XI] Electrical safety Flashcards
Who is concerned for protection for technical safety of medical device ?
Patients and operators
How much Amperes for let go current ?
10mA-100mA
How much amperes for begining of physical discomfort ?
100mA to 100A
what is the relashionship between the percieved mA and frequency of the excitation ?
Low treshold for low frequency
Name risks factors for the heart by electrical current
route in the body frequency of the applied current duration moment within the cardiac excitation (vulnerable Tphase) tritial state of the heart
what are the subcategorization fo laws for patient and operator protection ?
general instruction
special instruction for operators
special instructions for manufacturers
persecution and penalty of criminal facts
public authorities, tsting centern authorized experts
Cite the main ressources for device safety
MPG MDD AIMD IEC/EN 60601 EN ISO 14971 DIN EN ISO 13485
Cite the 3 main axes for protection against electronic shock
B(ody)
B(ody) F(loating)
C(ardiac) F(loating)
What current for housing leakage current without contact to patient ?
500
What current for housing leakage current with external contact to patient ?
100
What current for housing leakage current with direct contact to heart ?
100
What current for patient leakage current in µA when external patient contact ?
50
What current for patient leakage current in µA with direct contact to the heart ?
10
what are the 3 protection classes ?
Techniques with earth conductor
without earth conductor
safety extralow voltage
Cite the type of first failure
Interruption of the earth conductor
interruption of a conductor
occurence of a voltage at the signal in/output
Occurence of an external Voltage at signal in/output
Malfunctioning of a mechanical or electrical component
malfunctioning of a t° restrincting component