unit 11 Flashcards

1
Q

What are Artifacts?

A

Answer: Artifacts are errors in imaging which are not actual structures displayed in an image. The ultrasound systems are designed based on certain assumptions. Artifacts are caused by violations of these assumptions. The ultrasound artifacts include:

  • images that have reflectors of improper shape
  • images of reflectors in an inappropriate location
  • images of reflectors of incorrect brightness
  • images that do not correspond to anatomical structures

The artifacts can be caused by equipment malfunction, operator error, or patient motion.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

What is an axial resolution artifact?

A

Answer: Axial Resolution Artifact is the failure to resolve two separate reflectors located parallel to the main axis of the ultrasound beam. The axial resolution artifact appears when spatial pulse length is longer than the distance between two reflectors that are located in front of each other.
The two separate reflectors appear as one reflection.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Two ultrasound transducers are used to perform an exam.
Transducer A has frequency of 5 MHz and 2 cycles per pulse.
Transducer B has a frequency of 2.5 MHz and 4 cycles

A

Answer: The transducer A will have better axial resolution.
The axial resolution artifact appears when spatial pulse length is longer than the distance between two reflectors that are located in front of each other. The two separate reflectors appear as one reflection. The transducers with fewer cycles in an ultrasound pulse have better axial resolution.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

How can the axial resolution artifact be reduced?

A

Answer: The axial resolution artifact appears when a long ultrasound pulse strikes two closely spaced structures.
The axial resolution artifact can be reduced by the use of high frequency transducers. The high frequency transducers produce shorter pulses.
Shorter pulses improve axial resolution and create higher quality images.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

What is a lateral resolution artifact?

A

Answer: Lateral Resolution Artifact is the failure to resolve two separate reflectors located perpendicular to the main axis of the ultrasound beam.
The lateral resolution artifact appears when ultrasound beam is wider than the distance between two reflectors that are located side by side.
The two separate reflectors appear as one reflection.
The lateral resolution is related to the diameter or width of the ultrasound beam.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Two ultrasound transducers are used to perform an exam.
Transducer A has frequency of 7.5 MHz ultrasound beam diameter of 3 mm. Transducer B has a frequency of 5 MHz and ultrasound beam diameter of 8 mm. Which transducer will have better lateral resolution?

A

Answer: The transducer A will have better lateral resolution.
The lateral resolution is related to the diameter or width of the ultrasound beam. The lateral resolution artifact can be reduced by the use of ultrasound transducers which produce smaller diameter or narrow ultrasound beam.

The ultrasound beam with smaller diameter will have better lateral resolution.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

How can the lateral resolution artifact be reduced?

A

Answer: The lateral resolution is related to the width of the ultrasound beam.
The factors that affect ultrasound beam width are:

  • diameter of the piezoelectric element
  • frequency of the ultrasound beam
  • focusing of the ultrasound beam

Small diameter PZT crystals produce narrower ultrasound beams compared to large diameter PZT crystals which produce wider ultrasound beam. The ultrasound beams created by a small diameter crystal will have better lateral resolution and will produce better quality images.
High frequency ultrasound beams are narrower than low frequency ultrasound beams and have better lateral resolution.Focusing of ultrasound beam narrows its diameter within the focal zone area.

Narrower ultrasound beams have better lateral resolution and produce high quality images.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

What are reverberations?

A

Answer: Reverberations are the multiple echoes appearing on the display as a result of ultrasound pulse bouncing between two reflectors.
Reverberations are produced by a pair of strong reflectors that lie along the main axis of an ultrasound beam.

The reverberations are seen on the image as equally spaced bands of diminishing amplitude.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

What is a comet tail artifact?

A

Answer: The comet tail artifact is composed of thin lines of closely spaced echoes.
It appears as a single solid hyperechoic line directed downward.
The reflectors are located parallel to the main axis of the ultrasound beam. Comet tail artifact is produced by a strong reflector and is similar in appearance to reverberation.

The comet tail artifact can be seen posterior to an object that has either a much lower or much higher acoustic impedance than its surroundings.
The reflections of a comet tail artifact are closely spaced because of strong reflectors with high propagation speed and lying close to each other.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

What is the fundamental mechanism of comet tail artifact formation?

A

Answer: The fundamental mechanism in the formation of comet tail artifact is reflection. The ultrasound beam bounces between two closely spaced strong reflectors.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

What are the characteristics of the medium that produces comet tail artifact?

A

Answer: The characteristics of the medium that produces comet tail artifact are high propagation speed and closely spaced strong reflectors.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

What is a Shadowing Artifact?

A

Answer: Shadowing artifact occurs when ultrasound beam is unable to pass through a structure because the structure has a higher than usual attenuation.

Shadowing artifact appears as a hypoechoic or anechoic region extending downward from a very strong attenuating medium. The structures that lie deeper are not imaged or displayed. Structures such as gallstones, renal calculi, and bone produce shadowing artifact.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

How is shadowing artifact expressed?

A

Answer: The shadowing artifact is expressed as reflectors being absent on the image.
Shadowing artifact occurs when ultrasound beam is unable to pass through a structure because the structure has a higher than usual attenuation. The structures that lie deeper than the initial structure are not imaged or displayed.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

How is the shadowing artifact produced?

A

Answer: Shadowing artifact occurs due to attenuation of ultrasound wave.
When ultrasound waves strike a highly attenuating structure such as gall stone, the strength of the reflections from the structures located behind the attenuating structure decreases. Therefore, the structures located behind the attenuating structure cannot be displayed on the image. For example shadowing caused by a prosthetic valve can obscure the visualization of structures behind the prosthetic valve.
The weakening of ultrasound wave can occur with refraction, reflection, and absorption.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

What is an enhancement artifact?

A

Answer: Enhancement artifact appears as a hyperechoic region beneath tissues with low attenuation. Enhancement artifact is an increase in amplitude of echoes located behind a weak attenuating structure. Stronger reflections may be seen behind a fluid filled structure such as urine filled bladder.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

. Acoustic focusing of an ultrasound beam may create which artifact?

A

Answer: Acoustic focusing of an ultrasound beam may create focal enhancement or banding artifact.
Focal enhancement or banding is a form of acoustic enhancement within the focal zone. This results from the increase in beam intensity within the focal zone of a sound beam.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

What is vertical misregistration artifact?

A

Answer: Vertical misregistration occurs when ultrasound pulses travel at the speed less than 1540 m/sec in the body. The reflections on the image are placed deeper than the actual position of the reflector in the body.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

Which artifact is created by sound moving through a medium with a propagation speed less than soft tissue?

A

Answer: Vertical misregistration artifact is created by sound travelling through a medium with a propagation speed less than soft tissue.
If the sound beam passes from a medium of higher speed into a medium of a lesser speed, then the distance is calculated more than the actual distance and the reflectors are placed too deep which is called vertical misregistration.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

What is vertical misregistration artifact?

A

Answer: Vertical misregistration occurs when ultrasound pulses travel at the speed less than 1540 m/sec in the body. The reflections on the image are placed deeper than the actual position of the reflector in the body.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

An ultrasound wave travels through a mass at a speed of 1000 m/sec. What will happen to the position of echoes on the image produced from reflectors distal to the mass?

A

Answer: The reflected echoes on the image will appear farther than the actual position of the reflector.

The propagation speed of mass is less than the propagation speed of soft tissue (1540 m/sec). Therefore, the go return time from the structures located behind the mass will be longer. The ultrasound machine assumes that a long go return time for ultrasound pulse means that it came from a reflector that is located deep in the body. All structures that lie distal to the mass will be placed too deep on the image. The reflected echoes on the image will appear farther than the actual position of the reflector.

21
Q

An ultrasound wave travels through a large cyst at a speed of 1700 m/sec. What will happen to the position of echoes on the image produced from reflectors distal to the cyst?

A

Answer: The reflected echoes on the image will appear closer than the actual position of the reflector.

**The propagation speed of sound in the cyst is higher than the propagation speed of soft tissue (1540 m/sec). **Therefore, the go return time from the structures located behind the cyst will be shorter. The ultrasound machine assumes that a short go return time for ultrasound pulse means that it came from a reflector that is located shallower in the body. All structures that lie distal to the cyst will be placed shallower on the image. The reflected echoes on the image will appear closer than the actual position of the reflector.

22
Q

During an exam, the distance measured between two reflectors on an image is 120 mm. The actual distance between the two reflectors is 100 mm. If the actual distance between two reflectors is 50 mm, what will be the measured distance on the image by the same ultrasound system?

A

Answer: The measured distance will be 60 mm.
In this example, the sound is travelling in a medium which has propagation speed 20% less than the soft tissue and the system is over reporting distance by 20%.
Therefore, if two structures are 50 mm apart, the system will overestimate the distance by 20% and will be 60 mm when measured on the image.

23
Q

During an exam, the distance measured between two reflectors on an image is 40 mm. The actual distance between the two reflectors is 50 mm. If the actual distance between two reflectors is 100 mm, what will be the measured distance on the image by the same ultrasound system?

A

Answer: The measured distance will be 80 mm.
In this example, the sound is travelling in a medium which has propagation speed 20% faster than the soft tissue and the system is under reporting distance by 20%.
Therefore, if two structures are 100 mm apart, the system will underestimate the distance by 20% and will be 80 mm when measured on the image.

24
Q

Which artifacts are affected by the shape or dimensions of an ultrasound pulse?

A

Answer: Lateral resolution artifact, longitudinal resolution artifact and slice thickness artifact are affected by the shape or dimensions of an ultrasound pulse.

25
Which artifacts reduce lateral resolution?
Answer: **Side lobes, grating lobes** and **refraction artifacts** all reduce lateral resolution.
26
Which artifact results in **improper side by side positioning** of reflectors?
Answer: **Refraction artifact results in improper side by side positioning of reflectors.** Refraction is the change in direction of the sound beam as it passes from one medium to another. This causes a reflection to appear improperly positioned on the image. The reflection is placed on the image side by side of the true anatomic structure. **Refraction artifact degrades the lateral resolution.**
27
What are grating lobes?
Answer: **Grating lobes cause reflections to appear at improper**, off axis locations in the image. Grating lobes are produced as a result of **secondary sound waves transmitted from other than the main axis of the ultrasound beam**. **Grating lobes are seen with linear array transducers.**
28
Grating lobes are most common with which type of transducer technology?
Answer: **Grating lobes are seen with linear array transducers**. Grating lobes cause reflections to appear at improper, off axis locations in the image. **Grating Lobes and Side Lobes artifacts are produced by the same mechanism**. The ultrasound energy radiates in direction other than the main axis of the ultrasound beam.
29
Which technique of linear array transducers helps to eliminate the grating lobes artifacts in linear array transducers?
Answer: **Sub dicing technique** helps to eliminate the appearance of grating lobe artifact in linear array transducers. **Sub dicing is a technique in which each PZT crystal is divided into many smaller pieces**. The sub diced pieces are then fired simultaneously as if they were a single crystal.
30
What is apodization?
Answer: **Subdiced elements closer to the center of the beam get a higher voltage, and the outermost subdiced elements get a lower voltage**. This is called apodization. Apodization alters the electrical spike voltages applied to the subdiced elements and helps to reduce the side lobes. Apodization decreases the strength of the voltage sent to the outermost elements.
31
Are Grating lobes attributed only to array transducers?
Grating lobes are most common with linear arrays transducers.
32
What is a side lobes artifact?
Answer: **Inside lobes artifact the acoustic energy is transmitted in directions other than the main axis** of the ***ultrasound beam and generates reflections at improper***, off axis locations in the image. * **A second copy of the true reflector is placed on the image side by side of the true anatomic structure.** **Mechanical or single crystal transducers create side lobes.**
33
What is the cause of side lobe artifact?
Answer: Acoustic energy **radiating in directions other than the main axis of the ultrasound beam causes side lobe artifact.**
34
What are the effects of side lobes on the ultrasound image?
Answer: Side lobe artifact results in: * hollow structures appearing filled in on the image * reflectors appearing in improper locations on the image * reflectors appearing in multiple locations on the image
35
In diagnostic imaging, how does **multipath artifact** affect the image?
Answer: In diagnostic imaging, multipath artifact affects the image by **mild image degradation.** In multipath artifact the **returning echo signals do not follow the same path as the incident beam.** The time required for the echo signals to return to the transducer is different, **causing reflections to appear at incorrect depths.**
36
What is a mirror image artifact?
Answer: A mirror image artifact occurs **when an ultrasound pulse bounces off a strong reflector and then strikes a second reflector.** The path of ultrasound pulse is lengthened in mirror artifact. In mirror image artifact, the deeper image is the artifact. **The copy appears deeper than the true reflector**.
37
Which process causes mirror image artifact?
Answer: The process which causes mirror image artifact is **reflection**. Mirror image is generated when objects present on **one side of a strong reflector are also shown on the other side of the reflector**. Ultrasound system assumes that sound travels directly to a reflector and back to the transducer but sound may bounce off of a strong reflector in its path and be redirected. Such artifacts are commonly seen around the diaphragm.
38
A mirror image artifact can appear alongside of the true anatomy?
Answer: False Mirror image is generated when objects present on **one side of a strong reflector** are **also shown on the other side of the reflector.**
39
What is a slice thickness artifact?
Answer: Slice thickness artifact occurs **when the width of ultrasound beam is greater than the reflector.** Slice thickness artifact results in reflections on the image that do not correlate with the anatomical position of the reflector. **Extra echoes or debris appear in normally anechoic, or echo free structures such as cysts.**
40
What are the other names of slice thickness artifact?
Answer: The slice thickness artifact is also known as partial volume artifact and **section thickness artifact**.
41
What is a Refraction Artifact?
Answer: Refraction at the edge of a circular structure creates an artifact which is called refraction or **edge shadowing artifact**. **The sound beam bends at a curved surface, and changes direction producing a shadow**. **It appears as a hypochoic region extending along the edge of a curved reflector.** It prevents visualization of the true anatomy.
42
What is a ring down artifact?
Answer: Ring down artifact is **similar to reverberations, producing many parallel echoes.** It is produced by a resonance phenomenon and is **associated with gas bubbles**. Sometimes the echoes cannot be differentiated, giving the appearance of a continuous emission of sound.
43
How is range ambiguity artifact eliminated?
Answer: **The range ambiguity artifact can be eliminated by lowering the PRF.** The range ambiguity artifact occurs when **very deep reflections arrive at the transducer after the next pulse was emitted**. The ultrasound system assumes the reflection came from the second pulse and places the late arriving reflection too shallow on the image. The range ambiguity artifact can be fixed by lowering the PRF.
44
What is depth resolution artifact?
Answer: **Depth resolution is the ability of the ultrasound system to distinguish two reflectors that lie close to each other along the main axis of the beam**. The errors in depth resolution result in fewer reflections placed on the image than are actually in the body.
45
What is dead zone artifact?
Answer: **The dead zone artifact is an artifact which prevents ultrasound system to obtain images of structures which are located very close to the transducer.** **It is caused by the time interval required for the ultrasound system to switch from the transmit mode to the receive mode.** The piezoelectric crystals cannot receive signals while vibrating. There is a very small time gap when switching from transmit to receive mode. The echoes received from superficial structures during time gap cannot be processed. The size of dead zone is small and is approximately 5 mm. The size of the dead zone increases with use of low frequency transducers.
46
What is acoustic speckle?
Answer: **Noise arising from small amplitude sound waves interfering with each other is called acoustic speckle**. Acoustic speckle is produced by interference of small ultrasound waves reflected from small particles in the near field. Speckle is scatter in tissue, causing interference effects referred to as noise. It appears as very small amplitude echoes on the image. Speckle artifact has a grainy appearance and is caused by the interference effects of scattered sound. Acoustic speckle tends to slightly degrade the image.
47
What is acoustic speckle?
Answer: Noise arising from small amplitude sound waves interfering with each other is called acoustic speckle. Acoustic speckle is produced by interference of small ultrasound waves reflected from small particles in the near field. Speckle is scatter in tissue, causing interference effects referred to as noise. It appears as very small amplitude echoes on the image. Speckle artifact has a grainy appearance and is caused by the interference effects of scattered sound. Acoustic speckle tends to slightly degrade the image.
48
What is multi path artifact?
Answer: In multipath artifact a strong reflector is encountered on its way back to the transducer which changes the path, making it longer and placing the structure incorrectly deeper on the display. **Multipath artifact can be avoided by changing the transducer angle.**