week 2 - anesthesia machines Flashcards
Computer-driven machines should be turned ___ and restarted every __ hours, this is called a __-___
off
24
self-checkout
Anesthesia workstations and their components are covered by the ____
American Society for Testing and Materials (ASTM)
Only __ __ should by plugged into the machine’s electrical outlets.
anesthesia monitors
The backup battery power should last at least __ minutes
30
When gas travels from the high pressure cylinder to the machine, a ___ ___ decreases pressure to __ psig.
pressure regulator
50
The __ __ orients and supports the cylinder, provides a __ __ seal, and ensures ____ flow.
hanger yoke
gas tight
unidirectional
Three components exposed to high pressure are
hanger yoke
cylinder pressure gauge
cylinder pressure regulators
Check valves allow gas to enter the machine but prevent gas from exiting the machine when __
there is no cylinder in the yoke
Cylinder and pipeline pressure gauges are __ __ which stretch with pressure and point to correct pressure (“___” gauges)
metal needles
“Bourdon” gauges
Three pneumatic components of the intermediate pressure system include the __
master switch,
oxygen failure devices,
flow control devices
Four accessories supplied by oxygen include __
ventilators
accessory flowmeters
jet ventilation
oxygen flush
__ __ direct a high unmetered flow directly to the __ __ __ (35-75 L/min) and do not require activation of the __ __
Flush buttons
common gas outlet
master switch
Four flush hazards include __
leaks
sticking and obstructing flowmeter output
barotrauma
awareness
Machines possess a pneumatic or electronic alarm device that sounds a warning when the oxygen supply pressure decreases to below __ psig
30
The “__ __” valve (oxygen pressure sensor shut-off valve) will decrease supply of other gases as the oxygen supply pressure decreases, typically below __ to __ psig.
“fail-safe”
20-25 psig