There a two basic configurations of regulators what are they?
Low pressure (50 to 500), High pressure (50 to 2000)
The relief valve shall vent at least how many liters per minute?
100
The time required to fill the converter shall not exceed how many minutes? What is the filling pressure?
10 minutes
30 PSIG
What is the temperature of LOX in its liquid state? What is the expansion rate ?
- 297
862: 1
At 32,000ft ambient air shuts off and user receives what percent oxygen ?
100%
CRU-72?
High pressure regulator
50 to 2000
100% oxygen at 32,000
What is the maximum altitude of CRU-103?
50,000
When performing converter leakage test where is the hose assembly connected?
Bell Jar bottom coupling (C-1)
What is the relief valve setting on a LOX converter?
100-120 PSIG
When is the relief valve set to relieve excess pressure in the converter assembly?
110 PSIG
What CRU Reg does E-2/C2 uses?
CRU-53
When is the emergency position used on panel mount regulator?
When inadequate oxygen supply is suspected
What max allowable weight for a Lox converter in vented mode ?
5 Lbs
What is the max allowable weight for a lox converter in build up and supply mode?
3 lbs
What color are oxygen cylinders?
Green with 3 inch white stripes
How many people are required for handling Lox ?
Two or more
What is the manual for Gas cylinder handling and maintenance?
Navair 06-20-2 series
How much built up explosive pressure can a Lox converter make if in a confined space ?
12,000 PSIG
How high is a drip pan suppose to be ?
6 inches
Keep open flames away from oxygen equipment how many feet ?
100 ft
What is the operating pressure of a Lox converter?
55 to 90 PSIG
What is the delivery rate for a Lox converter?
120 LPM
What is the max buildup time for a lox converter?
5 minutes
What is burst disc rupture range ?
225psig at 72 degrees
What is the maximum allowable leakage for a lox converter with 95psig applied ?
0.01
What is the temperature range for ultrasonic cleaner?
32 F to 212F
What is the capacity of an ultrasonic cleaner ?
1.9 gallons
What is the weight of an ultrasonic cleaner?
41 1/2 pounds
What voltage does a ultrasonic cleaner use ?
110/115 volts
What is the operating pressure of a portable oxygen system?
50 to 2000psi
When does 100% oxygen begin in a portable oxygen system?
27,000ft
Air/oxygen dilution in a portable oxygen system begins at what feet ?
30,000
What is the operating altitude of a portable oxygen system?
43,000
What is the size of a portable oxygen system cylinder?
96 cubic inch
What is the recommended inter pressure for a Cru-103?
5 to 120 PSIG
The pressure-breathing feature maintains a positive pressure in the mask hot to how many inches in water ?
20in H2O at altitudes between 34,000 to 50,000 ft
RV-3 On LOX converter?
5-15 psi set at 10 psi
RV-4 on lox converter?
0-500 psi rand set at 180
C-2 on lox converter?
Bell Jar Top coupling