Analytical Methods Flashcards
measurement of light transmitted by a solution
SPECTROPHOTOMETER
Light or Radianr Source of Spectrophotometer
POLYCHROMATIC LIGHT
most commonly used light source
TUNGSTEN IODIDE LAMP
As the lamp envelope darkens with age, the instrument readout will not rise above the 90% T, instead 100% T. What should the MTOD do?
Replace the Sourcr Lamp
Light Source used in AAS
Hollow Cathode Lamp
TRUE or FALSE
The transmittance should be 100% at all times
TRUE
What are the components of a spectrophotometer?
Light Source
Entrance Slit
Monochromator
Exit Slit
Sample Cuvette
Photodetector
Readout Device
What is the most common cause of Loss of Linearity?
STRAY LIGHT
Most commin cause of stray light?
Second Order Spectra:
- Deteriorated Optics
- Light Dispersed by Darkened Envelope
- Extraneous Room Light
What are the type of filter is best in measuring stray light?
Sharp Cut Off Filters
What is used in verifying absorbance accuracy and linearity?
Neutral Density Filters and Dichromate Solution
Isolates specific wavelength of light.
Monochromator
Most commonly used monochromator
Diffraction Gratings
Short wavelengths are refracted more than long wavelengths, composed of sodium chloride, and considered to produce the purest light
Prism
Half band width of a monochromator is defined by the range of wavelength passed at ______ maximum transmittance.
50%
This is used to check the wavelength accuracy
Didymium Holmium Oxide Filter
It controls the bandpass or bandwidth, control the light reaching the sample cuvette
Exit Slit
If the wavelength is 540 nm and the bandpass allows 10 nm, what is the wavelength range?
535 nm to 545 nm
It is also called absorption cell, analyticsl cell or sample cell which holds the sample/solution.
Cuvette
Most commonly used type of cuvette
Alumina Silica
Unknown substances is DIRECTLY PROPORTIONAL to the absorbed light, and INVERSELY PROPORTIONAL to the TRANSMITTED LIGHT
Beer’s Law
Beer’s Law Absorbance Formula
2 - log % T
Most common type of photomultiplier and the most sensitive one
Photomultiplier Tube
simplest detector and do not require external voltage source
Barrier Layer Cell
Photocell
Photovoltaic Cell
simplest type of spectrophotometer and it measures one at a time
Single Beam Spectrophotometer
permit automatic correction of sample and uses two photodetectors
Double Beam Spectrophotometer
Double Beam Spectrophotometer that uses two photodetectors
In Space
A solution consisting of all the components of a reacting EXCEPT THE ANALYTE.
Blank
A blank contains seruk but without reagent to complete the assay
Blanking Technique
Contains the same reagent for the assay and corrects for absorbance caused by the color of the reagents.
Reagent Blank
Sample blank is used to subtract the intrinsic absorbance of the sample usually caused by hemolysis, icterus, turbidity, or drug interference, it is performed by substituting the ______ for reagent.
SALINE
In reflectance photometry, 100% reflectance is set with an ________ film called the white reference.
OPAQUE
In Automated Reagent Strip Reader, concentration of analyte/sample is ____________ to reflectance (light reflected).
INVERSELY PROPORTIONAL
Minimizes error caused by source lamp
Fluctuations
What principle is used in Bilirubinometer and Automated Reagent Strip Reader?
Reflectometry or Reflectance Photometry
What is the hottest flame color?
BLUE
Measures the light emitted by a single aton burned in a flame.
Flame Emission Photometry
In FES, Sodium emits ____color.
Yellow
In FES, Potassium emits ____color.
Violet/Purple
In FES, Lithium/Rubidium emits ____color.
Red
In FES, Magnesium/Copper emits ____color.
Blue
Type of spectrophotometer that uses an atomizer
Atomic Absorption Spectrophotometer
Element is not excited, but merely dissociated from its chemical bond and place un an ionized, unexcited, ground state
AAS
AAS Light Source
Hollow Cathode Lamp
Used by AAS to convert ions to atoms,
Atomizer/Graphite Furnace