Calibration Chain Flashcards
What is basic outline of calibration chain at NPL?
- Primary standard is graphite calorimeter, measures dose to graphite. Primary standard calibrates 3 NPL reference chambers in graphite phantom
- NPL reference chamber calibrations converted from D_graphite to D_water
- NPL calibrate hospital secondary standards against NPL reference chambers in water
How is dose measured in graphite?
Don’t measure temperature rise, maintain temperature and substitute electrical heating for radiation heating
D = VIt/m
Why do graphite calorimeters need to be thermally isolated?
Thermal diffusivity in graphite much larger than water, heat in graphite distribute itself in volume very quickly, too quickly to measure.
Why do NPL use graphite?
Reduce uncertainties
Bigger temperature rise from graphite
Why is graphite calorimeter not used to measure against secondary standards?
Too time consuming to use graphite calorimeter for everything. Reference chambers calibrated once a year against calorimeter.
What method do NPL use to calibrate user secondary standard with reference chambers?
Substitution method
What beams are used to calibrate graphite calorimeter against reference chambers?
60Co and a range of x-ray qualities between 4 and 25MV
What correction factors are used to correct raw dose for the measured chamber reading?
k_elec
k_TP
k_h
k_ion
k_pol
k_vol
What is k_elec correcting for?
Charge calibration coefficient for electrometer, which is calibrated by NPL separately, can be single factor or calculated by user
What is k_TP correcting for?
Temperature and pressure, account for the difference in mass of air in the chamber active volume
What is k_h correcting for?
Humidity correction. Can be taken as unity in 20-70% relative humidity, outside of this range NPL can advise
What is k_ion correcting for?
Losses because of ion recombination at finite collection voltages. Correction essential in pulsed radiation beams and can be substantial in FFF.
What is k_pol correcting for?
Polarity effects, which is negligible for 2611 chamber and only significant for small volume chambers. NPL don’t correct for polarity so long as it is constant during calibration process.
What is k_vol correcting for?
Beam non-uniformity, accounts for volume averaging and perturbation of secondary electron fluence in non-uniform beam profile such as FFF. For 2611 can be unity.
Which correction factors are usually unity
k_h
k_pol
k_vol