Week 3 Quality Assurance Flashcards
quality assurance
Program that is designed to control and maintain the standard of quality set for that program
Rad Onc: set of policies and procedures to maintain quality of patient care
set collectively by the profession
Dosimetric Accurary
+/- 5%: equipment calibration, treatment planning, patient setup
Errors: random and systematic,
Ion chamber uncertainty 1.5%
Estimated overall machine calibration uncertainty 2.5%
Acceptance testing required
New piece of equipment- linac, sim, brachy sources
Linac Acceptance testing
Radiation survey-does not exceed permissable doses, formal: head leakage, area survey, interloks, warning lights, emergency switches
jaw symmetry (error less than 1mm)
coincidence: collimator axis, light beam axis, cross hairs
light field with xray beam–+/-2mm, 1% on the side
Mechanical iso: collimater, rotate with rod attached, noting distance of rod to iso= tol 2mm diameter circle
Gantry rotation: same rod, with ring, rotate 360, not displacement, tol +/- 1mm
Acceptance: radiation isocenter collimator
collimator: place film on couch, open 1 set of coll wide, other small slit, rotate coll, do 6-7 exposures, repeat for opposite jaws with new film. intersection tolerance 2mm diameter circle
Acceptance: radiation iso treatment table
film on table, open small split in lower jaws, rotate couch, multiple exposures. tol 2mm diameter circle at intersection
Acceptance: radiation iso Gantry
film on table perpendicular, slit opening, 12 exposures at 30 degree intervals, 2mm circle
Acceptance: multiple beam alignment check
more than 1 beam
causes: focal spot displacement, assymmetry of col jaws, displacement in coll or gantry rotation
Split field test
Acceptance: MLCs
TG40
Projected leaf width at iso: film
Cal of leaf positions: film at iso with buildup, expose up to 8 fields, matchlines every 5cm
Leaf travel: max range of leaf travel in both directions
transmission: intra and interleaf, and beneath jaws/leaves
leakage: leakage when mlc leaves closed, check w ion chamber or film
Field shaping: series of irregularly shaped fields
IMRT additional tests
Acceptence Electrons: Energy
Depth dose distribution: practical method of specifying clinical beam energy
CAX curve is measured with ion chamber in a water phantom and compared to published data to specify the energy
10 x 10 cm field size, 100cm SSD, and 10cm depth
Acceptence Electrons:Field Flatness
Traditional definition: the transverse variation of dose relative to the CAX over the central 80% of the field size at 10cm depth plane perpendicular to the CAX
Tolerance: ± 3%
Flattening filter & penumbra
Acceptence Electrons: Field symmetry
Beam profiles for flatness can be used here
Tolerance: 2% at any pair of points situated symmetrically with respect to the CAX
Acceptance: wedge
10 x 10 field
Tolerance: ± 2 degrees of values specified
The angle between the isodose curve and the line perpendicular to the central axis at a specified depth (often 10cm).
Side Note: Wedge Angle = (180o – hinge angle)/2
Acceptance: Conventional Sim
Geometric and special accuracies
Performance evaluation of the x-ray generator and the associated imaging system
If used for mimicking the LINAC, then geometric accuracy should be comparable
Acceptance: CT sim
Accuracy of image from the CT scanner
Alignment of simulation hardware
Lasers
Couch alignment under typical load conditions
Accuracy of target localization & accuracy of DRRs
+/- 5HUs for daily tolerance