12. Diathermy Flashcards
What are the basic principles of
diathermy?
What is it
/ Used for
How do the heat
WHat is the heating effect dependent on
what are the modes
> Diathermy devices are surgical instruments
used to cut tissues and
coagulate blood vessels.
> They use the heating effect of high frequency AC (0.5–1.0 M Hz) passing through tissues of high impedance to burn or vaporise tissues in contact with the diathermy instrument.
> The heating effect of a current depends on the current density and duration.
> In diathermy the current density
at the point where the instrument makes
contact with the tissue is very high.
> In ‘cutting’ mode the current flows
in an alternating sine wave pattern
while in ‘coagulation’ mode the
current flows in a pulsed sine wave pattern.
> Diathermy devices can either be monopolar or bipolar.
> In monopolar diathermy,
what are the components
where is the current density high and low
The overall power that can be delivered
is in the region of -
> In monopolar diathermy,
current flows through a probe
(active electrode)
at a high current density
and then returns
via a diathermy plate
(neutral plate) at low current density.
100–400 W .
> In bipolar diathermy
> In bipolar diathermy,
the current is passed between
two probes within a
modified pair of forceps.
One probe delivers the current and
the second acts as the return circuit
so tissues between the probes
are exposed to the current and heated.
This system helps keep the electrical field focal.
The overall power this system
can generate is approximately
40 W .
Why is the diathermy pad always checked at the end of surgery?
In bipolar diathermy the current
makes a short journey
across the tissue and
is returned at a similar density,
heating all the tissue between the electrodes.
In monopolar diathermy the current exits
at a point distant
from the site of surgery,
so it is important to ensure that the
current density is low enough
not to cause injury at this point.
Hence the neutral plate has a large surface area and is applied over an area of good blood supply so that the current density is low
and
any heat generated can be dissipated by the blood flow.
If the plate is not applied properly the
reduced area
or
increased impedance
can lead to more heat generation and burns,
hence the reason the plate site
is checked at the end of surgery.
What are the hazards of diathermy?
1
> Burns
2
> Electric shocks
3
> Pacemaker interference
4
> Monitor interference
1
> Burns
- Incorrect positioning of the neutral plate
- Ignition of flammable skin preps
• Inadvertent activation of diathermy probe
(minimised by audible note when in use and use of a designated holder)
2
> Electric shocks
• Disconnection of the neutral plate
may lead to current passing through
an alternative route
(e.g. ECG electrodes
or
exiting through a site
where the patient may be in contact
with a conducting surface).
This can cause electrical shocks
and burns to the patient.
3
> Pacemaker interference
3
> Pacemaker interference
• Ideally, diathermy should be avoided
in patients with pacemakers and
implantable cardioverter defibrillators (ICD) as electrical interference
can cause these devices to malfunction.
• Pacemaker devices must be checked prior to surgery and alternative pacing techniques (external or transvenous) and cardiac arrest trolley should be available.
ICDs should be disabled during surgery in which
diathermy will be used.
- Surgical and theatre team must be informed.
- Bipolar is safer than monopolar diathermy.
• If monopolar diathermy is absolutely necessary,
its use should be limited to
short bursts of less than 5 seconds.
The neutral plate must be well adhered
and
sited as far away from the pacemaker a possible.
• ECG monitoring is paramount.
4
> Monitor interference
4
> Monitor interference
• Diathermy interferes with monitoring including
pulse oximetry,
ECG
and oesophageal Doppler.
What design features are incorporated within the diathermy system to minimise risk of electrical shock?
1
> Use of an isolated patient circuit
2
> Use of an isolating capacitor:
Reactance of capacitors reduces
with increasing current frequency;
therefore,
they have a higher impedance to
low-frequency currents
(i.e. they block conduction of AC
and minimise the risk of macro- and microshocks)
but a lower impedance
to high frequency diathermy currents
(i.e. they allow conduction of these currents)