x-ray circuit and tube heat management Flashcards
chapter 6
All of the following devices are located within the low voltage circuit and control console except the :
Step- up transformer
The autotransformer s primary purpose is to vary the:
voltage
which transformer is located in the filament circuit?
step-down transformer
The primary purpose of the filament circuit is to :
heat the x-ray tube filament for thermonic emission.
A timer that is capable of producing ultrashort exposure times is typical of an
electronic timer
The primary purpose of a rectifier in an x-ray circuit is to
change alternating current (AC) into direct current (DC).
the primary purpose of the high voltage circuit is to :
supply the x-ray tube with voltage high enough to produce x-rays.
The advantages of using a high frequency generator instead of a single phase generator include : 1. producing x-rays more efficiently. 2. requiring less exposure time to produce a given amount of exposure 3. producing the greatest amount of x-rays for the same exposure technique. ( 1,2,3,)
1 and 2 and 3
how much can the exposure time be decreased when using three phase x-ray equipment?
40 percent
automatic exposure control AEC automatically varies the : exposure time
exposure time
which of the following steps should be used to extend x-ray tube life
- warm up the anode 3. do not make repeated exposures near the tube limits
how many heat units are generated using high frequency generator at 200 mA 0.10 sec and 85 kvp ?
2380
For which type of x-ray exposure system can the technical factors be programmed into the system?
Anatomically programmed radiography (APR)
which transformer is located in the high -voltage circuit ?
step up transformer
how much higher is the percentage of x-rays produced in a three phase x-ray machine
40 percent