Apex Unit 6 Equipment & Monitors Flashcards
Identify the components of the low pressure system in the anesthesia machine. (Select 2.) Common gas outlet Oxygen flush valve Flowmeter tubes Cylinder pressure regulator
Flowmeter tubes
Common gas outlet
The anesthesia machine can be divided into three pressure systems: high, intermediate, and low.
The high pressure system contains the cylinder pressure regulator.
The intermediate pressure system contains the oxygen flush valve.
The low pressure system contains the flowmeter tubes and common gas outlet.
Match each component to its proper location in the SPDD model Supply Processing Delivery Disposal
Vaporizer
Cylinder
Scavenger
Circle system
Supply + Cylinder
Processing + Vaporizer
Delivery + Circle system
Disposal + Scavenger
The PISS system is used to:
filter and exchange air in the operating room.
diagram the pathway of gas through the anesthesia machine.
prevent the use of the wrong gas hose.
prevent the use of the wrong gas cylinder.
Prevent the use of the wrong gas cylinder
The pin index safety system (PISS) prevents the wrong cylinder from being attached to the anesthesia machine.
The diameter index safety system (DISS) prevents the wrong gas hose from being attached to the anesthesia machine.
The SPDD model describes the pathway of gas through the anesthesia machine.
Filtering and exchanging air in the operating room is not a function of the anesthesia machine.
The bourdon pressure gauge on an oxygen cylinder reads 500 psi. If the flow rate is 2 L/min, how long will this cylinder provide oxygen to the patient?
(Enter your answer as minutes and round to the nearest whole number)
75 - 87
Some books say a full oxygen e-cylinder contains 1900 psi, while others say 2000 psi. We accepted both values.
We’ll detail the calculation on the next page…
Identify the BEST practices for handling gas cylinders. (Select 2.)
Oiling cylinder valves every six months
Laying the cylinder on its side when changing the cylinder on the anesthesia machine
Placing two washers between the cylinder and the hanger yoke assembly
Removing the plastic cover on the port when installing the cylinder
Removing the plastic cover on the port when installing the cylinder
Laying the cylinder on its side when changing the cylinder on the anesthesia machine
Under normal circumstances, gas cylinders must be stored in the upright position AND secured. When changing the cylinder on the anesthesia machine, however, it is appropriate to temporarily place the old cylinder its side until it can be moved to its appropriate storage receptacle.
You should remove the plastic cover from the port before installing the cylinder. Failure to do so may obstruct gas flow when the cylinder is turned on.
The fire triad consists of an oxidizer, a fuel, and an igniter. Oiling the cylinder valve increases the risk of fire by combining oxygen or nitrous with the oil. Only a heat source is needed to complete the triad.
If there is a leak after you install a cylinder, do not be tempted to place more than one washer between the cylinder and the hanger yoke assembly. This may bypass the PISS and allow the wrong cylinder to be matched up with the wrong hanger yoke assembly.
The oxygen pressure failure device activates when the: (Select 2.)
oxygen pipeline crossover with nitrous oxide.
FiO2 in the inspiratory limb is less than 21 percent.
oxygen tank is exhausted.
oxygen pressure in the supply line is less than 20 psi.
Oxygen tank is exhausted
Oxygen pressure in the supply line is less than 20 psi
The oxygen pressure failure device monitors oxygen pressure (not concentration). It activates when oxygen pressure in the intermediate pressure system falls below 20 psi. By contrast, the oxygen analyzer monitors oxygen concentration in the low pressure system. It alarms when the FiO2 falls below 21 percent.
If pipeline pressure fails, and the auxiliary oxygen tank is open, the oxygen pressure failure device won’t activate until oxygen pressure (from the tank) is less than 20 psi. This explains why you don’t want to leave the O2 cylinder open if you aren’t using it.
The oxygen pressure failure device is NOT activated by a pipeline crossover or if a leak develops distal to the flowmeters.
The hypoxia prevention safety device on the anesthesia machine will:
limit the nitrous oxide flow to three times oxygen flow.
shut off the flow of nitrous oxide if the oxygen supply pressure is less than 30 psi.
alarm if the FiO2 is less than 21 percent.
will prevent a hypoxic mixture if a third gas is used.
Limit the nitrous oxide flow to three times oxygen flow
The hypoxia prevention safety device prevents you from accidently setting a hypoxic mixture with the flow control valves. It’s a pneumatic or mechanical device that limits the nitrous oxide flow to no more than 3 times the oxygen flow.
You are administering air 1L/min and oxygen 3 L/min. Calculate the fraction of inspired oxygen.
(Enter your number as a percentage)
80
A fresh gas flow of air 1 L/min and oxygen 3 L/min yields a FiO2 of 80%.
We’ll show you the math on the next page …
You are using an anesthesia machine that couples fresh gas flow to tidal volume and fully compensates for circuit compliance. Calculate the total tidal volume delivered to the patient.
Oxygen = 3 L/min Air = 1 L/min I:E = 1:2 Bellows = 500 mL Respiratory rate = 10 bpm
(Enter your answer in mL and as a whole number)
632 or 633
Well done! This is not an easy question.
The correct answer is 632 or 633 mL (depending if how you chose to complete this calculation).
A ventilator is programmed to deliver a tidal volume of 600 mL. If the breathing circuit compliance is 5 mL/cm H2O and the peak pressure is 25 cm H2O, what is the total tidal volume that is delivered to the patient?
500 mL
475 mL
450 mL
425 mL
475 mL
When the ventilator produces positive pressure inside the breathing circuit, some of this gas causes the circuit to expand. This quantity of gas does not reach the patient, and therefore does not contribute to the tidal volume that the patient receives.
In this example, 125 mL is lost to the compliance of the circuit. Since the tidal volume is set at 600 mL, the patient will only receive a tidal volume of 475 mL.
The isoflurane dial is set to two percent. What percent of fresh gas exiting the vaporizing chamber is saturated with isoflurane?
(Enter your answer as a percentage)
100
Modern variable bypass vaporizers split fresh gas into two parts.
Some fresh gas enters the vaporizing chamber and becomes 100% saturated with volatile agent. Indeed, this was the correct response to this question.
The rest of the fresh gas bypasses the vaporizing chamber and does not pick up anesthetic vapor.
Before leaving the vaporizer, these two fractions mix and determine the final anesthetic concentration exiting the vaporizer.
Choose the statement that BEST describes this vaporizer. (Select 2.)
Its output is increased inside a hyperbaric chamber.
It uses a flow-over design.
It is heated to 42 degrees.
It is pressurized to two atmospheres.
It is pressurized to two atmospheres
Its output is increased inside a hyperbaric chamber
The Tec 6 desflurane vaporizer injects anesthetic vapor into the fresh gas (it does not use a flow-over design). The chamber that contains the anesthetic agent is pressured to two atmospheres and heated to 39 (not 42) degrees C.
The vaporizer output varies inversely with elevation. Therefore, the vaporizer should be re-calibrated when it’s going to be used in high altitude locations.
Which device will be the FIRST to detect an oxygen pipeline crossover?
Fail-safe device
Oxygen analyzer
Proportioning system
Pulse oximeter
Oxygen analyzer
The oxygen analyzer will be the first monitor to detect an oxygen pipeline crossover.
The pulse oximeter would probably the be second monitor to detect this complication, however it’s unlikely that this complication would be high on your differential.
The oxygen pressure failure device (failsafe) and hypoxia prevention safety device (proportioning system) aren’t designed to detect an oxygen pipeline crossover.
Identify the MOST critical actions that should be carried out in the event of an oxygen pipeline crossover. (Select 2.)
Disconnect the pipeline supply.
Turn on the oxygen tank.
Ventilate with an Ambu with the auxiliary oxygen flowmeter.
Replace the oxygen analyzer.
Turn on the oxygen tank
Disconnect the pipeline supply
Turn ON the oxygen cylinder, and then disconnect the pipeline oxygen supply. This is a key step! If a crossover occurred, simply turning on the oxygen tank would not fix the problem. If an adequate oxygen pipeline pressure is present (regardless of the gas inside), it will prevent the oxygen tank from releasing its contents.
This is not the time to assume an equipment malfunction. Trust the oxygen analyzer, and do not attempt to fix it. This could waste precious time.
The auxiliary oxygen flowmeter on the anesthesia machine is supplied by the pipeline. If an oxygen crossover occurs, it will supply the wrong gas to the patient. This is why you should use an oxygen tank.
Select the true statements regarding the oxygen flush valve. (Select 2.)
Excessive use can lead to awareness.
It will deliver a continuous pressure of 35 – 75 psi.
The risk of barotrauma is minimized by ventilators with fresh gas decoupling.
It will cause the ventilator spill valve to close during inspiration.
Excessive use can lead to awareness
The risk of barotrauma is minimized by ventilators with fresh gas decoupling
Why where the other answer choices wrong?
The oxygen flush valve delivers a continuous oxygen flow of 35 – 75 L/min (not psi).
During inspiration, the ventilator drive gas closes the ventilator spill valve. Pressing the O2 flush valve does not affect this function.
Select the true statements about the pneumatic ventilator bellows. (Select 2.)
A leak in the bellows may cause the reading on the oxygen analyzer to increase.
A descending bellows cannot rise and fall with a circuit disconnect.
A hole in the bellows may cause barotrauma.
A descending bellows is made safer by fresh gas coupling.
A hole in the bellows may cause barotrauma.
A leak in the bellows may cause the reading on the oxygen analyzer to increase.
The pneumatic bellows is compressed by the ventilator drive gas. A leak in the bellows creates a direct line of communication between the ventilator drive gas and the breathing circuit. This can cause barotrauma.
If there is a bellows leak and oxygen is used as the ventilator drive gas, the FiO2 in the breathing circuit may increase.
A descending bellows may continue to rise and fall, even with a circuit disconnect. Fresh gas decoupling helps solve this problem.
The piston ventilator: (Select 2.)
preserves tank oxygen in the event of oxygen pipeline failure.
allows for more precise delivery of tidal volumes.
removes the risk of barotrauma.
relies on fresh gas flow coupling.
Allows for more precise delivery of tidal volumes
Preserves tank oxygen in the event of oxygen pipeline failure
The piston ventilator offers several advantages over a traditional pneumatic bellows ventilator.
The piston is compressed by an electric motor. Since this type of system doesn’t use oxygen as a drive gas, it won’t consume tank oxygen in the event of oxygen pipeline failure.
It allows for more precise delivery of tidal volumes (fresh gas is decoupled from the ventilator).
Barotrauma remains a risk, however.
Which statement regarding pressure control ventilation is true? (Select 2.)
The ventilator switches to expiration after a preset pressure is achieved.
Gas flow decelerates during inspiration.
Increased lung compliance will decrease tidal volume.
The risk of ventilator associated lung injury is decreased.
Gas flow decelerates during inspiration
The risk of ventilator associated lung injury is decreased
Why are the other answer choices wrong?
Because the peak pressure is fixed and the tidal volume is variable, an increase in lung compliance will increase (not decrease) tidal volume.
The ventilator achieves a peak pressure very early in the inspiratory cycle and holds it for a time determined by the I:E ratio. It does not cycle immediately after the peak pressure is achieved.
Which modes of mechanical ventilation are BEST suited for a laryngeal mask airway? (Select 2.)
Pressure support ventilation
Inverse ratio ventilation
Controlled mandatory ventilation
Synchronized intermittent mandatory ventilation
Synchronized intermittent mandatory ventilation
Pressure support ventilation
Although not always an accepted practice, mechanical ventilation is now considered a perfectly acceptable technique to use with an LMA. As you’ll see, some modes are better than others.
SIMV and PSV are commonly used with an LMA.
CMV and IMV are best for patients who don’t have a respiratory drive.
At what pH does the ethyl violet change to purple?
- 5
- 6
- 3
- 1
10.3
Ethyl violet is a dye that indicates when soda lime exhausts. It changes from colorless to purple when the pH falls below 10.3.
When compared to soda lime, what factor is increased with the use of calcium hydroxide lime (Amsorb Plus)?
Fire risk
Frequency of replacement
CO2 absorption capacity
Carbon monoxide
Frequency of replacement
Benefits of calcium hydroxide lime (Amsorb Plus):
No carbon monoxide production Very little or no compound A production Lower risk of fire when compared to soda lime Drawbacks of calcium hydroxide lime:
Lower CO2 absorption capacity
Requires more frequent replacement
Higher cost
What is the MOST common cause of low circuit pressure?
Circuit disconnect
Leak in corrugated tubing
Incompetent ventilator relief valve
Improper fitting of carbon dioxide absorbent
Circuit disconnect
Circuit disconnect (usually at the y-piece) is the most common cause of low circuit pressure. The second most common cause is a leak around the carbon dioxide absorbent.
During a general anesthetic with an endotracheal tube, the high peak pressure alarm sounds. After changing the bag selector switch from ventilator mode to bag mode, the peak inspiratory pressure returns to baseline. Which of the following is the MOST likely explanation for the rise in peak inspiratory pressure?
The endotracheal tube was kinked.
The patient experienced a bronchospasm.
The positive pressure relief valve on the scavenger failed.
The ventilator spill valve malfunctioned.
The ventilator spill valve malfunctioned
This question requires a solid understanding of how the ventilator and scavenger work together.
Changing from ventilator mode to bag mode would not remedy bronchospasm, a kinked endotracheal tube, or a faulty scavenger positive pressure relief valve.
While a patient is ventilating spontaneously with an endotracheal tube, you notice that a fresh gas flow of 10 L/min is required to fill the breathing bag and determine that the scavenger is malfunctioning. Which statement must be true?
There is an open scavenging system.
The negative pressure relief valve has failed.
There is a passive scavenging system.
The positive pressure relief valve has failed.
The negative pressure relief valve has failed
A scavenger can be active or passive – an active system uses suction, while a passive system relies on the positive pressure of fresh gas leaving the interface. Since the scavenger in this question is applying excess suction, it must be an active system.
A scavenger can be a closed or open system – a closed system uses valves, while an open system is open to the atmosphere. Since the scavenger in this question is applying excess suction, it must be a closed, active system.
When the negative pressure relief valve fails (closed, active system), it’s possible for the vacuum to remove gas from the breathing circuit.
When the positive pressure relief valve fails (closed, active system), fresh gas can accumulate inside the breathing circuit. This can cause barotrauma.
Which value exceeds OSHA’s recommended standards for occupational exposure to inhalation anesthetics in the operating room?
Sevoflurane 1.0 ppm + Nitrous oxide 15 ppm
Sevoflurane 1.8 ppm
Desflurane 0.8 ppm
Desflurane 0.3 ppm + Nitrous oxide 20 ppm
Sevoflurane 1.0 ppm + Nitrous oxide 15 ppm
OSHA recommends that anesthetic gas exposure should not exceed the following:
Nitrous oxide alone should be less than 25 ppm.
Halogenated agents alone should be less than 2 ppm.
Halogenated agents with nitrous oxide should be less than 0.5 ppm.
Of all the answers, only sevoflurane 1.0 ppm + nitrous oxide 15 ppm exceeds the maximum.
Match each breathing system with its classification.
Open + No rebreathing and no reservoir
Semi-open + No rebreathing with a reservoir
Semi-closed + Partial rebreathing with a reservoir
Closed + Complete rebreathing with a reservoir
While providing an anesthetic with a circle system, you observe this waveform. What is the BEST course of action at this time? (Select 2.)
Increase the fresh gas flow in excess of the patient’s minute ventilation.
Repair the expiratory valve.
Increase the patient’s minute ventilation.
Replace the carbon dioxide absorbent.
Replace the carbon dioxide absorbent.
Increase the fresh gas flow in excess of the patient’s minute ventilation.
This patient is rebreathing CO2. There are 2 ways to fix this. You can either replace the carbon dioxide absorbent or convert the breathing system to a semi-open system by increasing the FGF.
Increasing the minute ventilation does not prevent rebreathing. The patient will continue to inspire carbon dioxide, but now only at a faster rate.
This is not a problem with the expiratory valve. The trick to diagnosing unidirectional valve malfunction is to look at the beta angle during the inspiratory phase. If you aren’t familiar with this term, we have an image for you on the next slide.
The Bain circuit: (Select 2.)
is a modified Mapleson A.
delivers fresh gas through the inner tube.
is best for spontaneous but not controlled ventilation.
is tested with the Pethick test.
Is tested with the Pethick test
Delivers fresh gas through the inner tube
The Bain system uses a coaxial design, where:
Fresh gas is delivered to the patient through the inner tube.
Exhaled gas travels through the outer tube.
The inner tube is at risk for kinking and disconnection. The Pethick test should be performed as part of the preanesthetic checkout.
The Bain system is a modified Mapleson D (not A). It can be used for spontaneous as well as controlled ventilation.
Select the statements that MOST accurately describe pulse oximetry. (Select 2.)
Oxygenated hemoglobin absorbs light at 940 nm.
At the peak of the waveform, the ratio of arterial blood to venous blood is reduced.
It is based on the Doppler effect.
It is based on the Beer-Lambert law.
It is based on the Beer-Lambert law
Oxygenated hemoglobin absorbs light at 940 nm
Why are the other answer choices wrong?
The pulse oximeter utilizes the Beer-Lambert law (not the Doppler effect).
It emits two wavelengths of light: oxygenated blood better absorbs near-infrared light (940 nm) and deoxygenated blood better absorbs red light (660 nm).
The pulse oximeter looks at the ratio of light absorption during the peak of the waveform relative to the trough of the waveform. At the peak of the waveform, the ratio of arterial blood to venous blood is increased (not the other way around).
The pulse oximeter reads 80 percent. You estimate the PaO2 is approximately:
(Enter your answer as mmHg)
50
Know these numbers:
SpO2 90% = PaO2 60 mmHg
SpO2 80% = PaO2 50 mmHg
SpO2 70% = PaO2 40 mmHg
The pulse oximeter is a useful monitor of:
ventilation.
bronchial intubation.
vascular compression.
anemia.
Vascular compression
You need to know the limitations of your monitors. In the case of the pulse oximeter, it’s not a good monitor of ventilation, anemia, or bronchial intubation.
Of the answer choices presented, the pulse oximeter is best for detecting vascular compression. Think about innominate artery compression during mediastinoscopy.
Sometimes you’re going to look at the answer choices and not love any of them. Learn to reason through why the correct answer is correct, but more importantly, why the wrong answers are wrong.
Which condition is LEAST likely to affect the reliability of the pulse oximeter? Blue nail polish Jaundice Carboxyhemoglobin Left ventricular assist device
Jaundice
Why are the other answers wrong?
Unlike a healthy heart that delivers pulsatile flow to the body, a left ventricular assist device supplements the failing myocardium with non-pulsatile flow. This defeats the mechanism of pulse oximetry.
Carboxyhemoglobin absorbs the same wavelength as oxyhemoglobin. This causes the pulse oximeter to overestimate the degree of oxygen bound to hemoglobin.
Blue nail polish can cause a false reduction in SpO2.
What is the MOST common method of measuring exhaled gases inside the breathing circuit?
Mass spectrometry
Infrared absorption
Raman scattering
Piezoelectric crystals
Infrared absorption
Infrared absorption is the most common method of exhaled gas analysis in the operating room. Although the other answer choices are less commonly employed, you still need to understand them.
A blood pressure cuff that is too large:
requires less pressure to occlude the artery.
has a bladder width less than 40 percent of the circumference of the extremity.
falsely increases the blood pressure measurement.
increases the risk of radial neuropathy.
Requires less pressure to occlude the artery
An improperly sized NIBP cuff can produce inaccurate results.
The ideal bladder length equals 80 percent of the extremity circumference, while the ideal bladder width equals 40 percent of the extremity circumference.
A cuff that is too small overestimates SBP.
A cuff that is too large underestimates SBP.
Although rare, nerve injury due to NIBP measurement tends to affect the ulnar or median nerve.
When inserting a central line in the right internal jugular vein, how far should the catheter be advanced to achieve correct placement?
10 cm
15 cm
20 cm
25 cm
15 cm
The tip of the central venous catheter should reside at the junction of the vena cava and the right atrium. It should not enter the right atrium!
In this example, the correct distance is 15 cm from the skin to the junction of the VC and RA.
We’ll show you an easy way to calculate central line distances on the next page.
Central venous pressure is:
unchanged by a ventricular septal defect.
increased by PEEP.
falsely increased by placing the transducer above the zero point.
decreased by pericardial tamponade.
Increased by PEEP
PEEP increases pulmonary vascular resistance. This creates additional resistance against RV ejection, which can increase RVEDP and CVP.
Why were the other answers wrong?
A transducer above the zero point, underestimates (not overestimates) CVP.
Pericardial tamponade produces a compressive force around the heart. This reduces RA compliance and increases (not decreases) CVP.
A VSD typically increases RVEDV and CVP (they don’t stay the same).
Which conditions increase the amplitude of the a wave on the CVP waveform? (Select 2.)
Tricuspid stenosis
Atrial fibrillation
Diastolic dysfunction
Tricuspid regurgitation
Tricuspid stenosis
Diastolic dysfunction
If you understand the CVP waveform, then reasoning through this question becomes a whole lot easier!
The a wave correlates with atrial contraction. Therefore, if the a wave amplitude increases, it is due to a stenotic tricuspid valve or decreased compliance of the right ventricle.
With atrial fibrillation, there is no atrial kick. On the CVP waveform, the a wave is lost.
With tricuspid regurgitation, a portion of the RV volume travels through the incompetent tricuspid valve and re-enters the RA. On the CVP waveform, there will be a large v wave.
Which findings are observed when the tip of the pulmonary artery catheter enters the highlighted area? (Select 2.)
Dicrotic notch
Increased systolic blood pressure
Increased pulse pressure
Increased diastolic blood pressure
Dicrotic notch
Increased diastolic blood pressure
You should be able to explain the pressure changes as the tip of the PAC travels through the heart.
When the tip enters the pulmonary artery, the diastolic pressure increases and a dicrotic notch appears. We’ll walk you through all of it on the next page.
In which lung zone should the tip of the pulmonary artery catheter be placed?
(Enter your answer as a whole number)
Three
The tip of the PAC should be in zone III.
In this region, there is a continuous column of blood between the tip of the PAC and the left ventricle. This provides the most accurate estimation of LVEDP.
When does pulmonary artery occlusion pressure overestimate left ventricular end diastolic volume? (Select 2.) Aortic insufficiency PA catheter tip in West zone III PEEP Diastolic dysfunction
PEEP
Diastolic dysfunction
You should know when to trust your numbers, and more importantly, you should understand the conditions where your numbers may be inaccurate.
Anything that impairs the normal pressure gradient between the PAC tip and the LV can impact your interpretation of the PAOP.
PAOP overestimates LVEDV: PEEP and diastolic dysfunction
PAOP underestimates LVEDV: Aortic insufficiency
Which situation underestimates cardiac output obtained by the thermodilution method?
Right-to-left intracardiac shunt
Over warmed injectate
High injectate volume
Partially wedged pulmonary artery catheter
High injectate volume
When we measure cardiac output with the thermodilution method, we graph injectate temperature vs. time. Cardiac output is inversely proportional to the the area under the curve.
You must know the variables that can skew your results.
A high injectate volume underestimates CO, while a low injectate volume overestimates CO.
Injectate that is too warm and a partially wedged pulmonary artery catheter overestimate CO.
A right-to-left intracardiac shunt has an unpredictable effect on CO measurement.
Factors that increase mixed venous oxygen saturation include: (Select 2.)
sepsis.
thyroid storm.
anemia.
sodium nitroprusside toxicity.
Sodium nitroprusside toxicity
Sepsis
SvO2 is a function of four variables: cardiac output, arterial oxygen saturation, amount of hemoglobin, and oxygen consumption.
Cyanide toxicity (impaired O2 utilization) and sepsis (high cardiac output state + arterial admixture) increase SvO2. Anemia reduces oxygen delivery and decreases SvO2. Thyroid storm increases oxygen demand and decreases SvO2.
Preload responsiveness is expected to be present if a 250 mL fluid bolus increases the stroke volume in excess of:
(Enter your answer as a percentage)
10 percent
As a general rule, preload responsiveness is expected to be present if a 200 - 250 mL fluid bolus increases the stroke volume in excess of 10 percent.
Which pathway depolarizes the left atrium? Kent bundle Thorel tract Wenckebach tract Bachmann bundle
Bachmann bundle
There are three internodal tracts that travel from the SA node to the AV node:
Bachmann bundle (extends into the left atrium)
Wenckebach tract
Thorel tract
Kent’s bundle is a pathologic accessory pathway that is responsible for Wolff-Parkinson-White syndrome.
Match each phase of the ventricular action potential to its corresponding component on the EKG waveform.
Phase 0 + QRS complex
Phase 2 + QT interval
Phase 3 + T wave
There are five phases of the ventricular action potential: 0 = Rapid depolarization (QRS) 1 = Initial repolarization (QRS) 2 = Plateau phase (QT interval) 3 = Final repolarization (T wave) 4 = Resting phase (T → QRS)
Match each disease with the EKG abnormality that it is MOST likely to cause.
Pericarditis + PR interval depression
Hypokalemia + U wave
Intracranial hemorrhage + Peaked T wave
Wolff-Parkinson-White syndrome + Delta wave
Match each lead to the cardiac region that it monitors.
V1 + Septum
V3 + Anterior wall
aVF + Inferior wall
Lead I + Lateral wall
The mean electrical vector tends to point: (Select 2.)
towards areas of hypertrophy.
towards areas of myocardial infarction.
away from areas of hypertrophy.
away from areas of myocardial infarction.
Towards areas of hypertrophy
Away from areas of myocardial infarction
The mean electrical vector represents the average of all the electrical forces generated by the myocardium. The normal value is between -30 degrees and +90 degrees.
The mean electrical vector tends to point:
Towards areas of hypertrophy (there is more tissue to depolarize).
Away from areas of myocardial infarction (the vector must travel around these areas).
All of the following are effective for the treatment of atrial fibrillation EXCEPT:
metoprolol.
adenosine.
digoxin.
verapamil.
Adenosine
Adenosine is an endogenous nucleoside that slows conduction through the AV node. By stimulating the cardiac adenosine-1 receptor, adenosine activates K+ currents, which hyperpolarizes the cell membrane and reduces action potential duration.
It is efficacious for supraventricular tachycardia as well as WPW with a narrow QRS.
It is not efficacious for atrial fibrillation, atrial flutter, torsades des pointes, or ventricular tachycardia.
Match each antiarrhythmic agent with its drug class.
Lidocaine + Class I
Propranolol + Class II
Amiodarone + Class III
Verapamil + Class IV
Antiarrhythmic drugs can be divided into four classes:
Class I drugs inhibit fast sodium channels.
Class II drugs decrease the rate of depolarization.
Class III drugs inhibit potassium ion channels.
Class IV drugs inhibit slow calcium channels.
Wolff Parkinson White syndrome is associated with:
atrial-ventricular reentry.
SA nodal reentry.
atrial reentry.
ventricular reentry.
Atrial-ventricular reentry
Wolff-Parkinson-White syndrome occurs when an accessory pathway joins the atrium to the ventricle. This accessory pathway is called Kent’s bundle.
A patient with Wolff Parkinson White syndrome develops atrial fibrillation during surgery. Select the BEST treatment for this situation. (Select 2.)
Digoxin
Amiodarone
Procainamide
Verapamil
Procainamide
Amiodarone
This is a complex topic, so you may want to read the next page a few time
All of the following increase the likelihood of torsades de pointes in the patient with long QT syndrome EXCEPT:
furosemide.
metoprolol.
methadone.
hyperventilation.
Metoprolol
Patients with long QT syndrome are at risk for developing torsades de pointes. Beta-blocker therapy is associated with a reduction in sudden death in this patient population.
Patients with long QT syndrome should not receive drugs known to prolong the QT interval.
Methadone is the only narcotic known to increase the QT interval.
Furosemide can cause hypokalemia and hypomagnesemia. These metabolic disturbances can prolong the QT interval.
Hyperventilation shifts K+ into cells, decreases serum K+, and can prolong the QT interval.
Match the NBG pacemaker identification code to its designated function.
Position I + Chamber paced
Position II + Chamber sensed
Position III + Response to sensed event
Position IV + Programmability
A patient undergoing a bunionectomy has a VOO pacemaker with a rate of 80 bpm. During the procedure, there is failure to capture and the heart rate decreases to 50 beats per minute. Which of the following BEST explains why this complication occurred?
The patient was hyperthermic.
The EtCO2 was 20 mmHg.
An ultrasonic Harmonic scalpel was used.
The electrocautery setting was changed from “coagulation” to “cutting.”
The EtCO2 was 20 mmHg
The pacemaker failed to capture, because hypocarbia (which caused hypokalemia) made the myocardium more resistant to depolarization. The same electrical stimulus from the pacemaker was no longer sufficient to depolarize the heart. You’ll see the pacer spikes, but you will not see capture.
Why the other choices are wrong:
When compared to the surgical electrocautery, the use of an ultrasonic Harmonic scalpel decreases the chance of electromagnetic interference.
Changing from “coagulation” to “cutting” also reduces the risk of EMI.
Hypothermia (not hyper-) makes the myocardium more resistant to depolarization. As an aside, this is why hypothermia causes bradycardia.
Select the statement that BEST describes cerebral oximetry.
A > 25% change from baseline suggests a reduction in cerebral oxygenation.
It monitors arterial oxygen saturation in cerebral blood.
It monitors global cerebral oxygenation.
It is invasive.
A > 25% change from baseline suggests a reduction in cerebral oxygenation.
Cerebral oximetry is a noninvasive technique that utilizes near infrared spectroscopy (NIRS) to measure regional (not global) cerebral oxygenation.
It is based on the principle that ↓ cerebral oxygen delivery → ↑ cerebral oxygen extraction → ↓ venous hemoglobin saturation.
A > 25% change from baseline suggests a reduction in cerebral oxygenation.