7 - How do we know we are doing it right? Flashcards
1
Q
What is the path of the instruments through the LDU?
A
- dirty set down
- wash sink
- ultrasonic cleaner
- rinse sink
- washer disinfector
- inspection
- steriliser
- dispatch
2
Q
What guidance is used in the decontamination of instruments?
A
- Scottish Health Technical Memorandum (SHTM)
- SHTM 01-01 C
- SHTM 01-01 D
- SDCEP
3
Q
What legislation is used in the decontamination of instruments?
A
- Heath and safety at work act 1974
- medical device regulation (MDR 2017/745)
4
Q
What is the framework for staff development?
A
Level 1 : new trainee
Level 2 : support worker
Level 3 : senior healthcare assistant / technician
Level 4 : assistant or associate practitioner
Level 5 : practitioner
5
Q
What is the role of guidance?
A
- identifies equipment required for each stage
- define process requirements
- identifies baseline testing
6
Q
What are the quarterly tests for an AWD?
A
- verification of calibration
- thermometric tests for disinfection
- cleaning efficacy test (using test soil)
- doors and door interlocks
- chemical dosage (ensure parameter remains within tolerance)
- calibration, limits and functions, including alarm of independent monitoring system
7
Q
Who should carry out the quarterly tests?
A
Competent person (ie practitioner)