Cardiac SPECT and PET Instrumentation Flashcards
Higher system sensitivity implies higher image signal-to-noise and more efficient acquisition but with decreased spatial resolution. Okay
Depends on collimation parameters
Anger camera
Lead collimators –> large crystal surface (NaI(Tl) crystal) –> PMT –> discrimination circuit
Solid state detectors
Use cadmium-zinc telluride (CZT)
And cesium-iodide (CsI)
High resolution collimators vs “general-purpose” or “all-purpose” collimators
Longer hole length
Smaller hole diameter
Thinner septa
Filtered backprojection reconstruction (FBP)
Utilizes ramp filters that has the effect of minimizing the “rays that project outward from each point in the transverse image
Iterative image reconstruction
Use a stepwise approach
Crystals used for PET
BGO (bismuth germanate)
More recently, LSO (lutetium orthosilicate)