Analytical Techniques Flashcards
A tech is performing an enzyme assay at the 340 nm using a visible range spectrophotometer. After setting the wavelength and adjusting the readout to zero %T with the light path blocked, a civet with DI water is inserted. With the light path fully open and the 100% T control at a max, the instrument readout will not rise above 90% T. What is the most appropriate first course of action? (3)
- change the light source
- change the monochromator
- clean cuvett
a red colored solution would show highest transmittance at
620
a red colored solution would show highest absorbance at
450
a photomultiplier tube may be described by all of following except:
A. rapid response time
B. must be shielded from stay light
C. cannot be used with a chopper
D. amplifies current significantly.
C
what does this describe: cannot be used with a chopper
sudder
which type of photodetector has linear array. that allows it to respond to a specific wavelength resulting in complete UV/visible spectrum analysis
photodiode array
what isolates light within a narrow region of the spectrum
monochromator
what is the difference between a fluorimeter and a spectrometer
fluorimeter use filters to select the wavelengths of incident light. and spectrometers use monochrometers
what is the major difference in the arrangement of major components between spectrometers and atomic absorption spectrometer