. Fructosamine
B. Fructosamine
Rationale:
FRUCTOSAMINE: GLYCOSYLATED ALBUMIN
Fructosamine assays are the most widely used to assess short-term (3 to 6 weeks) glycemic control because the average half-life of the proteins is 2–3 weeks.
- Formation of glucose-6-phosphate from noncarbohydrate sources:
A. Glycolysis
B. Gluconeogenesis
C. Glycogenolysis
D. Glycogenesis
B. Gluconeogenesis
Rationale:
● Glycolysis: Metabolism of glucose molecule to pyruvate or lactate for production of energy
● Gluconeogenesis: Formation of glucose-6-phosphate from noncarbohydrate sources
● Glycogenolysis: Breakdown of glycogen to glucose for use as energy
● Glycogenesis: Conversion of glucose to glycogen for storage
● Lipogenesis: Conversion of carbohydrates to fatty acids
● Lipolysis: Decomposition of fat
- HYPERGLYCEMIC FACTOR produced by the pancreas is:
A. Epinephrine
B. Glucagon
C. Growth hormone
D. Insulin
B. Glucagon
Rationale:
Glucagon is the primary hormone responsible for increasing glucose levels. It is synthesized by the ALPHA-cells of islets of Langerhans in the pancreas and released during stress and fasting states.
ACTION OF GLUCAGON
● Increases glycogenolysis: glycogen → glucose
● Increases gluconeogenesis: fatty acids → acetyl-CoA → ketone, proteins → amino acids
- HYPOGLYCEMIC FACTOR produced by the pancreas is:
A. Epinephrine
B. Glucagon
C. Growth hormone
D. Insulin
D. Insulin
Rationale:
Insulin is the only hormone that decreases glucose levels and can be referred to as a hypoglycemic agent.
Insulin is the primary hormone responsible for the entry of glucose into the cell. It is synthesized by the beta cells of islets of Langerhans in the pancreas.
It decreases plasma glucose levels by increasing the transport entry of glucose in muscle and adipose tissue by way of nonspecific receptors. It also regulates glucose by increasing glycogenesis, lipogenesis, and glycolysis and inhibiting glycogenolysis.
ACTION OF INSULIN
● Increases glycogenesis and glycolysis: glucose → glycogen → pyruvate → acetyl-CoA
● Increases lipogenesis
● Decreases glycogenolysis
- What would an individual with CUSHING SYNDROME tend to exhibit?
A. Hyperglycemia
B. Hypoglycemia
C. Normal blood glucose level
D. Decreased 2-hour postprandial glucose
A. Hyperglycemia
Rationale:
Hypercortisolism = Cushing’s Syndrome = Thyroid Problem
CORTISOL INCREASES BLOOD GLUCOSE.
- Symptoms of HYPOGLYCEMIA usually occur when blood glucose has fallen below ___ mg/dL.
A. Below 50 mg/dL
E. Below 60 mg/dL
F. Below 70 mg/dL
G. Below 80 mg/dL
A. Below 50 mg/dL
- Beta cell destruction, usually leading to absolute insulin deficiency:
A. Type 1 DM
B. Type 2 DM
A. Type 1 DM
- May range from predominantly insulin resistance with relative insulin deficiency to a predominantly secretory defect with insulin resistance:
A. Type 1 DM
B. Type 2 DM
B. Type 2 DM
- Usual dose of LACTOSE in the oral lactose tolerance test is:
A. 25 grams
B. 50 grams
C. 75 grams
D. 100 grams
B. 50 grams
Rationale:
ORAL LACTOSE TOLERANCE TESTS, measuring the increase in plasma glucose or galactose following ingestion of lactose, have been used to diagnose lactase deficiency.
- Which of the following is the most common application of IMMUNOELECTROPHORESIS (IEP)?
A. Identification of the absence of a normal serum protein
B. Structural abnormalities of proteins
C. Screening for circulating immune complexes
D. Diagnosis of monoclonal gammopathies
D. Diagnosis of monoclonal gammopathies
- PRE-ECLAMPSIA, also referred to as TOXEMIA OF PREGNANCY is marked by specific symptoms including:
A. Water retention (with swelling particularly in the feet, legs, and hands)
B. High blood pressure
C. Protein in the urine
D. All of these
D. All of these
- A sensitive, although not specific indicator of damage to the kidneys:
A. Urea
B. Creatinine
C. Proteinuria
D. Cystatin C
C. Proteinuria
- At pH 8.6, proteins are _________ charged and migrate toward the _________.
A. Negatively, anode
B. Positively, cathode
C. Positively, anode
D. Negatively, cathode
A. Negatively, anode
- For albumin assay, absorbance at 630 nm is less likely to be affected by bilirubin or hemoglobin in the sample. Which dye gives a much greater absorbance change at 630 nm than it would at 500 nm?
A. HABA (Hydroxyazobenzene-benzoic acid)
B. BCG (Bromcresol green)
B. BCG (Bromcresol green)
- Which of the following dyes is the MOST SPECIFIC for measurement of albumin?
A. Bromcresol green (BCG)
B. Bromcresol purple (BCP)
C. Tetrabromosulfophthalein
D. Tetrabromphenol blue
B. Bromcresol purple (BCP)
Rationale:
BCP is more specific for albumin than BCG.
- In what condition would an increased level of serum albumin be expected?
A. Malnutrition
B. Acute inflammation
C. Dehydration
D. Renal disease
C. Dehydration
- ARTIFACTUAL INCREASE in albumin concentration:
A. Prolonged tourniquet application
B. Dehydration
C. Nephrotic syndrome
D. Inflammation
A. Prolonged tourniquet application
- Identification of which of the following is useful in early stages of glomerular dysfunction?
A. Microalbuminuria
B. Ketonuria
C. Hematuria
D. Urinary light chains
A. Microalbuminuria
- Most abundant amino acid in the body:
A. Glutamine
B. Lysine
C. Phenylalanine
D. Tyrosine
A. Glutamine
Rationale:
Glutamine is the most abundant amino acid in the body, being involved in more metabolic processes than any other amino acid.
- Precursor of the adrenal hormones epinephrine, norepinephrine, and dopamine and the thyroid hormones, including thyroxine:
A. Glutamine
B. Lysine
C. Phenylalanine
D. Tyrosine
D. Tyrosine
Rationale:
Tyrosine is a precursor of the adrenal hormones epinephrine, norepinephrine, and dopamine and the thyroid hormones, including thyroxine.
It is important in overall metabolism, aiding in the functions of the adrenal, thyroid, and pituitary glands.
- The plasma protein mainly responsible for maintaining colloidal osmotic pressure in vivo is:
A. Albumin
B. Hemoglobin
C. Fibrinogen
D. Alpha2-macroglobulin
A. Albumin
Rationale:
Albumin is responsible for nearly 80% of the colloid osmotic pressure (COP) of the intravascular fluid, which maintains the appropriate fluid balance in the tissue.
- Sensitive marker of poor nutritional status:
A. Prealbumin
B. Fibrinogen
C. Gc-globulin
D. Orosomucoid
A. Prealbumin
Rationale:
A low prealbumin level is a sensitive marker of poor nutritional status.
When a diet is deficient in protein, hepatic synthesis of proteins is reduced, with the resulting decrease in the level of the proteins originating in the liver, including prealbumin, albumin, and β-globulins. Because prealbumin has a short half-life of approximately 2 days, it decreases more rapidly than do other proteins.
- Nutritional assessment with poor protein-caloric status is associated with:
A. A decreased level of prealbumin
B. A low level of γ-globulins
C. An elevated ceruloplasmin concentration
D. An increased level of α1-fetoprotein
A. A decreased level of prealbumin
- Retinol (vitamin A) binding protein:
A. Albumin
B. Alpha1-antitrypsin
C. Fibronectin
D. Prealbumin
D. Prealbumin
Rationale:
Prealbumin is the transport protein for thyroxine and triiodothyronine (thyroid hormones); it also binds with retinol-binding protein to form a complex that transports retinol (vitamin A) and is rich in tryptophan.
- Which of the following conditions is the result of a LOW ALPHA1-ANTITRYPSIN LEVEL?
A. Asthma
B. Emphysema
C. Pulmonary hypertension
D. Sarcoidosis
B. Emphysema
Pulmonary Emphysema
Juvenile Hepatic Cirrhosis
- All are conditions associated with an elevated AFP, EXCEPT:
A. Neural tube defects
B. Spina bifida
C. Anencephaly
D. Down syndrome
D. Down syndrome
Rationale:
Conditions associated with an elevated AFP level include spina bifida, neural tube defects, abdominal wall defects, anencephaly (absence of the major portion of the brain), and general fetal distress.
Low levels of maternal AFP indicate an increased risk for Down syndrome (Trisomy 21) and trisomy 18 (Edward Syndrome), while it is increased in the presence of twins and neural tube defects.
- In nephrotic syndrome, the levels of serum ______ may increase as much as 10 times because its large size aids in its retention.
A. Alpha2-macroglobulin
B. Ceruloplasmin
C. Orosomucoid
D. Transferrin
A. Alpha2-macroglobulin
Rationale:
In nephrosis, the levels of serum α2-macroglobulin may increase as much as 10 times because its large size aids in its retention. The protein is also increased in diabetes and liver disease.
Use of contraceptive medications and pregnancy increase the serum levels by 20%.
- Orosomucoid:
A. Alpha1-antitrypsin
B. Alpha1-chymotrypsin
C. Alpha1-fetoprotein
D. Alpha1-acid glycoprotein
D. Alpha1-acid glycoprotein
Rationale:
α1-Acid Glycoprotein (Orosomucoid)
α1-Acid glycoprotein (AAG), a major plasma glycoprotein, is negatively charged even in acid solutions, a fact that gave it its name. This protein is produced by the liver and is an acute-phase reactant.
- Which of the following is a low-weight protein that is found on the cell surfaces of nucleated cells?
A. Alpha2-macroglobulin
B. Beta2-microglobulin
C. C-reactive protein
D. Ceruloplasmin
B. Beta2-microglobulin
Rationale:
β2-microglobulin is a single polypeptide chain that is the light chain component of human leukocyte antigens (HLAs). It is found on the surface of nucleated cells and is notably present on lymphocytes. Increased plasma levels of β2-microglobulin are associated with renal failure, lymphocytosis, rheumatoid arthritis, and systemic lupus erythematosus.
- Variants demonstrate a wide variety of cellular interactions, including roles in cell adhesion, tissue differentiation, growth, and wound healing:
A. Beta-trace protein
B. Cystatin C
C. Fibronectin
D. Troponin
C. Fibronectin
- A glycoprotein used to help predict the short-term risk of PREMATURE DELIVERY:
A. Adiponectin
B. Alpha-fetoprotein
C. Amyloid
D. Fetal fibronectin
D. Fetal fibronectin
- BIOCHEMICAL MARKER OF BONE RESORPTION that can be detected in serum and urine:
A. Beta-trace protein
B. Crosslinked C-telopeptides
C. Fibronectin
D. Troponin
B. Crosslinked C-telopeptides
Rationale:
Cross-linked C-telopeptides (CTXs) are proteolytic fragments of collagen I formed during bone resorption (turnover).
CTX is a biochemical marker of bone resorption that can be detected in serum and urine.
- An accurate marker of CSF leakage:
A. Beta-trace protein
B. Crosslinked C-telopeptides (CTX)
C. Fibronectin
D. Troponin
A. Beta-trace protein
Rationale:
β-Trace protein (BTP; synonym prostaglandin D synthase) is a 168–amino acid, low-molecular-mass protein in the lipocalin protein family. Recently, it was verified that BTP was established as an accurate marker of CSF leakage. It has also been reported recently as a potential marker in detecting impaired renal function, although no more sensitive than cystatin C.
- Recently proposed new marker for the early assessment of changes to the glomerular filtration rate:
A. Adiponectin
B. Beta-trace protein
C. Cross-linked C-telopeptides (CTX)
D. Cystatin C
D. Cystatin C
- Supplemental tests to help differentiate a diagnosis of ALZHEIMER DISEASE from other forms of dementia:
A. Amyloid β42 (Aβ42) and Tau protein
B. Crosslinked C-telopeptides (CTX)
C. Fibronectin
D. Troponin
A. Amyloid β42 (Aβ42) and Tau protein
Rationale:
In a symptomatic patient, low Aβ42 along with high Tau reflects an increased likelihood of Alzheimer disease, but it does not mean that the person definitely has Alzheimer disease. If a patient does not have abnormal levels of these proteins, then the dementia is more likely due to a cause other than Alzheimer disease.
- As a cardiac biomarker, this protein has been used in conjunction with troponin to help diagnose or rule out a heart attack:
A. Brain natriuretic peptide (BNP)
B. Cross-linked C-telopeptides (CTX)
C. Cystatin C
D. Myoglobin
D. Myoglobin
Rationale:
As a cardiac biomarker, myoglobin has been used in conjunction with troponin to help diagnose or rule out a heart attack. When striated muscle is damaged, myoglobin is released, elevating the blood levels.
In an AMI, this increase is seen within 2 to 3 hours of onset and reaches peak concentration in 8 to 12 hours. Myoglobin is a small molecule freely filtered by the kidneys, allowing levels to return to normal in 18 to 30 hours after the AMI.
Because of the speed of appearance and clearance of myoglobin, it is also a useful marker for monitoring the success or failure of reperfusion.
- “Gold standard” in the diagnosis of acute coronary syndrome (ACS):
A. Brain natriuretic peptide (BNP)
B. Cross-linked c-telopeptides
C. Myoglobin
D. Troponin
D. Troponin
Rationale:
ACUTE CORONARY SYNDROME/MYOCARDIAL INFARCTION
Cardiac troponin (cTn) represents a complex of regulatory proteins that include troponin I (cTnI) and troponin T (cTnT) that are specific to heart muscle.
cTnI and cTnT are the “gold standard” in the diagnosis of acute coronary syndrome (ACS). cTn should be measured in all patients presenting with symptoms suggestive of ACS, in conjunction with physical examination and ECG.
- Which test, if elevated, would PROVIDE INFORMATION about risk for developing coronary artery disease?
A. CK-MB
B. hs-CRP
C. Myoglobin
D. Troponin
B. hs-CRP
Rationale:
HIGH-SENSITIVITY CRP
Considered a GOOD PREDICTOR TEST for assessing cardiovascular risk
CK-MB, TROPONIN AND MYOGLOBIN
Used to assess if a myocardial infarction HAS OCCURRED
- If elevated, which laboratory test would support a diagnosis of CONGESTIVE HEART FAILURE?
A. Albumin cobalt binding
B. B-type natriuretic peptide
C. Homocysteine
D. Troponin
B. B-type natriuretic peptide
Rationale:
B-type (brain) natriuretic peptide (BNP) is used to determine if physical symptoms are related to congestive heart failure.
- Which two tests detect swelling of the ventricles that occurs in congestive heart failure?
A. BNP and electrocardiogram
B. BNP and echocardiogram
C. Troponin T and electrocardiogram
D. Troponin I and echocardiogram
A. BNP and electrocardiogram
Rationale:
DIAGNOSIS OF CONGESTIVE HEART FAILURE:
Until recently, this condition was diagnosed strictly on the basis of symptomatology and/or as a result of procedures such as echocardiography, but more recently a biomarker for this condition is the brain form or B-type natriuretic peptide (BNP), which has been approved as a definitive test for this condition and appears to be an excellent marker for early heart failure.
- Which of the following laboratory tests is a marker for ISCHEMIC HEART DISEASE?
A. Albumin cobalt binding
B. CK-MB isoforms
C. Free fatty acid binding protein
D. Myosin light chain
A. Albumin cobalt binding
Rationale:
Albumin cobalt binding is a test that measures ischemia-modified albumin, which is a marker for ischemic heart disease.
- The turbid, or milky, appearance of serum after fat ingestion is termed postprandial lipemia, which is caused by the presence of what substance?
A. Bilirubin
B. Cholesterol
C. Chylomicron
D. Phospholipid
C. Chylomicron
Rationale:
These chylomicrons enter the blood through the lymphatic system, where they impart a turbid appearance to serum
- When the plasma appears OPAQUE AND MILKY, the triglyceride level is probably:
A. Less than 100 mg/dL
B. Less than 200 mg/dL
C. Greater than 300 mg/dL
D. Greater than 600 mg/dL
D. Greater than 600 mg/dL
Rationale:
The appearance of the plasma or serum can be observed and noted after a minimum 12-hour fast.
If the plasma is clear, the triglyceride level is probably less than 200 mg/dL. When the plasma appears hazy or turbid, the triglyceride level has increased to greater than 300 mg/dL, and if the specimen appears opaque and milky (lipemic, from chylomicrons), the triglyceride level is probably greater than 600 mg/dL.
Note:
- Clear plasma: TAG < 200 mg/dL
- Hazy or turbid plasma: TAG > 300 mg/dL
- Opaque or milky plasma: TAG > 600 mg/dL
- Which of the following tests would most likely be included in a routine lipid profile?
A. Total cholesterol, triglyceride, fatty acid, chylomicron
B. Total cholesterol, triglyceride, HDL cholesterol, phospholipid
C. Triglyceride, HDL cholesterol, LDL cholesterol, chylomicron
D. Total cholesterol, triglyceride, HDL cholesterol, LDL cholesterol
D. Total cholesterol, triglyceride, HDL cholesterol, LDL cholesterol
- To produce reliable results, when should blood specimens for lipid studies be drawn?
A. Immediately after eating
B. Anytime during the day
C. In the fasting state, approximately 2 to 4 hours after eating
D. In the fasting state, approximately 12 hours after eating
D. In the fasting state, approximately 12 hours after eating
Rationale:
Blood specimens for lipid studies should be drawn in the fasting state at least 9 to 12 hours after eating. Although fat ingestion only slightly affects cholesterol levels, the triglyceride results are greatly affected. Triglycerides peak at about 4 to 6 hours after a meal, and these exogenous lipids should be cleared from the plasma before analysis.
- Which of the following lipid tests is LEAST affected by the fasting status of the patient?
A. Cholesterol
B. Triglyceride
C. Fatty acid
D. Lipoprotein
A. Cholesterol
Rationale:
Total cholesterol screenings are commonly performed on nonfasting individuals.
Total cholesterol is only slightly affected by the fasting status of the individual, whereas triglycerides, fatty acids, and lipoproteins are greatly affected.
- The kinetic methods for quantifying serum triglyceride employ enzymatic hydrolysis. The hydrolysis of triglyceride may be accomplished by what enzyme?
A. Amylase
B. Leucine aminopeptidase
C. Lactate dehydrogenase
D. Lipase
Rationale:
It is first necessary to hydrolyze the triglycerides to free fatty acids and glycerol. This hydrolysis step is catalyzed by the enzyme lipase.
The glycerol is then free to react in the enzyme-coupled reaction system that includes glycerokinase, pyruvate kinase, and lactate dehydrogenase or in the enzyme-coupled system that includes glycerokinase, glycerophosphate oxidase, and peroxidase.
- The largest and the least dense of the lipoprotein particles:
A. LDL
B. HDL
C. VLDL
D. Chylomicrons
D. Chylomicrons
Rationale:
Chylomicrons, which contain apo B-48, are the largest and the least dense of the lipoprotein particles.
HDL, the smallest and most dense lipoprotein particle, is synthesized by both the liver and the intestine.
- The smallest and most dense lipoprotein particle:
A. LDL
B. HDL
C. VLDL
D. Chylomicrons
B. HDL
- An abnormal lipoprotein present in patients with biliary cirrhosis or cholestasis:
A. LDL
B. B-VLDL
C. Lp(a)
D. LpX
D. LpX
Rationale:
Lipoprotein X is an abnormal lipoprotein present in patients with biliary cirrhosis or cholestasis and in patients with mutations in lecithin:cholesterol acyltransferase (LCAT), the enzyme that esterifies cholesterol.
- Exogenous triglycerides are transported in the plasma in what form?
A. Phospholipids
B. Cholestryl esters
C. Chylomicrons
D. Free fatty acids
C. Chylomicrons
Rationale:
From the epithelial cells, the chylomicrons are released into the lymphatic system, which transports chylomicrons to the blood. The chylomicrons may then carry the triglycerides to adipose tissue for storage, to organs for catabolism, or to the liver for incorporation of the triglycerides into very-low-density lipoproteins (VLDLs). Chylomicrons are normally cleared from plasma within 6 hours after a meal.
- Select the lipoprotein fraction that carries most of the endogenous triglycerides.
A. VLDL
B. HDL
C. LDL
D. Chylomicrons
A. VLDL
Rationale:
VLDL transports the majority of endogenous triglycerides, while the triglycerides of chylomicrons are derived entirely from dietary absorption.