Apex- Anesthesia Machine Flashcards
Where does the high pressure system begin and end?
intermitediate pressure system?
low pressure system?
Cylinders → Cylinder Regulators
Pipeline → Flowmeter VALVES
Flowmeter TUBES → common gas outlet
What tests the integrity of the low-pressure circuit from the flowmeter valves to the common gas outlet?
how is it done?
what would signify an issue?
low-pressure leak nest
attach a bulb to the CGO and create neg pressure
- must be performed with teh vaporizers off, and then again with each vaporizer turned on one at a time
- if the bulb inflates, theres a leak in the system
Components of the high-pressure system (4)
Components of the intermediate-pressure system (7)
Components of the low-pressure system (4)
Pipeline vs tank psi
Pipeline = 50
Tank = 45
T/F: when doing a low-pressure leak test, the FGF must be off
True
if you have a vent with a minimum FGF- you have to turn off the machine to do the test
when doing a low-pressure test, should you be on the bag or vent?
bag
low pressure tests from flowmeter VALVES to CGO
How often should the low-pressure test be performed?
before the first case of each day
lol
PISS vs DISS
PISS- pin index - prevents wrong cylinder
DISS- diameter index - prevents wrong pipeline
PISS configurations for Air
O2
Nitrous
Air 1,5
o2 2,5
nitrous 3,5
The O2 cylinder pressure drops from ~2,000psi to what psi upon entry into the intermediate system
why?
~45 psi
to ensure gas is preferentially puplled from the pipeline if a cylinder is left open
(pipeline ~ 50psi)
Why should there only be one washer between the hanger yoke assembly and the stem of the tank?
bc more than one may allow for the PISS system to be bypassed
Service pressure and max volume of:
O2
Air
N20
O2= 1900psi/660L
Air = 1900psi/625L
N20= 745psi/ 1590L
service pressure of nitrous oxide
745psi
1590L
maximum volume of nitrous oxide in an e-cylinder
1590L
745psi
why does nitrous oxide exist as a liquid inside the cylinder?
bc it’s critical temp is above room tempearure (~36.5)
-critical temp is the highest temp a gas can exist as a liquid
How many L of nitrous oxide gas is remaining in the tank when there is no more liquid present
what about 400psi
250L
400psi = 136L
t/f: the cylinder valve is the most delicate component of the cylinder
True
the silver part on top - this is why a cylinder needs to be stored in an upright position
bc if it falls down and goes boom, and the top silver piece breaks off (the cylinder valve) - the o2 tank becomes a missle
why do we need to stand up the cylinder tank in a locked postion when not in use? what wil lhappen if it falls?
if it falls, the cylinder valve (the top silver piece) can break off (it’s delicate) and then the o2 tank can become a missle
What makes a o2 tank MRI safe and how can you tell that it is?
-made of aluminum
-mainly silver with the top portion being the color of the gas it contains (green = o2)
*but the best way to determine the contents is to READ the label- not examine the color
What kind of safety relief device is made by woods metal
what’s in woods metal
when would it provide safety and how?
fusible plug
BLT with Cheese- Bismuth, Lead, Tin, Cadmium
in the event of a fire, it would melt, to slowly release o2 contents - instead of exploding
What’s in woods metal
Bismuth
Lead
Tin
Cadmium
BLT with Cheese
Weight of a full nitrous oxide tank vs empty
full = 20.7#
empty= 14.1#
oh come on
World Health Organization colors for Air, O2, and Nitrous
Air = black and white
O2 = white
Nitrous = blue
Does Carbon dioxide exist as a liquid or gas inside the cylinder
max volume?
service pressure?
liquid
1590L (same as nitrous)
838psi
T/F: if youre changing the cylinder on the andesthesia machine, it’s ok to lay the cylinder on its side temporarily
True
If you remove a cylinder from the anesthesia machine but don’t have a replacement, what do you have to do?
Insert a yoke plug
If you open the cylinder and hear a hissing sound- what does that indicate and what do you need to do?
leak
-tighten the connection
T/F: if there is a leak, add an additional washer between the cylinder and hanger yoke assembly
FALSE- dont do this - allows bypassing PISS
Which gases are oxidizers
fire triangle components?
nitrous oxide and oxygen
oxidizer, fuel, igniter
Gas cylinder should never be exposed to temps above what degrees (F + C)
130 degrees F
57 degrees C
why shouldn’t you oil the cylinder valve
bc it increases risk of fire
What would happen if you didn’t insert a yoke plug in an empty cylinder spot?
If the check valve were to fail & there is no cylinder or yoke plug - gas that should be going to the patient will exit the anesthesia machine
Match: American Society for Testing and Materials (ASTM), FDA, OSHA, DOT:
-sets the standards for the required components of the anesthesia machine
-sets standards for compressed gas cylinders
-sets standards for acceptable occupational exposure to volatile anesthetics
-creates standard machine checkout procedures
-sets the standards for the required components of the anesthesia machine - ASTM
-sets standards for compressed gas cylinders- DOT
-sets standards for acceptable occupational exposure to volatile anesthetics- OSHA
-creates standard machine checkout procedures- FDA
A cylinder must be tested how often?
it’s tested to how many times it’s service pressure?
every 5 years -10 years with a specia permit indicated by a 5-pointed star on the label
1.66x it’s service pressure
7 pieces of information required on the cylinder label according to US DOT
- Government agency (DOT)
- type of metal
- max filling pressure
- serial number
- manufactorer
- owner
- date of last inspection
What monitors for low oxygen pressure in the anestheisa machine
another name for it
3 instances it would alarm
Oxygen-Pressure Failure Device
Fail-safe device
- depleted o2 tank
- drop in pipeline pressure
- disconnected o2 hose
T/F: the failsafe device will alarm in the setting of an o2 pipeline crossover
False- the failsafe device monitors pressure - not o2 concentration
2 components of the failsafe device (o2 pressure failure device)
does it reside in the high, intermediate, or low pressure system?
- **a threshold alarm **that sounds when o2 pipeline pressure falls below 28-30psi
- A Pneumatic device that *reduces or stops the flow of nitrous oxide *when the pressure in the o2 pipeline falls *below 20psi *
Does the fail-safe device monitor pressure or concentration
pressure
What would you do if you suspect a pipeline cross over? 2 steps
- open o2 cylinder on machine
- disconnect o2 pipeline from machine
Which machines failsafe device stops nitrous commpletely when o2 psi < 20 vs proportionatlly :
drager vs GE Datex-Ohmeda)
Drager - proportional decrease
GE Datex-Ohmeda - all or nothing
How would you know if the o2 pressure failure device is working?
okay so the o2 pressure failure device alarms when psi <30
apex says to turn on o2 and nitrous flow
verify o2 cylinder is closed
then d/c the o2 pipeline
once o2 source is disconnected, watch the flow meeters and the nitrous oxide flow should stop before the o2 flow stops
retinruducing the o2 pipeline should result in both gases restored to previous flow rates
The hypoxia prevention safety device will:
A. Shut off the flow of nitrous oxide if the o2 supply pressure is < 30psi
B. limit the nitorus flow to 3x the o2 flow
C. Alarm if the Fio2 is < 21%
D. prevent a hypoxic mixture if a third gas is used
B
When you’re administering nitrous oxide, the proporitioning device ensure that the FiO2 never drops below what % ?
25%
The proportioning device maintsins a max of what ratio of nitrous to o2?
3:1
T/F: the proportioning device will prevent hypoxia with the administration of a third gas
False
ie) helium
T/F: the o2 pressure failure device and the hypoxia prevention safety device are not the same thing
True
pressure failure device = fail-safe device
hypoxia prevention safety device = proporitoning device
What is the oxygen-ratio monitor controller
monitors the ratio of o2 to n20 (drager machine)
automatically adjusts n20 to go no grater than 3:1 (pneumatic device)
What is another name for the proportion device?
Hypoxia prevention safety device
The internal diameter of the flow tubes are (narrowest/widest) at the base and (narrows/widens) with ascent
narrowest at the base
widens as it goes up
To monimize the risk of hypoxic mixture in the setting of a cracked flowmeter, the O2 flowmetere should be positionined where?
where is this location on US machines?
closest to the common gas outlet
Far right
T/F: flowmeters are calabritated for each specific gas
True
What is the space between the ball float and the wall of the flowmeter called
the annular space
T/F: glass flowmeters resists the buildup of static electricity
True
Should o2 be positioned first or last in the sequence of flowmeters
last/downstream
ie) closest to the CGO/pt
incase there is a crack in any of the upstream flowmeters - this prevents a hypoxic mixture
but a crack in the o2 flow meter can still produce a hypoxic mixture
At high FGF, gas flow is (turbulant/laminar) and depdendent on gas (density/viscostiy) - what law
At low FGF, gas flow is (turbulant/laminar) and depdent on gas (density/viscosity) - what law
High FGF- turbulent- density- grahams
Low FGF - laminar, viscosity, pousilles
What is the o2% delivered with 1L air and 3L o2
80%
[Air flow rate x 21) + (O2 flow rate x 100)] / Total flow rate
1 x 21 = 21 + 300 = 321/4 = 80.25
O2 nasal cannula %s:
1L-
2L-
3L-
4L-
5L-
6L-
1L- 24%
2L- 28%
3L - 32%
4L- 36%
5L - 40%
6L - 44%
RA = 21% - then just go to 24% and add by 4
% of O2 delivered with a nasal cannula can range from what
24-44%
Steps in figuring out tidal volume with FGF is coupled to it
- convert FGF from L/min to ml/min
- multiply FGF by I:E ratio
- Calculate the tv per breath (FGF/RR)
- Add volume set on vent to FGF during inspiration
- I:E ratio of 1:2 means over 1 minute, 20 seconds is spent in inspiration and 40 in expiration
- said another way, 1/3 spent in inspiration and 2/3 spent in exhalation
- so for #2- if I:E ratio is 1:2, your multipling by 1/3 or 0.33333
A vent if programmed to deliver a tidal volume of 600mls. If the breathing circuit compliance is 5ml/cm H20 and the peak pressure is 25cm H20, what is the total tidal volume delivered to the patient?
A. 425mls
B. 450mls
C. 475mls
D. 500mls
C. 475 mls
5ml/cmh20 x 25cm H20 = 125ml
600 - 125ml = 475mls
*when the vent produces postive pressure inside the breathing circuit, some of this gas causes the circuit to expand. This quantiy of gas doesnt reach the patient and doesnt contribute to the tidal volume the pt recieves
Change in volume/ Change in pressure = what?
compliance
The isoflurane dial is set to 2%. What % of fresh gas exiting the vaporizing chamber is saturated with isoflurane? Enter your answer as a percentage
100%
Modern variable bypass vaporizers split fresh gas into 2 parts:
- some fresh gas enters the vaporizing chamger and becomes 100% saturated with the volatile agent
- The rest of the fresh gas bypasses the vaporizing chamber and does not pick up the anesthetic vapor
Before leaving the vaporizer, these two fractions mix, and this determines the final anesthetic concentration exitin the vaporizer
What does it mean when a vaporizer is “variable bypass?”
it means when fresh gas enters the vaporizer, someo f it encounters the liquid anesthetic and the rest bypasses it.
You determine the splitting ratio (how much fresh gas contacts the liquid) by setting the concentration on the dial.
What does it mean when a vaporizer is “flow-over”
so as fresh gas enters the vaporizing chamber with the liquid anesthetic, it flows over a series of baffles and wicks.
The baffles and wicks increase surface area and turbulence, ensuring the fresh gas incide the vaporizing chamber becomes 100% saturated with the anesthetic agent
variable bypass vaporiziers:
-flow-over wick
-temperature compensated
-out-of-circuit
-agent specific
T/F: the temperature compensating valve on a vaporizer adjusts the ratio of the vaporizing chamber flow to bypass flow and guarantees a constant vaporizer output over a wide range of temperatures
True
variable bypass vaporiziers:
-flow-over wick
-temperature compensated
-out-of-circuit
-agent specific
T/F: the variable bypass vaproizer is positioned OUTSIDE of the breathing circuit
true
variable bypass vaporiziers:
-flow-over wick
-temperature compensated
-out-of-circuit
-agent specific
T/F: variable bypass vaporizers are agent specific
True - each vaporizer is calibrated to one anesthetic agent. Fillling a vaporizer with the incorrect anesthetic can lead to catastrophic errors in output
variable bypass vaporiziers:
-flow-over wick
-temperature compensated
-out-of-circuit
-agent specific