Ultrasound Flashcards

1
Q

Describe Sound

A

Answer: Sound is simply a pressure wave (a form of mechanical energy) that travels in a longitudinal wave

A sound wave is created when a vibrating object (such as a cone in a speaker) sets the molecules of a medium (such as air) into motion

When the vibrating object contacts a medium, the molecules of the medium are compressed together, creating an area of high pressure. This is called compression

When the vibrating object moves away from the medium, it pulls the molecules of the medium along with it. This creates an area of low pressure called a rarefaction.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

What is the difference between Frequency, Amplitude and Wavelength?

A

Frequency is the measure of a pitch.
- It tells us how many cycles occur in a given period of time.
- Measured in Hertz (Hz) or cycles per second

Wavelength is the distance between two identical points on adjacent cycles
- Frequency and wavelength are closely related.
- A higher frequency produces a shorter wavelength
- A lower frequency produces a longer wavelength

Amplitude represents the sound’s loudness (measured in decibels). It is determined by the degree of pressure fluctuations from the displacement of the molecules within the medium.
- A higher amplitude produces a greater pressure change and a louder sound
- A lower amplitude produces a lower pressure change and a softer sound

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

How does sound propagate through tissue?

A

Answer: Propagation velocity of sound through several mediums relavant to ultrasound:
- Air: 343 m/sec
- Soft tissue: 1,540 m/sec
- Bone: 3,000 - 5,000 m/sec

NOTE: when no medium is present (such as in a vacuum or outer space), there is no sound

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

What is echolocation?

A

Answer: The use of sound to visualize physical structures.

EXP: Bats use echolocation as the emission of ultrasonic sound waves by chirping and “listening” as the sound waves bounce off nearby objects.

By calculating how long it took the sound to make a roundtrip and the volume of the returning signal, we can determine the nature of objects around us and their position in space.

By emitting different frequencies, the bat can filter out non-essential information while it focuses on vital information

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

What is the piezoelectric effect?

A

Answer: a piezoelectric material can transducer electrical energy to mechanical energy and vice versa

If you apply electric current to a piezoelectric material, it will vibrate and emit sound

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

How does ultrasound technology work?

A

Answer: The transducer emits ultrasound waves into the body at a fixed rate then listens for echos between each pulse (process is repeated many times each second)

This sequence allows for a rapid refresh rate on the screen –> why the images appear to move dynamically

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Break down one ultrasound pulse listen cycle step by step.

A
  1. THe ultrasound system applies an electric current to the piezoelectric elements inside the transducer, causing them to vibrate
  2. The piezoelectric elements generate ultrasound waves that descend into the body
  3. When an ultrasound wave encounters a boundary between two tissue types (ie: tissues of different acoustic impedance), the wave will echo off the structure and return to the transducer
  4. Echos returning to the transducer make the piezoelectric elements vibrate, causing them to transducer the mechanical energy into an electrical signal
  5. This electrical information is fed into an algorithm that plots the corresponding dot on the screen. This algorithm compares the total sound energy emitted from the transducer and the number and intensity of the returning echos, ultimately revealing a 2D, B-mode image of the patient’s anatomy
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

What determines the placement of each dot on an ultrasound machine?

A

Answer: The vertical placement of each dot is determined by how long it takes for the echo to return to the transducer (time delay)

The horizontal placement of each dot is determined by the particular crystal that receives the returning echo.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

What determines the brightness of each dot?

A

Answer: the amplitude of the returning signals signal and echogenicity describe a tissue’s ability to transmit or reflect sound waves in the context of the surrounding tissues

The relative level of brightness is defined as:

Hyperechoic: they produce a strong, bright (high amplitude) echos
- these tissues have high impendance (EXP: BONE)

Hypoechoic: structures appear as a darker shade of grey
- they produce weak (low amplitude) echos
- these tissues have a lower impedance (IE: Solid organs, skin, adipose and cartilage)

Muscle also tends to be hypo echoic but muscle surfaces often produce hyper echoic fascial lines
- Muscle spindles may appear as small bright dots

Anechoic: structures appear black (they do not produce any echo)
- EXP: Vascular structures, cysts and ascites

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

How do vascular structures appear in short-axis view vs. long-axis view?

A

Answer: Vascular structures appear in black circles in short axis and black tunnels in long axis view

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

How do you compare arteries vs. veins on the ultrasound?

A

Answer: Arteries are pulsatile and relatively non-compressible (although they may be easier to compress if the patient is hypovolemic)

Veins dont pulsate, and they’re easy to compress (although a distended and non-compressible vein should raise suspicion for DVT)

Another way to differentiate veins is to look for their valves in the long axis view
- You can use color Doppler to visualize flow through a vessel

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

How does a nerve structure appear on the ultrasound?

A

Answer: PN imaging can be tough because it can appear hyper echoic or anechoic depending on the region of the body

PN near the neurosis tend to appear anechoic (black)
- Nerves may be confused with vascular structures
- PN are NOT collapsible nor pulsatile

Distal PN are hyper echoic (white) with a characteristic honeycomb appearance
- Distal nerves tend to be enveloped in fascia and connective tissue which accounts for a greater degree of echogenicity while the black circles inside the structures represent the nerves themselves

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

How do you determine if you are looking at a tendon or a nerve on ultrasound?

A

Answer: Tendons may look similar to peripheral nerves but typically without the honeycomb appearance

Best method to differentiate a nerve from the tendon is to scan proximally from a joint towards the muscle

Nerves are continuous and will not change in size as you scan

Tendons become flat and disappear as they connect to muscle. Additionally, they appear “rope-like” at joints where they connect to the bone

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Discuss the resolution of an ultrasound

A

Answer: Axial resolution displays beam depth, lateral resolution determines beam width and elevational resolution determines beam thickness

Axial resolution:
ability to differentiate structures that exist along the length of the ultrasound beam

Lateral resolution
ability to differentiate structures that exist in the width of the ultrasound beam

Elevational resolution
ability to differentiate structures that exist in the thickness of of the ultrasound beam

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

What are the three zones to an ultrasound beam? Where is the image resolution best?

A

Answer: Focal zone, near zone and far zone. Resolution is best in the focal zone but also good in the near zone

Focal Zone = the region where the beam is the narrowest (x- and y-axis) and thinnest (z-axis)

Near zone (Fresnel zone) = the region between the transducer and the focal zone

Far zone (Fraunhofer Zone) = the region beyond the focal zone

Resolution in the far zone is reduced because the distance between each sound wave increases as the beam diverges

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

What is attenuation? What is this result of (four processes)?

A

Answer: Attenuation is when a sound wave propagates through the body and its strength naturally decreases. Additionally some of the sound waves never return to the transducer

Absorption occurs as the ultrasound waves are lost to the body as heat – explains why the wave’s strength naturally degrades as it travels through tissue

Reflection is the process where a sound wave bounces off a tissue boundary of differing acoustic impedance. The greater the difference, the stronger the echo that travels back to the transducer

Scatter occurs when the ultrasound encounters an object smaller than the wave. this causes the echos to scatter in all directions, so the signal never returns to to the transducer

Refraction is the bending of the ultrasound wave that encounters a tissue boundary at an oblique angle – based on Snell’s Law (a formula used to calculate the refraction of light when passing between two media with different refractive indices

17
Q

What is the tradeoff between image resolution and the depth of tissue penetration.

A

The lower frequencies (which have longer wavelengths) allows us to see the deeper inside of the body, but we sacrifice image resolution

Higher frequencies (which have shorter wavelengths) produce the best resolution at the expense of not visualizing deep structures

The best transducer for a given clinical situation produces the highest resolution image while allowing insonation of the relevant anatomy

18
Q

What transducer is best for a given situation? What array configuration should you use?

A

High frequency (>10 MHz): Interscalene, Supravlavicular, Axillary, Forearm, Wrist, Femoral, Ankle and superficial blood vessels.
- Linear Array Transducer

Less frequency (<5 MHz): lumbar plexus, celiac ganglion and neuraxial blocks. Patients with high BMI
= Curvilinear

19
Q

What does the gain on the ultrasound machine do?

A

Answer: Allows you to adjust the strength of the returning echos displayed on the screen

While adjusting the gain can enhance or reduce the screen’s overall brightness, it doe snot change the relative contrast of the anatomy

Excessive gain makes the screen appear too bright

Too little gain makes the screen appear unacceptably dark

20
Q

Describe B-mode vs. M-mode on the Ultrasound.

A

Answer: B mode stands for the brightness of the pixels on the screen. B-mode imaging produces a real-time image of the sonoanatomy, and most bedside ultrasound procedures utilize this modality

M-mode stands for the motion. It is an alternative ultrasound modality. The Y-axis represents the degree of movement of the structures in a defined plane of the body. The x-axis represents time

M-mode is frequently used in echocardiography where it provides useful information about valvular integrity, ventricular function, wall thickness, chamber size and aortic root diameter

21
Q

What is the doppler used for on the ultrasound?

A

Answer: color doppler helps us detect flow through vascular structures. To understand how this works, we must first review the doppler effect.

Doppler effect describes the change in perceived frequency of a sound wave when there’s relative motion between the sound’s source and an observer.

The color of the doppler describes the degree of the doppler shift.

If the direction of the blood flow is moving towards the doppler, it will be red.

If the direction of the blood flow is moving away from the doppler, it will be blue

The doppler shift is greatest when the ultrasound beam is parallel to the blood flow

The doppler shift is zero when the ultrasound is perpendicular to the blood flow. This Is because the cosine of 90 degrees = 0.

21
Q

How should you handle the ultrasound probe?

A

Answer: Optimal imaging necessities that the transducer’s face lay flat on the patents skin. Ensure the angle of incidence is 90 degrees. If the beam meets a structure at 90 degree (perpendicular), more ultrasound waves will reflect towards the transducer, producing a higher-quality image

Correct: holding the transducer near the base (between your fingers and the thumb like a pencil) allows for more exacting movements. Anchoring your wrist and forearm not he patient provides a base of support

Incorrect: Holding the transducer higher up (near the cable) reduces precision and is not recommended

22
Q

What are five ways to move the transducer on the patient?

A

Answer: Tilt, Rock, Slide, Compress, Rotate

Tilt - move the transducer backwards and forward at the same point of contact changes the angle of incidence
- Helps improve image quality by reducing signal loss

Rock - moving the transducer from side to side (while staying in the long axis of the transducer)
- Helpful for better contact between the patient and transducer and helpful for imaging inside a narrow acoustic window

Slide- moving the short axis of the transducer up or down while maintaining the same angle of incidence

Compress - applying force to the transducer so it presses further into the patient’s body

Rotate: moving the transducer in a clockwise or counterclockwise fashion in the same axis of compression

23
Q

What are some clinical examples of artifacts on the ultrasound?

A

Answer: Air, Shadow, Acoustic Enhancement, Mirror image, Reverberation, Bayoneting

24
Q

How are air and shadow different artifacts on the ultrasound? How do we solve these?

A

Air occurs when any part of the transducer’s footprint fails to contact the skin. – APPLY MORE GEL for surface contact

Shadow is a hyper echoic structure that produces strong echoe. Acoustic energy does not penetrate bone well so you will observe an acoustic shadow deep to the hyper echoic border. – Adjust the scanning plane to find a better acoustic window to minimize the influence of a shadow

25
Q

What is the point of point-of-care cardiac ultrasound?

A

Answer: Its greatest value is to help us answer binary (yes/no) questions about the patient’s clinical status.

For example:
1. is there RV or LV failure?
2. Are regional wall motion abnormalities present?
3. Is there valvular dysfunction?
4. Is there a pericardial effusion
5. Is the patient’s volume status adequate?
6. Will the patient be responsive to fluids?
7. Are there gross signs of chronic heart disease?
8. Is there an intracardiac mass

26
Q

What are the three standard imaging windows for cardiac ultrasound.

A

Answer: Parasternal, Apical, Subcostal

27
Q

Discuss the parasternal long-axis (PLAX)

A

Patient postion: left lateral decubitus (brings the heart closer to the anterior chest and reduces influence of lung artifact)

Place phased-array transducer just left of the sternum at the third or fourth intercostal space

Point orientation marker to wards the patient’s right shoulder

Structures seen: LA, LV, MV, AV, Aorta and Pericardium

Interpretation: LV Function, LVH, MV and AV lesions, Pericardial effusions

TEE Equivalent: Midesophageal long axis

28
Q

Discuss the parasternal short axis (PSAX)

A

Pt position: Start with he patient supine. Starting with the position used for the parasternal long axis view, rotate the transducer 90 degrees clockwise towards the patient’s left shoulder

Structures: LV, RV, Pericardium

Interpretation: LV and RV Function, Pericardial effusion

TEE Equivalent: Transgastric short axis

29
Q

Discuss the Apical 4-Chamber (A4CH) view.

A

patient position: left lateral decubitus. Place transducer at point of maximum impulse. Usually just inferolateral to the left nipple in men and under the inferolateral quadrant of the left breast in women

Point transducer orientation marker to pt’s left side with the ultrasound beam pointing towards the patient’s right shoulder.

Structures: RA, RV, LA, RV, AV, Pericardium

Interpretation: LV and RV Function, AV lesions, Pericardial effusion

TEE Equivalent: Midesophageal 4 chamber

30
Q

Describe the 4-Chamber Subcostal (Subcostal 4HC) view/

A

Patient position: Supine

Place transducer in midline just inferior to xiphoid process

Transducer marker should point to patient’s left side. Will need decent amount of pressure – can cause some discomfort

Tip: Ask pt to bend knees to help relax the abdominal musculature.

Structures: RA, RV, LA, LV, AV, pericardial space

Interpretation: RV Function and pericardial effusion

TEE Equivalent: Midesophageal 4-chamber

31
Q

Describe the Subcostal IVC View

A

Patient position: supine in the subcostal 4-chamber view position.

Rotate the transducer 90 degrees

Then tilt beam in the posterior direction

Structures: IVC, RA, Liver

Interpretation: Volume status (IVC collapse suggests hypovolemia)

TEE Equivalent: Bicaval