8 Mechanical Testing Flashcards
What is medical device testing?
Th eprocess of demonstrating that a device will reliably and safely perform in use
Name 5 types of mechanical test
Drop test
Temperature test
Accelerated aging test
Slip test
Humidity test
Describe drop testing
Test to ensure the device is resistance to the stress caused by a fall
Performed by dropping a device from a specific height onto a 50mm thick piece of hardware on a concrete surface
Describe humidity testing
Done to conform the operation of the device following a time period in a humidity chamber
Performed to prove the device meets safety standards after being exposed to humidity
The duration in the chamber varies but can be up to seven days
Describe temperature testing
Performed to ensure the device can operate in a range of temperatures
Typical operating temperature is between 0-40 degrees celcius
Temperature in a hospital is fixed, it can vary at home
Transportation testing is done at a wider range of values
Describe slip testing
Relevent to floor standing devices
Test is performed on a ramp with a specific angle and specific material