Anesthesia Equipment Chapter 12 Flashcards
1
Q
A hole or tear in the bellows my result in?
A
Hyperventilation
Hyperoxia
Barotrauma
Patient awareness
2
Q
Excessive airway pressure maybe caused by?
A
- Activating the oxygen flush during inspiration
- A hole in the bellow housing
- A stuck spill valve
3
Q
A standing bellows
A
- expands during exhalation
- makes it easy to detect disconnections
4
Q
when a ventilator with a piston is used
A
- Air entrainment may occur
- The reservoir bag may expand and contract with mechanical ventilation
5
Q
Regarding pressure control ventilation
A
- Flow is highest at the beginning of inspiration
- Tidal volume is not constant
- It is often used with supraglotic devices
6
Q
Regarding SIMV
A
- The ventilator has a secondary source of gas flow for spontaneous breathing
- It is often used to wean patients off mechanical ventilation
- A mandatory tidal volume and minimum mechanical ventilation rate must be selected.
7
Q
Regarding PSV
A
- Most ventilators have a backup or apneic rate.
2. The pressure waveform has a square, flat top shape