Anesthesia Machine Flashcards

1
Q

These are part of the (low/int/high system)?
Common gas outlet
Thorpe tubes
Vaporizers

A

Low pressure system

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

These are part of the (low/int/high system)?
Flowmeter valves
Oxygen flush valve
Oxygen pressure failure device
Pipeline inlets

A

Intermediate Pressure System

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

These are part of the (low/int/high system)?
Hanger yoke
Cylinder pressure regulator
Cylinder pressure gauge

A

High Pressure System

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

How is a low-pressure leak test performed?

A

This test is performed by attaching a bulb to the common gas outlet and creating negative pressure. Failure occurs if the bulb reinflates within 10 seconds.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

How much negative pressure should be created for a proper negative-pressure test?

A

– 65 cm H2O

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Should the vaporizer be on or off during a low-pressure leak test?

A

Both on and off. The vaporizers should be off at first, but then the test should be repeated with vaporizer turned on.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

List the 4 components of the SPDD model.

A

Supply
Processing
Delivery
Disposal

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

The PISS system is used to:

A

prevents the wrong cylinder from being attached to the anesthesia machine.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

The DISS system is used to:

A

prevents the wrong gas hose from being attached to the anesthesia machine.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

What is the pin configuration for oxygen

A

2, 5

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

What is the pin configuration for air

A

1, 5

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

What is the pin configuration for N2O

A

3, 5

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

What is the maximum volume of an O2 cylinder?

A

660 L

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

What is the maximum volume of an air cylinder?

A

625 L

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

What is the maximum volume of an N2O cylinder?

A

1, 590 L

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

What is the maximum pressure inside an O2 cylinder?

A

1,900 psi

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

What is the maximum pressure inside an air cylinder?

A

1,900 psi

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

What is the maximum pressure inside an N2O cylinder?

A

745 psi

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

Can an cylinder be laid on its side while being changed?

A

Yes, only when empty and being changed out

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

What agency sets the standards for the required components of the anesthesia machine.

A

American Society for Testing and Materials (ASTM)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

What agency developed the Anesthesia Machine Pre-Use Checkout procedures.

A

Food and Drug Administration (FDA)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

What agency sets the standards for acceptable occupational exposure to volatile anesthetics.

A

Occupational Safety and Health Administration (OSHA)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
23
Q

What agency sets the requirements for compressed gas cylinders.

A

United States Department of Transportation (DOT)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
24
Q

The oxygen pressure failure device activates when the (2)

A

oxygen pressure in the supply line < 20 psi and O2 tank exhaustion

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
25
Q

What is the purpose of the oxygen failsafe device?

A

To monitor for (and protect against) low oxygen pressure in the anesthesia machine.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
26
Q

The oxygen failsafe device will alert you to what 3 problems?

A

Depleted O2 tank
A drop in pipeline pressure
Disconnected O2 supply hose

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
27
Q

In what pressure system does the oxygen failsafe device reside?

A

Intermediate pressure system

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
28
Q

When does the oxygen failsafe device fail to detect a hypoxic mixture?
(2)

A

Oxygen-Pipeline Crossover
Flowmeter Leak

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
29
Q

What is another name for the oxygen failsafe device?

A

Oxygen pressure failure device

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
30
Q

The hypoxia prevention safety device on the anesthesia machine will:

A

limit the nitrous oxide flow to three times oxygen flow.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
31
Q

What is another name for the proportion device?

A

Hypoxia prevention safety device

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
32
Q

What is the minimum FiO2 permitted by the proportioning device?

A

25%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
33
Q

How does the proportioning device work?

A

It limits N2O flow to 3 times O2 flow (N2O max ~ 75%)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
34
Q

Which type of flowmeter float is read at the middle?

A

Ball

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
35
Q

How are the skirted, plumb bob, and nonrotating floats read?

A

at the top of the float.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
36
Q

What is the annular space?

A

The area between the float and sidewall of the flowmeter.
(ball float)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
37
Q

Describe the internal geometry of the flowmeter.

A

The flowmeter is narrowest at the base and widens along its ascent. This creates a “variable orifice” architecture.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
38
Q

What is the safest position for the oxygen flowmeter as it relates to other gas flowmeters?

A

Positioned on the far right (closest to the manifold outlet)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
39
Q

These things (increase/decrease tidal volume)
Increased I:E ratio
Increased bellows height
Decreased respiratory rate
Increased gas flow

A

Increases Tidal Volume

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
40
Q

These things (increase/decrease tidal volume)
Decreased gas flow
Decreased bellows height
Increased respiratory rate
Change the I:E ratio from 1:2 to 1:3

A

Decreases Tidal Volume

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
41
Q

How does circuit compliance affect the tidal volume delivered to the patient?

A

If it is not compensated for, the tidal volume will decrease. Some machines automatically compensate.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
42
Q

What should you do if the vaporizer is tipped over? (2)

A

Drain the vaporizer to remove all the liquid anesthetic.
Turn the dial to the highest setting and run a high FGF for 20–30 minutes.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
43
Q

What is the “T” setting on the vaporizer?

A

Setting the vaporizer to transport mode (T) eliminates the worry of tipping the vaporizer during transport or removal.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
44
Q

What is the most common cause of a vaporizer leak?

A

Loose filler cap

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
45
Q

Does a variable bypass vaporizer need to be calibrated for elevation?

A

No

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
46
Q

How many atmospheres is the Tec 6 desflurane vaporizer pressurized to?

A

2 atmospheres

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
47
Q

What temperature is the Tec 6 desflurane vaporizer heated to?

A

39 degrees C

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
48
Q

What vaporizer type varies inversely with elevation?

A

Tec 6

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
49
Q

This matches (variable bypass vaporizer/tec 6)?
Compensates for elevation
Has temperature compensation valve
Variable bypass
Flow over design

A

Variable Bypass Vaporizer

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
50
Q

This matches (variable bypass vaporizer/tec 6)?
Pressurized to 2 ATM
Requires electrical power
Dual circuit design
Does not compensate for elevation
Heated to 39 C

A

Datex-Ohmeda Tec 6

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
51
Q

Which device will be the FIRST to detect an oxygen pipeline crossover?

A

Oxygen analyzer

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
52
Q

Why will a pyrexic 70-kg patient become hypoxic with an oxygen flow rate at 250 mL/min in a closed-circuit system?

A

The VO2 for the average adult is 250 mL/min. Conditions such as fever, sepsis, and thyrotoxicosis will increase O2 consumption.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
53
Q

Normal oxygen consumption, VO2 for an adult

A

250 mL/min of O2

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
54
Q

The oxygen analyzer resides in what limb of the breathing circuit?

A

Inspiratory limb

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
55
Q

What are the MOST critical actions to perform during an oxygen pipeline crossover? (2)

A

Disconnect the pipeline supply.
Turn on the oxygen tank.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
56
Q

What 2 immediate actions must you take if the oxygen pipeline supply fails?

A

Turn ON the oxygen cylinder
Disconnect the pipeline oxygen supply.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
57
Q

What should be at the top of your differential when the O2 analyzer alarms?

A

O2 pipeline crossover until other causes can be ruled out.

58
Q

Is the risk of barotrauma minimized by ventilators with fresh gas decoupling

59
Q

Can excessive use of the O2 flush lead to awareness?

A

Yes due to dilution of anesthetic gasses

60
Q

What 2 functions does the drive gas on an a pneumatic ventilator serve?

A

It compresses the bellows
It opens and closes the ventilator spill valve

61
Q

What is the position of the ventilator spill valve during inspiration?

A

It’s closed. This ensures that the tidal volume goes to the patient and not to the scavenger.

62
Q

What is the minimum expiratory pressure needed to open the ventilator spill valve?

63
Q

Pressing the O2 flush exposes to breathing circuit what pressure?

A

50 psi (pipeline pressure)

64
Q

Pressing the O2 flush exposes to breathing circuit what flow rate?

A

35–75 L/min

65
Q

List 2 complications caused by overuse of the oxygen flush valve.

A

Barotrauma
Awareness

66
Q

A leak/hole in the bellows may cause what 2 things

A

Increased oxygen analyzer reading and barotrauma

67
Q

How does a leak in the bellows lead to barotrauma?

A

The ventilator drive gas compresses the pneumatic bellows. A leak in the bellows creates a direct line of communication between the ventilator drive gas and the breathing circuit.

68
Q

How does a leak in the bellows increase O2 analyzer readings?

A

If there is a bellows leak and oxygen is used as the ventilator drive gas, the FiO2 in the breathing circuit may increase.

69
Q

What is the problem with descending bellows during circuit disconnects?

A

A descending bellows may continue to rise and fall

70
Q

How does an ascending bellows move during inspiration?

71
Q

Will a descending bellows fill during a circuit disconnect?

A

Yes—it can fill with entrained room air.

72
Q

How will a bellows leak affect the inspired concentration of anesthetic agent?

A

It will dilute the anesthetic concentration (it goes down).

73
Q

2 distinct advantages of a piston ventilator

A

preserves oxygen tank in the event of oxygen pipeline failure.
allows for more precise delivery of tidal volumes.

74
Q

Are these things (true/false) of a piston ventilator?
Contains a positive pressure relief valve
Contains a negative pressure relief valve
Decouples FGF from Vt
Reservoir bag inflates during inspiration

75
Q

Are these things (true/false) of a piston ventilator?
Requires a drive gas
Supplies 3 cm H2O of intrinsic PEEP
Reservoir bag inflates during expiration

76
Q

Gas flow (accelerates/decelerates) during inspiration with pressure control ventilation

A

Decelerates

77
Q

What ventilator mode decreases the risk of ventilator-associated lung injury

A

Pressure control

78
Q

What parameters are fixed and what is variable in pressure control

A

Fixed: peak pressure
Variable: tidal volume

79
Q

This is (volume control/pressure control)?
Vt is fixed
Inspiratory flow rate is fixed
PIP is variable

A

Volume Controlled Ventilation

80
Q

This is (volume control/pressure control)?
Inspiratory flow rate is variable
PIP is fixed
Vt is variable

A

Pressure Controlled Ventilation

81
Q

Which 2 modes of mechanical ventilation are BEST suited for a laryngeal mask airway?

A

Synchronized intermittent mandatory ventilation.
Pressure support ventilation.

82
Q

Is this (true/false) of mechanical ventilation?
A risk of assist control is respiratory alkalosis
IRV requires a paralyzed and sedated patient
CMV is best suited for apneic patients
PSV-Pro is useful for a patient with an LMA

83
Q

Is this (true/false) of mechanical ventilation?
Pressure support provides backup for apnea
BiPAP requires you to set the Vt
SIMV increases the risk of patient ventilator asynchrony

84
Q

At what pH does the ethyl violet change to purple?

85
Q

What are 2 major problems that can occur with carbon dioxide absorbent?

A

The absorbent is exhausted.
The absorbent is desiccated.

86
Q

Why is silica added to soda lime?

A

It provides hardness and minimizes dust production.

87
Q

What is the ideal mesh size for CO2 absorbent?

A

4–8 mesh

88
Q

Does indicator dye alert you to CO2 absorbent desiccation?

A

No. It only tells you if the absorbent is exhausted.

89
Q

When compared to soda lime, what factor is increased with the use of calcium hydroxide lime?

A

Frequency of replacement

90
Q

3 benefits of calcium hydroxide lime (Amsorb Plus):

A

No carbon monoxide production.
Very little or no compound A production.
Lower risk of fire when compared to soda lime.

91
Q

3 Drawbacks of calcium hydroxide lime:

A

Lower CO2 absorption capacity.
Requires more frequent replacement.
Higher cost.

92
Q

Which component scavenger system determines if it’s an open or closed system?

A

The scavenging interface

93
Q

What type of scavenging system communicates to the atmosphere with pressure valves?

A

Closed system

94
Q

What 2 valves must be present on a closed interface - active scavenger?

A

Negative pressure relief valve
Positive pressure relief valve

95
Q

What scavenger valve is malfunctioned when a patient requires FGF to fill the breathing bag?

A

Negative pressure relief valve

96
Q

What is the MOST common cause of low circuit pressure?

A

Circuit disconnect

97
Q

What part of the circuit is the most common cause of low circuit pressure?

98
Q

Second most common cause of low circuit pressure

A

leak around the carbon dioxide absorbent.

99
Q

What is the most common source of preventable equipment related complications?

A

Circuit disconnect

100
Q

Recite 4 way to detect a circuit disconnect.

A

Pressure
Volume
EtCO2
Your vigilance

101
Q

What should you do if the breathing circuit doesn’t hold pressure in the middle of a procedure?

A

Ventilate the patient with a bag-valve-mask and oxygen tank. Provide TIVA if necessary.

102
Q

High peak pressure alarm sounds. After changing the bag selector switch from ventilator mode to bag mode, PIP returns to baseline. What causes the rise in pressure?

A

The ventilator spill valve malfunctioned.

103
Q

What are 4 CV consequences caused by excessive pressure in the breathing circuit?

A

Decreased venous return
Decreased cardiac output
Hypotension
Cardiovascular collapse

104
Q

Volatile anesthetics OSHA recommended exposure in ppm

A

less than 2 ppm.

105
Q

Nitrous oxide OSHA recommended exposure in ppm

A

less than 25 ppm

106
Q

VA with N2O, OSHA recommended exposure in ppm

A

less than 0.5 ppm

107
Q

Which 2 components of the anesthesia machine could be affected if the low-pressure leak test fails?

A

Flowmeter.
Vaporizer.

108
Q

The low-pressure leak test (negative pressure test) assesses

A

the low-pressure circuit’s integrity from the flowmeter valves to the common gas outlet

109
Q

Which 2 situations would produce negative pressure inside the breathing circuit?

A

An orogastric tube placed in the trachea
Malfunction of a closed-system scavenger

110
Q

A negative deflection is observed on the airway pressure waveform when the mechanical ventilator is activated. Which 2 mode of ventilation is the patient MOST likely receiving?

A

Pressure-support ventilation protect.
Biphasic positive airway pressure.

111
Q

A negative deflection just before the breath indicates

A

a patient triggered breath

112
Q

Inverse ratio ventilation and controlled mandatory ventilation require

A

an apneic patient

113
Q

What is the MOST likely cause of a lower than expected concentration of sevoflurane inside the breathing circuit?

A

Bellows leak

114
Q

Which agency regulates health care worker exposure to anesthetic gases in the operating room?

A

Occupational Safety and Health Administration

115
Q

According to the US Department of Transportation, the gas cylinder label must include (2)

A

date of last inspection.
serial number.

116
Q

The transfer tubing joins which components of the scavenger? (2)

A

APL valve.
Scavenger interface.

117
Q

The gas disposal assembly tubing connects

A

scavenging interface to the gas disposal assembly.

118
Q

The pressure in which type of cylinder will remain constant regardless of how much liquid is present inside the tank?

A

Carbon dioxide

119
Q

The oxygen analyzer measures

A

O2 concentration inside the breathing circuit, making it the only safety device on the anesthesia machine that will alert you to an oxygen-pipeline crossover.

120
Q

The failsafe device and hypoxia prevention safety devices help to prevent

A

a hypoxic mixture

121
Q

What is the MINIMUM recommended pressure when performing a high-pressure leak test on the anesthesia machine? (Enter your answer in cm H2O)

122
Q

What is the MAXIMUM flow rate delivered by the oxygen flush valve?

123
Q

Soda lime that has changed color from purple to white occurs during

A

periods of nonuse

124
Q

Does the soda lime color change from white to purple indicate desiccation?

125
Q

What is the indicator dye used in soda lime?

A

Ethylene violet

126
Q

Identify the final step in the soda lime reaction.

A

Na2CO3 + Ca(OH)2 ​ —> ​ CaCO3 + 2NaOH

127
Q

What is the water content of soda lime?

128
Q

Which components creates the GREATEST resistance to airflow?

A

Endotracheal tube

129
Q

The combination of sevoflurane and Baralyme increases the risk of:

A

breathing circuit fire.

130
Q

flow tubes are tapered at

131
Q

flow tubes are wide at

132
Q

According to the World Health Organization, what color should an oxygen tank be painted?

133
Q

Which 2 actions in the formulation of calcium hydroxide lime reduce compound A production?

A

Removal of NaOH.
Addition of Ca(OH)2.

134
Q

DISS stands for

A

diameter index safety system

135
Q

PISS stands for

A

pin index safety system

136
Q

The SPDD model is used to understand

A

the pathway of gas through the anesthesia machine.

137
Q

​When replacing an oxygen tank on the anesthesia machine, using more than one washer between the cylinder and the yoke can lead to a/an:

A

inadvertent attachment of a nitrous oxide cylinder.

138
Q

Which vaporizers are approved for the use of desflurane? ​ (2)

A

GE-Datex-Ohmeda Tec 6. Drager D Vapor

139
Q

best method to detect a vaporizer leak.

A

​The low pressure circuit leak test

140
Q

When the ambient temperature is increased, the bi-metallic strip in a variable bypass vaporizer directs (more/less) fresh gas to (bypass/vaporizing) chamber (2)

A

More fresh gas to the bypass chamber.
Less fresh gas to the vaporizing chamber.