What is the purpose of 500A Basic Oxygen Analyzer?
Continuously monitor the content of a clean, dry and non-corrosive sample gas stream by measuring its relatively magnetic susceptibility
500A Internal Case Temp
122 degrees Fahrenheit automatically controlled by power supply and temp control
500A Power Requirements
115 VAC
48-62 Hz
300 Volts Amps
500A max flow
1000 cc/m
500A Component Case
-drip proof, corrosion resistant, cast aluminum
500A Heat Indicator Function
- pulse/flash when interior casing temperatures stabilizes at 122 degrees Fahrenheit
- temperature stabilization takes approximately 4 hours
500A Gain Check Switch Function
- O2 position
- should not change the output but more than 1.5% Oxygen
500A Meter Test Switch Function
Should read +188.88
Calibration Span
- O2
- right side
- Trim-pot B
Calibration Zero
- N2
- left side
- Trim-pot A
500A Output Range Selector Switch Function
The voltage output of the 510 module