A single point calibration is performed more frequently on ABG instruments in order to:

A

Correct electrode and instrument drift

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Which of the following will shift the O2 dissociation curve to the left?

A

Alkalosis

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Which electrode on an ABG instrument measures the amount of flow between an anode and cathode?

A

pO2 electrode

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

How many Ion Selective Electrodes does an ABG instrument have? Which electrode(s) is/are the ISE?

A

Two: pH electrode and pCO2 electrode

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Which of the following causes a shift to the left in an oxygen dissociation curve?

A

Hyperventilation

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

A difference between the voltage of the measuring and the reference electrode is set up across a selective membrane on the measuring electrode. This describes the principle of:

A

Potentiometry

How well did you know this?
1
Not at all
2
3
4
5
Perfectly