190722_Scavenging Capnography and CO2 Absorption Flashcards
Scavenging collection of excess gases
from equipment used in administering anesthesia, or exhaled by patient.
Scavenging removal of excess gases
to an appropriate place of discharge outside the working environment
NIOSH Recommended Levels of Anesthetic Gases in OR: Volatile Halogenated Anesthetic alone
2 ppm
NIOSH Recommended Levels of Anesthetic Gases in OR: Nitrous Oxide
25 ppm
NIOSH Recommended Levels of Anesthetic Gases in OR: Volatile Anesthetic with Nitrous Oxide
0.5 ppm
5 basic components of scavenging system:
- Gas collecting assembly
- Transfer means
- Scavenging interface
- Gas disposal tubing
- Gas disposal assembly
Gas Collecting Assembly
- Captures excess gases at the site of emission (from circuit).
- Delivers them to the transfer means tubing.
- Outlet connection usually 30mm (19mm on older machines) male-fitting.
- Size of connections is important so that it doesn’t connect to other components of breathing system.
- APL Valve
- APL By-pass Valve
- Exhaust Valve
Transfer Means
- Also called exhaust tubing or hose and transfer system.
- Conveys gas from the collecting assembly to the interface.
- Usually a tube with female-fitting connectors on both ends.
- Tubing is short and large diameter, to carry a high flow of gas w/o a significant increase in pressure.
- Must be kink resistant.
- Must be different from breathing tubes
- Color coded yellow and stiffer plastic
Scavenging Interface
- Prevents pressure increases or decreases in the scavenging system from being transmitted to the breathing system.
- Also called the balancing valve, or balancing device.
- Interface limits pressures immediately downstream of the gas-collecting assembly to between -0.5-+3.5cm H2O.
- Inlet should be 30mm male connector.
- Should be situated as close to gas-collecting assembly as possible.
3 basic elements of the Scavenging Interface
- Positive pressure relief-protects patient and equipment in case of occlusion of system.
- Negative pressure relief-limit sub-atmospheric pressure.
- Reservoir capacity-matches the intermittent gas flow from gas collecting assembly to the continuous flow of disposal system.
Scavenging Interface - 2 Types
Open or Closed
Open Scavenging Interface
- No valves - is open to the atmosphere via “relief ports” in reservoir, avoiding buildup of positive or negative pressures.
- Require use of a central vacuum system and a reservoir (open canister –size should allow for high waste gas flows).
- Gas enters the system at the top of the canister and travels through a narrow inner tube to the base.
- Vacuum control valve can be adjusted – varies the level of suction on the canister/reservoir – must be > excess gas flow rate to prevent OR pollution
Closed Scavenging Interface:
POSITIVE-PRESSURE RELIEF ONLY
- Single positive-pressure relief valve opens when a max. pressure is reached
- Passive disposal – no vacuum used, no reservoir bag needed
Closed Scavenging Interface:
POSITIVE-PRESSURE AND NEGATIVE-PRESSURE RELIEF
• Has a positive-pressure relief valve, negative-pressure relief valve, and a reservoir bag.
• Used with an active disposal systems -Vacuum control valve adjusted so that the reservoir bag is NOT over distended or completely deflated
• Gas is vented to the atmosphere if the system pressure exceeds + 0.5 cm H2O
• Room air is entrained if the system pressure is less than -0.5 cm H2O.
- A backup negative-pressure relief valve opens at -1.8 cm H2O if the primary negative-pressure relief valve becomes occluded.
Gas-Disposal Tubing
- Connects the scavenging interface to the disposal assembly.
- Should be different in size and color from the breathing system.
- With a passive system the hose should be short and wide.
- Tubing running overhead ideal to prevent accidental obstruction and kinking
- If connected to an active gas disposal system it must be a DISS connector
Gas-Disposal Assembly
• Consists of components used to remove waste gases from the OR.
Gas-Disposal Assembly - Active
a mechanical flow-inducing device moves the gases (produces negative pressure in disposal tubing; must have negative pressure relief)
• These systems connect the exhaust of the breathing system to the Hospital vacuum system via an interface controlled by a needle valve.
Gas-Disposal Assembly: Passive
pressure is raised above atmospheric by the patient exhaling, manual squeezing of the reservoir bag or ventilator (needs positive pressure)
The waste gases is directed out of the building via:
• An open window
• A pipe passing through an outside wall
• An extractor fan vented to the outside air
in what type of Gas-Disposal Assembly?
Passive
Gas-Disposal Assembly - Passive
Advantages:
inexpensive to set up
simple to operate
Gas-Disposal Assembly - Passive
Disadvantages:
may be impractical in some buildings
Gas-Disposal Assembly - Active
Advantages:
convenient in large hospitals where many machines are in use in different locations.
Gas-Disposal Assembly - Active
Disadvantages:
vacuum system and pipework is a major expense.
Needle valve may need continual adjustment.
Most common Gas-Disposal Assembly used in hospitals
Active system