Unit 1 Flashcards
6 foot cord with deadman switch
6 foot rule
Most popular type of mobile equipment
Battery operated
When was fluoroscopy invented
1896 by Thomas Edison
What is dynamic imaging
Live
Fluoro equipment can be over table or under table tube units T/F
True
Fluoro X-ray tubes use what kind of mA
Lower mA - 0.5-5mA. (Diagnostic operates at 50-1200 mA)
Fixed tube target to prevent SOD of ______ than 15”
Less
Designed to convert remnant radiation from the patient to an amplified light image
Image intensification tube
Emission of light is called
Photoemission
The input fluorescent coating is
0.1-0.2mm sodium-activated cesium iodine (Csl)
T/F the output screen absorbs the electrons and emits light photons
True
Silver activated zinc-cadmium sulfide phosphor
Output screen
The measurement of the increase in image intensity achieved by an image intensification tube
Total brightness gain
Total brightness gain/ minification gain formula
Minification gain= input screen diameter (squared)/ output screen diameter (squared)
The measurement of the increase in light photons due to the conversion efficiency of the output screen
Flux gain