5. Mammo Flashcards
difference in mammo XR
- lower energy monoenergetic beam to enhance attenuation differences
- increase in spatial resolution to see microcalculations
mammo – optimal kVp
ideal kVp: 16-23 keV ; voltage of 25-30 kVp is used (Dx is usually 50-120)
-PE dominates at low kvP; PE proportional to Z^3 and indirectly with energy 1/E^3
mammo – target, characteristic x-rays
mammo uses molybednum or rhodium anode
moly: 18 keV ; rhodium: 20.2 keV
why is moly used for mammo
low atomic number, generate low K characteristic x-rays
peak characteristic x-rays with low Bremsstrahlung
k-edge filtration
filter placed outside of tube to create mono-energetic beam (<15 keV and >20 filtered out)
when is Rh vs Mo used?
Rh used for denser breasts since it’s higher energy
why would Mo anode and Rh filter be used?
intermediate energy spectrum
would you ever use Rh target (21 kev) and Mo filter (20 kev K edge)?
NO
when is Mo anode used with aluminium filter?
harder beam to penetrate denser breasts
increasing the filter k edge will result in beam ? and contrast ?
increases average energy of beams
reduces contrast
mammo – focal spot size
0.1-3 small; better spatial resolution
general x ray 0.6-1.2 mm
problems with small focal spot size
cannot tolerate heat well, must have lower mA (and requires longer exposure time)
mA 50 for 0.1 mm focal size
mA 100 for 0.3 mm focal size
mammo – exposure time
longer than xray
mammo ~1 sec
abdominal x ray 0.05 sec
CXR 0.005 sec
heel effect and mammography; which side does cathode go on?
cathode goes on chest wall
tube angled up 20 degrees
effective anode angle
effective anode angle = anode + tube tilt angle
diagnostic tube for mammo vs xr?
mammo: beryllium exit window (less attenuation)
Dx: glass
reason for breast compression in mammo?
reduced thickness/scatter so lower kVP can be used
lower kVP/scatter so improved contrast
improved tissue penetration, so less dose/scatter
reduced thickness so less mAs/dose needed
no motion so less motion/focal blur and spatial resolution
less geometric magnification
less tissue overlap
antiscatter grid in mammo
4-5 grid ratio (height/width)
smaller grid ratio and lower kVp (reduces scatter)
problems with grid
increased dose with grid since grid removes scatter
bucky factor for mammo vs XR
mammo: 2
XR: 5
mag view of breast?
move breast away from detector and closer to the source/focal spot
cutting distance to source in half doubles the mag
is there a grid in mag view?
no. instead the air gap is used as a grid
the air gap allows off axis scatter to miss the target
major changes with mammography and XR
smaller focal spot less mA (to not melt focal spot) longer exposure time to compensate for increase distance between breast tissue and image receptor
difference between mammo and mag view
no grid, even smaller focal spott, increase exposure time, less mA
optical density in mammo
high maximum optical density, most useful at periphery of breast (skin) where thickness decreases