Anesthesia Machine Flashcards

1
Q

What is the fail-safe valve?

A

Prevents delivery of hypoxic gas mixtures.
Based on pressure from oxygen supplies.
Turns off nitrous oxide if in use. Can be proportional or on/off

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

How does the fail-safe valve work?

A

In the event of a failure of the oxygen supply (oxygen delivery line pressure decreases below 30 psi), proportionally decreases the flow of all other gases.

Prevents unrecognized exhaustion of oxygen cylinder or if hospital supply line failure.

Does not prevent delivery of 100% N2O when O2 flow is zero. The provider will also need to monitor the oxygen analyzer to detect hypoxic gas mixtures.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Diameter Index Safety System (DISS)

A

Fittings that are noninterchangeable to prevent misconnections of pipeline gases (from the wall source).

The pipeline gases come into the wall with color-coded outlets.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Hospital pipeline pressure.

A

Approximately 50 psi

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Pin Indexed Safety System (PISS)

A

Prevents misconnection of the cylinders to the hanger yoke assembly. Makes it impossible to attached the wrong cylinder.

Medical air has pins 1 and 5
N20 pins 3 and 5
O2 pins 2 and 5

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Oxygen E-cylinder

A
Green
Pins at 2,5
Full tank pressure: 1900-2000 psi
Full tank volume: 625-660L
Volume proportional to psi/pressure, ex: psi 1000 = 1/2 full or ~330L
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Nitrous Oxide (N2O) E-cylinder

A

Blue
Pins: 3,5
Pressure: 745-750 psi until 75% empty or all liquid nitrogen has vaporized
Determine tank volume via weight
Once pressure beings to fall below 750 psi less 25% left or ~400L
Total volume of a full tank: 1590

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Why does frost accumulated on E-cylinder valves and tanks?

A

The expansion of compressed O2 and vaporization of liquified nitrous oxide absorbs heat. Heat is extracted from the cylinder and atmosphere. This causes condensation from the water in our atmosphere.

There will never be internal icing because gases in the tank are free of water vapor

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

How do flowmeters measure flow?

A

Measurement based on the principle that flow past a resistance is proportional to the pressure.

Proportionality between pressure and flow is determined by the shape of the tube (resistance) and the physical properties (density and viscosity) of the gas.

Calibrated to a specific gas, and not interchangeable. Color coded.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Flowmeters design

A

Gas enters at the bottom of vertically positioned and tapered glass flow tube.
Gas as it enters raises the bobbin or ball shaped float
Gas is off when gravity is balanced by the decrease in pressure.

*The cross sectional area increases upward from the site of gas entry.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

The oxygen flush valve

A

Bypasses flowmeters. Direct communication between high and low pressure circuits.
Delivers 35-75L/min
Pressure of 50 psi or hospital pipeline
Can cause barotrauma. Avoid using during mechanical inspiration. No relief valve.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

High Pressure system pressure value

A

Related to tank pressure. Oxygen tanks at full are 2200psi. N2O is 745psi

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Components of the hight pressure system.

A
Cylinder
Hanger yoke assembly
PISS- pin indexed safety system
Pressure gauge, indicator for cylinders
Ends at the pressure reducing valve
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Intermediate pressure system pressure?

A

35-55psi

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Intermediate pressure system components

A
Pipeline inlet connects for wall source
Pipeline pressure gauges
piping
gas power outlet
master switch
O2 pressure failure devices 
O2 flush valve
Secondary stage reducing device if GE
Flow control valves
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Low pressure system pressure

A

About 16 psi, safe for patients. 14-26psi depending on machine

17
Q

Low pressure system includes

A
Flow indicators
unidirectional valve
pressure relief device
vaporizers and mounts
common gas outlet
18
Q

What type of valves are the check valves

A

Free floating, unidirectional

Back flow closes the valve

19
Q

Where are check valves placed

A

after the the hanger yolk
after the pipeline inlets
after vaporizers
and in the breathing circuit