Beam Restriciting devices Flashcards
Why do we need to employ restricting devices?
- to reduce patient dose
- reduce scatter radiation
- achieve optimal radiographs within reasonable KVP and mAs ranges
What are the two principle factors that the amount of scatter produced?
- patient size
2. field size
As field size increases, so does__________
scatter radiaton
Reducing the beam, reduces scatter which also results:
in lower optical density, occasionally technical factors must be increased
Limit the size of the x-ray field to only anatomic structure of interest
beam restrictors
Reason for beam restrictors
- only the tissue being examined should be exposed
2. larger x-ray fields increasing patient exposure, scatter radiation and reduce image contrast
what are types of beam restricting devices?
Aperature diaphragm
cones and cylinders
variable aperature diaphragm
Aperture diaphragm is
-lead lined metal diaphragm attached to the x-ray tube head
Aperture diaphragm projects the image:
1 cm smaller on all side that the size of the image receptor, leaving a border around edges of the exposed film
Cones and cylinders are:
extended metal structure that restrict the beam to the required field size
Cones & cylinders are appropriate for
skull and sinus work
Cones & cylinders improve
image contrast
Most common beam restricting device
variable - aperture collimator
variable - aperture collimator restricts:
patient dose by restriction the volume of tissue irradiated, and decreasing scatter radiation
Positive beam limitation (PBL) mandiated in _____ and removed in _____
1974;1994
Penumbra
- Fuzzy border; image blur that results from the size of the focal spot; geometric artificially
How can penumbra be minimized
by using correct SID
Off focus radiation absorbed by
the collimator shutters made of @ 3mm lead
Off focus radiation increases
image blur
Off focus radiation definition
the xrays that are produced when the projectile electrons stray and interact at positions on the anode other than the focal spot
Reduce penumbra along the periphery of the beam due to greater distance fomr the focal spot
bottom shutters
Reduce the amount of off-focus radiation reaching the film by absorbing this radiation before it exits the tube
Upper shutters
The light field is provided by mounting on the path of the x-ray beam at a
40-45 degree angle
Light field
- The light field is provided by mounting on the path of the x-ray beam at a 40-45 degree angle
- A light source is placed opposite the mirror and the light is projected through the collimator
- the light source and x-ray source should be equidistant from each other to ensure their light fields are the same size
lead blocker
- lead-impregnated rubber that into various sizes and shapes
- absorbs scatter radiation
Lead masks
lead that is cut to a particular size or shape for a specific patient and exam
-cerebral angiograms, radiation therapy